Accuracy of ventilation feedback devices compared with a reference standard during asynchronous ventilation with continuous chest compressions in a simulated cardiac arrest: a prospective non-randomized cross-over study

AIM This study aimed to evaluate the accuracy of two commercially available ventilation feedback devices (VFD) in measuring ventilation rate and tidal volume during simulated cardiopulmonary resuscitation (CPR) with continuous chest compressions and asynchronous ventilations in the presence of reversed airflow. Methods Prospective, simulation-based, non-randomized cross-over study. Twenty emergency department nurse practitioners performed manual ventilation on an intubated high-fidelity manikin during two 1-minute trials (one per device), targeting 10 breaths/min during continuous chest compressions (100–120/min). Ventilation parameters displayed by EOLifeX® and AccuVent® were compared with measurements from a high-fidelity pneumotachograph (HFRP). Data is presented as mean (standard deviation, min-max range). CI = confidence interval. Results EOLifeX®: From HFRP flow data, 254 ventilations were identified (rate 12/min [1; 8–14], tidal volume 544 mL [32; 482–604]). All showed reversed airflow with an average of 49 mL [13;19–72]. VFD displayed values overestimated ventilation rate and underestimated tidal volume (111/min [5; 99–121] and 104 mL [14; 68–240]; biases +99 breaths/min, 95% CI 96.6–101.4, and −440 mL, 95% CI −456.3 to −423.7). AccuVent®: From HFRP flow data, a total of 222 ventilations were identified (rate 12/min [1; 8–13], tidal volume 532 mL [35; 460–614]). All showed reversed airflow with an average of 56 mL [15; 22–78]. VFD displayed values were similar for rate (12/min [1; 8–13]) but underestimated tidal volume (360 mL [24; 292–398]; bias −172 mL, 95% CI −191.9 to −152.1). Conclusion During simulated CPR with continuous compressions and asynchronous ventilations, VFDs demonstrated clinically relevant inaccuracies. Validation under realistic resuscitation conditions are needed.

Authors

Institutions

Publication Details

Journal
Resuscitation Plus
Published
2026-09-01
DOI
https://doi.org/10.1016/j.resplu.2026.101475
Primary Topic
Cardiac Arrest and Resuscitation
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of ventilation feedback devices compared with a reference standard during asynchronous ventilation with continuous chest compressions in a simulated cardiac arrest: a prospective non-randomized cross-over study

Christophe Duchatelet, Thomas Tackaert, Maxim Vanwulpen, Kristof Segher et al.
Resuscitation Plus
Cardiac Arrest and Resuscitation
article

Accuracy of ventilation feedback devices compared with a reference standard during asynchronous ventilation with continuous chest compressions in a simulated cardiac arrest: a prospective non-randomized cross-over study

Christophe Duchatelet, Thomas Tackaert, Maxim Vanwulpen, Kristof Segher, Said Hachimi-Idrissi, Bert Ingels, Dimitri Muylaert
article en

Abstract

AIM This study aimed to evaluate the accuracy of two commercially available ventilation feedback devices (VFD) in measuring ventilation rate and tidal volume during simulated cardiopulmonary resuscitation (CPR) with continuous chest compressions and asynchronous ventilations in the presence of reversed airflow. Methods Prospective, simulation-based, non-randomized cross-over study. Twenty emergency department nurse practitioners performed manual ventilation on an intubated high-fidelity manikin during two 1-minute trials (one per device), targeting 10 breaths/min during continuous chest compressions (100–120/min). Ventilation parameters displayed by EOLifeX® and AccuVent® were compared with measurements from a high-fidelity pneumotachograph (HFRP). Data is presented as mean (standard deviation, min-max range). CI = confidence interval. Results EOLifeX®: From HFRP flow data, 254 ventilations were identified (rate 12/min [1; 8–14], tidal volume 544 mL [32; 482–604]). All showed reversed airflow with an average of 49 mL [13;19–72]. VFD displayed values overestimated ventilation rate and underestimated tidal volume (111/min [5; 99–121] and 104 mL [14; 68–240]; biases +99 breaths/min, 95% CI 96.6–101.4, and −440 mL, 95% CI −456.3 to −423.7). AccuVent®: From HFRP flow data, a total of 222 ventilations were identified (rate 12/min [1; 8–13], tidal volume 532 mL [35; 460–614]). All showed reversed airflow with an average of 56 mL [15; 22–78]. VFD displayed values were similar for rate (12/min [1; 8–13]) but underestimated tidal volume (360 mL [24; 292–398]; bias −172 mL, 95% CI −191.9 to −152.1). Conclusion During simulated CPR with continuous compressions and asynchronous ventilations, VFDs demonstrated clinically relevant inaccuracies. Validation under realistic resuscitation conditions are needed.

Resuscitation Plus
Flanders Marine Institute (BE), Ghent University Hospital (BE), Ghent University (BE), Onze Lieve Vrouwziekenhuis Hospital (BE), Alma (BE), French Community of Belgium (BE)
Fonds Wetenschappelijk Onderzoek, Universitair Ziekenhuis Gent, Universiteit Gent
Openalex Percentile: Top 7%
Cardiac Arrest and Resuscitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.