Interval-Level Validation of a Wearable Dry-Electrode ECG Biosensor Platform: A Proof-of-Concept Study

Wearable dry-electrode electrocardiographic (ECG) systems may enable longer and more comfortable rhythm monitoring, but their interval-level measurement fidelity requires validation against reference acquisition systems. In this single-center proof-of-concept study, 20 participants (10 healthy volunteers and 10 cardiology patients) underwent simultaneous ECG recordings with a U.S. Food and Drug Administration (FDA)-cleared wearable dry-electrode platform (X-trodes System M) and a wired reference system (MP150 with ECG100C module, Biopac Systems). Three-minute paired segments were recorded and underwent temporal synchronization, bandpass filtering, baseline correction, automated discrete-wavelet peak detection in NeuroKit2 (version 0.2.10), and standardized manual adjudication. After predefined exclusions for temporal alignment and P-wave detection quality, agreement was assessed in nine cardiac and ten healthy participants for RR intervals and in seven cardiac and eight healthy participants for P-to-R peak intervals. RR interval agreement was excellent in the healthy and cardiac cohorts, with a mean bias of 0.4 ± 3.1 ms and 0.1 ± 4.6 ms and mean absolute error (MAE) of 0.97 ms and 1.28 ms, respectively. Intraclass correlation coefficient (ICC) was over 0.99 for both cohorts. P-to-R peak interval agreement was strong, with an MAE of 1.76 ms and 7.14 ms and ICC of 0.96 and 0.98, respectively. These findings provide preliminary evidence of interval-level agreement for the X-trodes wearable dry-electrode ECG biosensor platform under controlled, short-duration daytime conditions. Because the matched interval pairs were clustered within a small number of participants, the findings should be regarded as proof-of-concept evidence supporting larger ambulatory validation studies rather than as definitive clinical validation.

Authors

Institutions

Publication Details

Journal
Biosensors
Published
2026-09-01
DOI
https://doi.org/10.3390/bios16090481
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interval-Level Validation of a Wearable Dry-Electrode ECG Biosensor Platform: A Proof-of-Concept Study

Daniel Possti, Avshalom Shaffer, Oliana Vazhgovsky, Yael Hanein et al.
Biosensors
ECG Monitoring and Analysis
article

Interval-Level Validation of a Wearable Dry-Electrode ECG Biosensor Platform: A Proof-of-Concept Study

Daniel Possti, Avshalom Shaffer, Oliana Vazhgovsky, Yael Hanein, Shai Tejman‐Yarden, Deganit Barak‐Shinar, David Hochstein, Roy Beigel, Rona Haker, Yizhaq Shmayahu, Hila Meiri
article en

Abstract

Wearable dry-electrode electrocardiographic (ECG) systems may enable longer and more comfortable rhythm monitoring, but their interval-level measurement fidelity requires validation against reference acquisition systems. In this single-center proof-of-concept study, 20 participants (10 healthy volunteers and 10 cardiology patients) underwent simultaneous ECG recordings with a U.S. Food and Drug Administration (FDA)-cleared wearable dry-electrode platform (X-trodes System M) and a wired reference system (MP150 with ECG100C module, Biopac Systems). Three-minute paired segments were recorded and underwent temporal synchronization, bandpass filtering, baseline correction, automated discrete-wavelet peak detection in NeuroKit2 (version 0.2.10), and standardized manual adjudication. After predefined exclusions for temporal alignment and P-wave detection quality, agreement was assessed in nine cardiac and ten healthy participants for RR intervals and in seven cardiac and eight healthy participants for P-to-R peak intervals. RR interval agreement was excellent in the healthy and cardiac cohorts, with a mean bias of 0.4 ± 3.1 ms and 0.1 ± 4.6 ms and mean absolute error (MAE) of 0.97 ms and 1.28 ms, respectively. Intraclass correlation coefficient (ICC) was over 0.99 for both cohorts. P-to-R peak interval agreement was strong, with an MAE of 1.76 ms and 7.14 ms and ICC of 0.96 and 0.98, respectively. These findings provide preliminary evidence of interval-level agreement for the X-trodes wearable dry-electrode ECG biosensor platform under controlled, short-duration daytime conditions. Because the matched interval pairs were clustered within a small number of participants, the findings should be regarded as proof-of-concept evidence supporting larger ambulatory validation studies rather than as definitive clinical validation.

BiosensorsVol. 16(9)
Reichman University (IL), Tel Aviv University (IL), Sheba Medical Center (IL), Herzliya Medical Center (IL), Edmond and Lily Safra Children's Hospital (IL)
Openalex Percentile: Top 10%
ECG Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.