Long‐Term Performance of Single‐Chamber Atrioventricular‐Synchronous Leadless Pacing: Results From 4 Years of Serial Follow‐up

ABSTRACT Introduction Micra AV enables atrioventricular (AV)–synchronous pacing by sensing atrial mechanical contraction (A4) using device's accelerometer. Short‐term studies showed good AV synchrony (AVS), but long‐term data are lacking. Methods Within the OPTIVALL cohort, we identified patients implanted with a first‐generation Micra AV (June 2020–February 2024) who remained in sinus rhythm and programmed in VDD for ≥ 1 year. Device‐derived AVS was collected annually (tracking index and total AVS = AM‐VP + AM‐VS) up to 4 years. A4 amplitude was recorded at each visit. A subgroup underwent 24‐h Holter‐ECG to assess ambulatory AVS at long‐term follow‐up. Atrial sensing had been optimized early after implant with no further adjustment at follow‐up. Results Of 54 implanted patients, 44 met inclusion criteria. Median (IQR) total AVS remained stable over time: 90.6% (85.4%–93.6%), 90.6% (83.1%–92.1%), 89.2% (84.3%–92.2%), and 89.3% (86.2%–91.6%) at 1, 2, 3, and 4 years, respectively ( p = 0.789). Median (IQR) tracking index was 94.3% (88.9%–97.1%), 93.0% (85.0%–98.0%), 90.2% (84.4%–98.0%), and 90.0% (87.0%–93.9%) ( p = 0.059). Median (IQR) A4 signal amplitude was 3.1 (2.0–5.1), 5.1 (2.1–5.1), 3.7 (1.9–5.1), and 5.1 (2.5–5.1) m/s 2 ( p = 0.993). In 20 patients with 24‐h Holter‐ECG at 2.7 ± 1.1 years, ambulatory AVS was 85.5% (80.0%–90.7%) and was comparable to the 3‐month value (87.9% (76.8%–93.1%); p = 0.811). Conclusions When atrial sensing is optimized early, Micra AV can provide high and stable AV‐synchronous pacing for up to 4 years, with preserved A4 signal amplitude.

Authors

Institutions

Publication Details

Journal
Journal of Cardiovascular Electrophysiology
Published
2026-09-03
DOI
https://doi.org/10.1111/jce.70490
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Long‐Term Performance of Single‐Chamber Atrioventricular‐Synchronous Leadless Pacing: Results From 4 Years of Serial Follow‐up

Silvia Jiménez-Loeches, S Briongos Figuero, Miguel E. Jáuregui-Abularach, David Vaqueriza Cubillo et al.
Journal of Cardiovascular Electrophysiology
Cardiac pacing and defibrillation studies
article

Long‐Term Performance of Single‐Chamber Atrioventricular‐Synchronous Leadless Pacing: Results From 4 Years of Serial Follow‐up

Silvia Jiménez-Loeches, S Briongos Figuero, Miguel E. Jáuregui-Abularach, David Vaqueriza Cubillo, Ana Sánchez-Hernández, Elena Sánchez‐López, Álvaro Estévez‐Paniagua, Roberto Muñoz‐Aguilera, Eloy Gómez‐Mariscal, Beatriz Jiménez‐Gutiérrez
article en

Abstract

ABSTRACT Introduction Micra AV enables atrioventricular (AV)–synchronous pacing by sensing atrial mechanical contraction (A4) using device's accelerometer. Short‐term studies showed good AV synchrony (AVS), but long‐term data are lacking. Methods Within the OPTIVALL cohort, we identified patients implanted with a first‐generation Micra AV (June 2020–February 2024) who remained in sinus rhythm and programmed in VDD for ≥ 1 year. Device‐derived AVS was collected annually (tracking index and total AVS = AM‐VP + AM‐VS) up to 4 years. A4 amplitude was recorded at each visit. A subgroup underwent 24‐h Holter‐ECG to assess ambulatory AVS at long‐term follow‐up. Atrial sensing had been optimized early after implant with no further adjustment at follow‐up. Results Of 54 implanted patients, 44 met inclusion criteria. Median (IQR) total AVS remained stable over time: 90.6% (85.4%–93.6%), 90.6% (83.1%–92.1%), 89.2% (84.3%–92.2%), and 89.3% (86.2%–91.6%) at 1, 2, 3, and 4 years, respectively ( p = 0.789). Median (IQR) tracking index was 94.3% (88.9%–97.1%), 93.0% (85.0%–98.0%), 90.2% (84.4%–98.0%), and 90.0% (87.0%–93.9%) ( p = 0.059). Median (IQR) A4 signal amplitude was 3.1 (2.0–5.1), 5.1 (2.1–5.1), 3.7 (1.9–5.1), and 5.1 (2.5–5.1) m/s 2 ( p = 0.993). In 20 patients with 24‐h Holter‐ECG at 2.7 ± 1.1 years, ambulatory AVS was 85.5% (80.0%–90.7%) and was comparable to the 3‐month value (87.9% (76.8%–93.1%); p = 0.811). Conclusions When atrial sensing is optimized early, Micra AV can provide high and stable AV‐synchronous pacing for up to 4 years, with preserved A4 signal amplitude.

Journal of Cardiovascular Electrophysiology
Research Institute Hospital 12 de Octubre (ES), Hospital Universitario Infanta Leonor (ES)
Reduced inequalities
Openalex Percentile: Top 10%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.