Optimisation of Q.Clear (BSREM) Reconstruction Parameters for 123I and 177Lu Imaging with a 360° CZT Detector

Accurate quantitative SPECT depends strongly on the reconstruction protocol, particularly for penalised-likelihood reconstruction algorithms, such as Q.Clear (BSREM). This study aimed to optimise Q.Clear reconstruction parameters for 123I and 177Lu imaging. SPECT images of a NEMA IEC PET Body phantom filled with 123I and 177Lu were acquired on a ring-shaped CZT camera and on a conventional Anger camera. The images were reconstructed using Q.Clear and OSEM and evaluated using quantitative image-quality metrics. Q.Clear was optimised for each radionuclide. The 123I dataset was reconstructed using different combinations of the regularisation parameters β and γ. The 177Lu dataset was reconstructed using literature-reported parameters, and the best-performing reconstruction was fine-tuned. The best-performing Q.Clear reconstructions were compared with corresponding OSEM reconstructions. Overall, increasing β reduced the recovery coefficients, background variability, and contrast recovery coefficients while increasing the signal-to-noise ratio. γ had no appreciable influence on the metrics within the investigated range. The candidate optimised StarGuide/Q.Clear reconstructions outperformed the standard Symbia/OSEM reconstructions in terms of activity recovery, whereas Symbia/OSEM produced lower image noise for most of the analysed VOIs. This study identified settings associated with improved quantitative metrics within the evaluated parameter space, with potential implications for quantitative imaging requiring clinical validation. Nevertheless, multiple reconstruction parameter combinations remain unexplored.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-08-28
DOI
https://doi.org/10.3390/bioengineering13091004
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimisation of Q.Clear (BSREM) Reconstruction Parameters for 123I and 177Lu Imaging with a 360° CZT Detector

Vera Antunes, J. M. R. Teixeira, João Santos, Rafaela Carvalhais
Bioengineering
Medical Imaging Techniques and Applications
article

Optimisation of Q.Clear (BSREM) Reconstruction Parameters for 123I and 177Lu Imaging with a 360° CZT Detector

Vera Antunes, J. M. R. Teixeira, João Santos, Rafaela Carvalhais
article en

Abstract

Accurate quantitative SPECT depends strongly on the reconstruction protocol, particularly for penalised-likelihood reconstruction algorithms, such as Q.Clear (BSREM). This study aimed to optimise Q.Clear reconstruction parameters for 123I and 177Lu imaging. SPECT images of a NEMA IEC PET Body phantom filled with 123I and 177Lu were acquired on a ring-shaped CZT camera and on a conventional Anger camera. The images were reconstructed using Q.Clear and OSEM and evaluated using quantitative image-quality metrics. Q.Clear was optimised for each radionuclide. The 123I dataset was reconstructed using different combinations of the regularisation parameters β and γ. The 177Lu dataset was reconstructed using literature-reported parameters, and the best-performing reconstruction was fine-tuned. The best-performing Q.Clear reconstructions were compared with corresponding OSEM reconstructions. Overall, increasing β reduced the recovery coefficients, background variability, and contrast recovery coefficients while increasing the signal-to-noise ratio. γ had no appreciable influence on the metrics within the investigated range. The candidate optimised StarGuide/Q.Clear reconstructions outperformed the standard Symbia/OSEM reconstructions in terms of activity recovery, whereas Symbia/OSEM produced lower image noise for most of the analysed VOIs. This study identified settings associated with improved quantitative metrics within the evaluated parameter space, with potential implications for quantitative imaging requiring clinical validation. Nevertheless, multiple reconstruction parameter combinations remain unexplored.

BioengineeringVol. 13(9)
Universidade do Porto (PT), Instituto Português de Oncologia Francisco Gentil (PT), IPO Porto (PT)
Openalex Percentile: Top 11%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.