Biomechanical comparison of mini-plate and screw fixation in minimally invasive distal metatarsal osteotomy for hallux valgus

Aims: To biomechanically compare the stability of mini-plate and dual-screw fixation following minimally invasive distal metatarsal osteotomy for hallux valgus, with an emphasis on early-stage stability under clinically relevant loading conditions. Methods: Three fixation configurations were evaluated: mini-plate fixation with two (PS2) or three (PS3) locking screws and dual-screw fixation (DS). Finite element analysis (FEA) was performed to quantify displacement and stress distribution under controlled bending and torsional loading, providing insight into internal load transfer mechanisms not accessible experimentally. Fresh-frozen human cadaveric specimens then underwent distal chevron metatarsal osteotomy and fixation according to each configuration (n = 3 per group). Cyclic and static bending tests were performed to simulate early postoperative weightbearing, and displacement responses were recorded. FE predictions were compared with experimental displacement trends. Results: Under bending loads, both FEA and cadaveric testing consistently demonstrated that PS3 exhibited the smallest displacement, followed by PS2, with DS showing the greatest instability. FE analysis further revealed that mini-plate fixation reduced bone stress at the osteotomy site but increased implant stress, whereas DS showed higher bone stress concentration. Under torsional loading, FE results indicated lower displacement but higher bone stress in DS. Cadaveric cyclic tests confirmed the superior stability of PS3, showing the least displacement accumulation over 1,000 cycles (p < 0.05). In static testing, PS3 achieved the highest failure load (~423 N), followed by PS2 (~248 N), while DS showed the lowest (~205 N). Conclusion: Mini-plate fixation, particularly PS3, provides superior resistance to bending-induced micromotion compared with DS, with consistent agreement between FE predictions and experimental findings. FEA further elucidated internal stress distribution, highlighting increased bone stress concentrations in DS. Overall, mini-plate fixation offers a more stable and clinically reliable construct for maintaining correction during early postoperative weightbearing.

Authors

Institutions

Publication Details

Journal
Bone and Joint Research
Published
2026-09-01
DOI
https://doi.org/10.1302/2046-3758.159.bjr-2026-0011.r1
Primary Topic
Foot and Ankle Surgery
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biomechanical comparison of mini-plate and screw fixation in minimally invasive distal metatarsal osteotomy for hallux valgus

Hsuan-Wen Wang, Ya-Han Chan, Shao-Fu Huang, Chia-Hung Chu et al.
Bone and Joint Research
Foot and Ankle Surgery
article

Biomechanical comparison of mini-plate and screw fixation in minimally invasive distal metatarsal osteotomy for hallux valgus

Hsuan-Wen Wang, Ya-Han Chan, Shao-Fu Huang, Chia-Hung Chu, Chun‐Li Lin, Keen-Wai Chong, Yan-Yu Chen
article en

Abstract

Aims: To biomechanically compare the stability of mini-plate and dual-screw fixation following minimally invasive distal metatarsal osteotomy for hallux valgus, with an emphasis on early-stage stability under clinically relevant loading conditions. Methods: Three fixation configurations were evaluated: mini-plate fixation with two (PS2) or three (PS3) locking screws and dual-screw fixation (DS). Finite element analysis (FEA) was performed to quantify displacement and stress distribution under controlled bending and torsional loading, providing insight into internal load transfer mechanisms not accessible experimentally. Fresh-frozen human cadaveric specimens then underwent distal chevron metatarsal osteotomy and fixation according to each configuration (n = 3 per group). Cyclic and static bending tests were performed to simulate early postoperative weightbearing, and displacement responses were recorded. FE predictions were compared with experimental displacement trends. Results: Under bending loads, both FEA and cadaveric testing consistently demonstrated that PS3 exhibited the smallest displacement, followed by PS2, with DS showing the greatest instability. FE analysis further revealed that mini-plate fixation reduced bone stress at the osteotomy site but increased implant stress, whereas DS showed higher bone stress concentration. Under torsional loading, FE results indicated lower displacement but higher bone stress in DS. Cadaveric cyclic tests confirmed the superior stability of PS3, showing the least displacement accumulation over 1,000 cycles (p < 0.05). In static testing, PS3 achieved the highest failure load (~423 N), followed by PS2 (~248 N), while DS showed the lowest (~205 N). Conclusion: Mini-plate fixation, particularly PS3, provides superior resistance to bending-induced micromotion compared with DS, with consistent agreement between FE predictions and experimental findings. FEA further elucidated internal stress distribution, highlighting increased bone stress concentrations in DS. Overall, mini-plate fixation offers a more stable and clinically reliable construct for maintaining correction during early postoperative weightbearing.

Bone and Joint ResearchVol. 15(9)
National Yang Ming Chiao Tung University (TW), Singapore General Hospital (SG), Chang Bing Show Chwan Memorial Hospital (TW)
National Science and Technology Council
Openalex Percentile: Top 9%
Foot and Ankle Surgery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.