Open-field and closed-field aberrometry to measure ocular refraction in a prospective semi-randomized crossover study

To assess the agreement and repeatability of closed-field and open-field pyramidal wavefront aberrometers in measuring non-cycloplegic spherocylindrical refraction, compared to subjective refraction and open-field autorefractometry. Fifty healthy adults (27 non-presbyopic, 23 presbyopes) were tested using four methods: closed-field aberrometer (CFA), open-field aberrometer (OFA), open-field autorefractor (OFAUTO), and subjective refraction. Measurements were performed monocularly and binocularly. Significant differences in the M component of the power vector were primarily observed when comparing OFA and subjective refraction, and OFAUTO and subjective refraction, across both age groups and in both eyes, and CFA comparing subjective refraction limited to presbyopic participants. In non-presbyopic participants, CFA showed more negative M values than OFA, consistent with a possible accommodative response or instrument myopia under closed-field non-cycloplegic conditions. No clinically relevant differences were found between monocular and binocular measurements when using the OFA and OFAUTO devices, regardless of age group. All reliability coefficients for the various procedures were relatively high in the Presbyopic group for both eyes. In the Non-presbyopic group, the reliability of M measurements showed a notable decline in the case of CFA in both eyes. OFA demonstrated high reliability comparable to OFAUTO, but a non-negligible clinical difference compared to subjective procedure. The outcomes of CFA in non-presbyopic participants highlights the challenge of accommodation control in non-cycloplegic conditions for closed-field systems. The effect of pupil size warrants further investigation to enhance understanding of objective refraction accuracy compared to the subjective procedure.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-09-01
DOI
https://doi.org/10.1038/s41598-026-68704-2
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Open-field and closed-field aberrometry to measure ocular refraction in a prospective semi-randomized crossover study

Giulia Carlotta Rizzo, Francesco Versaci, Silvia Tavazzi, Fabrizio Zeri et al.
Scientific Reports
Ophthalmology and Visual Impairment Studies
article

Open-field and closed-field aberrometry to measure ocular refraction in a prospective semi-randomized crossover study

Giulia Carlotta Rizzo, Francesco Versaci, Silvia Tavazzi, Fabrizio Zeri, Alessandro Borghesi, AZIZA M.S. OBAID
article en

Abstract

To assess the agreement and repeatability of closed-field and open-field pyramidal wavefront aberrometers in measuring non-cycloplegic spherocylindrical refraction, compared to subjective refraction and open-field autorefractometry. Fifty healthy adults (27 non-presbyopic, 23 presbyopes) were tested using four methods: closed-field aberrometer (CFA), open-field aberrometer (OFA), open-field autorefractor (OFAUTO), and subjective refraction. Measurements were performed monocularly and binocularly. Significant differences in the M component of the power vector were primarily observed when comparing OFA and subjective refraction, and OFAUTO and subjective refraction, across both age groups and in both eyes, and CFA comparing subjective refraction limited to presbyopic participants. In non-presbyopic participants, CFA showed more negative M values than OFA, consistent with a possible accommodative response or instrument myopia under closed-field non-cycloplegic conditions. No clinically relevant differences were found between monocular and binocular measurements when using the OFA and OFAUTO devices, regardless of age group. All reliability coefficients for the various procedures were relatively high in the Presbyopic group for both eyes. In the Non-presbyopic group, the reliability of M measurements showed a notable decline in the case of CFA in both eyes. OFA demonstrated high reliability comparable to OFAUTO, but a non-negligible clinical difference compared to subjective procedure. The outcomes of CFA in non-presbyopic participants highlights the challenge of accommodation control in non-cycloplegic conditions for closed-field systems. The effect of pupil size warrants further investigation to enhance understanding of objective refraction accuracy compared to the subjective procedure.

Scientific Reports
Aston University (GB), University of Milano-Bicocca (IT)
European Commission
Quality Education
Openalex Percentile: Top 11%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.