Impact of additive solutions on haematocrit measurement and red blood cell integrity: A comparative study of SAGM and PAGGSM

Abstract Background and Objectives Accurate haematocrit (Ht) measurement of red cell concentrates (RCCs) is critical because it is used to calculate the level of haemolysis, a key RCC quality control parameter. The hypothesis was that Ht depends on both the measurement method (microcapillary vs. haematology analysers) and the additive solution (hypertonic saline–adenine–glucose–mannitol [SAGM] vs. isotonic phosphate–adenine–glucose–guanosine–saline–mannitol [PAGGSM]). We quantified method‐ and solution‐related effects on Ht measurement and assessed the downstream consequences for haemolysis. Storage‐induced micro‐erythrocytes (SMEs) were evaluated as a secondary marker of morphological preservation. Materials and Methods RCCs were produced from whole blood ( N = 16). Using a pool‐and‐split design, eight paired RCCs were assigned to SAGM or PAGGSM and stored for 21 days. Ht was measured by HT200 microcapillary and haematology analysers. Red blood cell (RBC) morphology and SMEs were assessed by imaging flow cytometry. Results At Day 1, microcapillary Ht in PAGGSM exceeded analyser‐derived Ht by 3.6%–10%, whereas SAGM Ht was more consistent. Using different Ht sources materially altered haemolysis estimates. Over storage, PAGGSM showed a decline in Ht (65.3%–62.6%) accompanied by reduced RBC projected surface area (73.9–71.01 μm 2 ). SAGM showed stable Ht but reduced RBC projected surface area (72.6–69.1 μm 2 ). SME accumulation was lower in PAGGSM at Day 21 (13.6%) than in SAGM (18.8%). Conclusion Ht measurement method materially impacts haemolysis, underscoring the need to standardize Ht assessment and to prefer microcapillary Ht when calculating haemolysis. Additive solution also influences storage quality: PAGGSM better preserved RBC morphology and reduced SMEs relative to SAGM up to Day 21.

Authors

Institutions

Publication Details

Journal
Vox Sanguinis
Published
2026-08-27
DOI
https://doi.org/10.1111/vox.70358
Primary Topic
Blood properties and coagulation
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of additive solutions on haematocrit measurement and red blood cell integrity: A comparative study of SAGM and PAGGSM

Pascal Amireault, Nadine Marpaux, Michaël Dussiot, Anaïs Lagrange et al.
Vox Sanguinis
Blood properties and coagulation
article

Impact of additive solutions on haematocrit measurement and red blood cell integrity: A comparative study of SAGM and PAGGSM

Pascal Amireault, Nadine Marpaux, Michaël Dussiot, Anaïs Lagrange, Quentin Brebant
article en

Abstract

Abstract Background and Objectives Accurate haematocrit (Ht) measurement of red cell concentrates (RCCs) is critical because it is used to calculate the level of haemolysis, a key RCC quality control parameter. The hypothesis was that Ht depends on both the measurement method (microcapillary vs. haematology analysers) and the additive solution (hypertonic saline–adenine–glucose–mannitol [SAGM] vs. isotonic phosphate–adenine–glucose–guanosine–saline–mannitol [PAGGSM]). We quantified method‐ and solution‐related effects on Ht measurement and assessed the downstream consequences for haemolysis. Storage‐induced micro‐erythrocytes (SMEs) were evaluated as a secondary marker of morphological preservation. Materials and Methods RCCs were produced from whole blood ( N = 16). Using a pool‐and‐split design, eight paired RCCs were assigned to SAGM or PAGGSM and stored for 21 days. Ht was measured by HT200 microcapillary and haematology analysers. Red blood cell (RBC) morphology and SMEs were assessed by imaging flow cytometry. Results At Day 1, microcapillary Ht in PAGGSM exceeded analyser‐derived Ht by 3.6%–10%, whereas SAGM Ht was more consistent. Using different Ht sources materially altered haemolysis estimates. Over storage, PAGGSM showed a decline in Ht (65.3%–62.6%) accompanied by reduced RBC projected surface area (73.9–71.01 μm 2 ). SAGM showed stable Ht but reduced RBC projected surface area (72.6–69.1 μm 2 ). SME accumulation was lower in PAGGSM at Day 21 (13.6%) than in SAGM (18.8%). Conclusion Ht measurement method materially impacts haemolysis, underscoring the need to standardize Ht assessment and to prefer microcapillary Ht when calculating haemolysis. Additive solution also influences storage quality: PAGGSM better preserved RBC morphology and reduced SMEs relative to SAGM up to Day 21.

Vox Sanguinis
Inserm (FR), Université Paris Cité (FR), Sorbonne Paris Cité (FR), Methodology and Quality of Life Unit in Oncology (FR), Macopharma (France) (FR), Biologie intégrée du globule rouge et de l'Erythropoïèse (FR), Institut des Maladies Génétiques Imagine (FR)
Agence Nationale de la Recherche
Openalex Percentile: Top 11%
Blood properties and coagulation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.