Incorporation of the Sysmex DI ‐60 Into the Laboratory's Manual Slide Review Criteria: Assessment of Laboratory Efficiency in a Routine Practice

ABSTRACT Introduction Digital morphology (DM) analyzers may improve laboratory efficiency; however, their practical integration into manual slide review (MSR) criteria and the resulting impact on overall efficiency remain underexplored. We evaluated the efficiency of the DI‐60 (Sysmex, Kobe, Japan) by assessing its potential to modify or replace existing MSR criteria in a routine practice and comparing its turnaround time (TAT) with that of manual counting. Methods A total of 907 samples were consecutively selected based on our laboratory's MSR criteria. We simulated modified MSR criteria incorporating the DI‐60 to assess the potential reduction in laboratory workload. We compared TATs of the DI‐60 and manual counting according to white blood cell (WBC) counts (range, 0.01–158.6 × 10 9 /L). Results DI‐60 achieved complete analysis in 717/907 samples (79.1%), showing stable performance in samples with normal WBC counts. However, the overall TAT of the DI‐60 was significantly longer than that of manual counting (263 vs. 120 s, p < 0.001). Incomplete analysis (< 200‐cell count) occurred in 190 samples (20.9%), predominantly in hematology‐oncology samples (94.7%) and samples with severe leukopenia. Simulation of modified MSR criteria incorporating the DI‐60 demonstrated a potential workload reduction of 11.6% (105/907). Conclusions DM analyzers may have the potential to reduce laboratory workload by being incorporated into each laboratory's MSR criteria. With incomplete analyses in specific sample types, however, DM analyzers cannot fully replace MSR. The effective use of DM analyzers requires streamlined integration into the existing laboratory workflow and further optimization considering each laboratory's situation and unmet needs.

Authors

Institutions

Publication Details

Journal
International Journal of Laboratory Hematology
Published
2026-09-01
DOI
https://doi.org/10.1111/ijlh.70236
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Incorporation of the Sysmex DI ‐60 Into the Laboratory's Manual Slide Review Criteria: Assessment of Laboratory Efficiency in a Routine Practice

Yeo‐Min Yun, Hee‐Won Moon, Hanah Kim, Mi-Hyun Hong et al.
International Journal of Laboratory Hematology
Clinical Laboratory Practices and Quality Control
article

Incorporation of the Sysmex DI ‐60 Into the Laboratory's Manual Slide Review Criteria: Assessment of Laboratory Efficiency in a Routine Practice

Yeo‐Min Yun, Hee‐Won Moon, Hanah Kim, Mi-Hyun Hong, Eunju Shin, Mina Hur
article en

Abstract

ABSTRACT Introduction Digital morphology (DM) analyzers may improve laboratory efficiency; however, their practical integration into manual slide review (MSR) criteria and the resulting impact on overall efficiency remain underexplored. We evaluated the efficiency of the DI‐60 (Sysmex, Kobe, Japan) by assessing its potential to modify or replace existing MSR criteria in a routine practice and comparing its turnaround time (TAT) with that of manual counting. Methods A total of 907 samples were consecutively selected based on our laboratory's MSR criteria. We simulated modified MSR criteria incorporating the DI‐60 to assess the potential reduction in laboratory workload. We compared TATs of the DI‐60 and manual counting according to white blood cell (WBC) counts (range, 0.01–158.6 × 10 9 /L). Results DI‐60 achieved complete analysis in 717/907 samples (79.1%), showing stable performance in samples with normal WBC counts. However, the overall TAT of the DI‐60 was significantly longer than that of manual counting (263 vs. 120 s, p < 0.001). Incomplete analysis (< 200‐cell count) occurred in 190 samples (20.9%), predominantly in hematology‐oncology samples (94.7%) and samples with severe leukopenia. Simulation of modified MSR criteria incorporating the DI‐60 demonstrated a potential workload reduction of 11.6% (105/907). Conclusions DM analyzers may have the potential to reduce laboratory workload by being incorporated into each laboratory's MSR criteria. With incomplete analyses in specific sample types, however, DM analyzers cannot fully replace MSR. The effective use of DM analyzers requires streamlined integration into the existing laboratory workflow and further optimization considering each laboratory's situation and unmet needs.

International Journal of Laboratory Hematology
Konkuk University (KR), Gachon University Gil Medical Center (KR)
Openalex Percentile: Top 11%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.