An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of piperacillin in human serum and plasma

OBJECTIVES: Piperacillin is a broad-spectrum antibiotic used to treat critically ill patients with severe infections, requiring therapeutic drug monitoring (TDM). We report the development of an isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure (RMP) to quantify piperacillin in human plasma and serum. METHODS: Primary reference material was characterized by quantitative nuclear magnetic resonance (qNMR) to ensure traceability to the International System of Units (SI). Piperacillin was analyzed using LC-MS/MS operating in positive electrospray ionization and multiple reaction monitoring mode. Method validation evaluated selectivity, matrix effects, precision, accuracy, and measurement uncertainty (MU) according to GUM guidelines. RESULTS: This RMP allowed quantification of piperacillin within the range of 0.773 µmol/L (0.400 µg/mL) to 464 µmol/L (240 µg/mL), with selectivity, sensitivity and matrix-independence. Intermediate precision was <2.3 % and <1.1 % for spiked analyte in free human serum and native patient pools, respectively. The repeatability CV ranged from 0.7 to 2.0 % and relative mean bias ranged from -1.5 to 2.9 % across matrices and concentrations. Single measurement expanded MU ranged from 2.8 % to 5.0 %. Expanded MU (k=2) for target value assignment ranged from 2.0 % to 3.4 % across the primary therapeutic range, reaching 7.1 % total error at the lower limit of the measuring interval (LLMI). CONCLUSIONS: This candidate RMP enables accurate determination of piperacillin in human serum and plasma, providing a standardized platform to facilitate reliable clinical TDM and routine assay harmonization.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2026-09-01
DOI
https://doi.org/10.1515/cclm-2026-0427
Primary Topic
Antibiotics Pharmacokinetics and Efficacy
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of piperacillin in human serum and plasma

Friederike Bauland, Judith Taibon, Michael Vogeser, Carina Schuster et al.
Clinical Chemistry and Laboratory Medicine (CCLM)
Antibiotics Pharmacokinetics and Efficacy
article

An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of piperacillin in human serum and plasma

Friederike Bauland, Judith Taibon, Michael Vogeser, Carina Schuster, Arber Rexhaj, Sandra Fleischer, Andrea Geistanger, Elie Fux, Neeraj Singh
article en

Abstract

OBJECTIVES: Piperacillin is a broad-spectrum antibiotic used to treat critically ill patients with severe infections, requiring therapeutic drug monitoring (TDM). We report the development of an isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure (RMP) to quantify piperacillin in human plasma and serum. METHODS: Primary reference material was characterized by quantitative nuclear magnetic resonance (qNMR) to ensure traceability to the International System of Units (SI). Piperacillin was analyzed using LC-MS/MS operating in positive electrospray ionization and multiple reaction monitoring mode. Method validation evaluated selectivity, matrix effects, precision, accuracy, and measurement uncertainty (MU) according to GUM guidelines. RESULTS: This RMP allowed quantification of piperacillin within the range of 0.773 µmol/L (0.400 µg/mL) to 464 µmol/L (240 µg/mL), with selectivity, sensitivity and matrix-independence. Intermediate precision was <2.3 % and <1.1 % for spiked analyte in free human serum and native patient pools, respectively. The repeatability CV ranged from 0.7 to 2.0 % and relative mean bias ranged from -1.5 to 2.9 % across matrices and concentrations. Single measurement expanded MU ranged from 2.8 % to 5.0 %. Expanded MU (k=2) for target value assignment ranged from 2.0 % to 3.4 % across the primary therapeutic range, reaching 7.1 % total error at the lower limit of the measuring interval (LLMI). CONCLUSIONS: This candidate RMP enables accurate determination of piperacillin in human serum and plasma, providing a standardized platform to facilitate reliable clinical TDM and routine assay harmonization.

Clinical Chemistry and Laboratory Medicine (CCLM)
Evonik (Germany) (DE), LMU Klinikum (DE), Roche Pharma AG (Germany) (DE), Ludwig-Maximilians-Universität München (DE)
Partnerships for the goals
Openalex Percentile: Top 12%
Antibiotics Pharmacokinetics and Efficacy
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.