What Do Infrequent Goldmann Applanation Tonometry Measurements Reveal About Intraocular Pressure Profiles After Nonpenetrating Glaucoma Surgery?

Purpose: To evaluate how reliably single Goldmann applanation tonometry (GAT) measurements reflect the preceding month's intraocular pressure (IOP) profile in operated open-angle glaucoma. Methods: This longitudinal study included 24 patients with a suprachoroidal telemetric IOP sensor analyzing 156 in-hospital GAT measurements, 154 simultaneous telemetric readings, and 13,734 at-home measurements across follow-up visits. Telemetric data from the 30 days prior to each visit were stratified by time of day and matched to corresponding GAT measurements, which, in turn, were ranked relative to the interquartile range (IQR) of the preceding at-home measurements. Peak-and-trough analyses were performed using hourly aggregated data. Results: The GAT and telemetric IOP measurements showed good agreement (mean ΔIOP = -0.1 ± 3.0 mm Hg). However, 59.7% of the GAT-measured IOP values exceeded the IQR, and 14.3% fell below the IQR of the corresponding telemetric distributions; only 26.0% were within the IQR. Deviations from the telemetric median were 4.2 ± 2.4 mm Hg (above IQR), -2.3 ± 1.5 mm Hg (below IQR), and 0.5 ± 1.2 mm Hg (within IQR). Of the 99 statistically significant extrema (45 peaks and 54 troughs), most peaks occurred within 3 hours before GAT measurements. Troughs were observed both before and after GAT, but with larger nyctohemeral misalignment to GAT. Conclusions: Single GAT measurements frequently failed to represent the underlying 30-day IOP profile and often missed clinically relevant peaks and troughs, despite good pointwise agreement with telemetric measurements. Translational Relevance: High-frequency telemetric monitoring provides a more comprehensive assessment of IOP dynamics and may improve individualized glaucoma management compared with infrequent GAT snapshot measurements.

Authors

Institutions

Publication Details

Journal
Translational Vision Science & Technology
Published
2026-08-27
DOI
https://doi.org/10.1167/tvst.15.8.28
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

What Do Infrequent Goldmann Applanation Tonometry Measurements Reveal About Intraocular Pressure Profiles After Nonpenetrating Glaucoma Surgery?

Marc J. Mackert, Karl T. Boden, Colya N. Englisch, André M. Trouvain et al.
Translational Vision Science & Technology
Glaucoma and retinal disorders
article

What Do Infrequent Goldmann Applanation Tonometry Measurements Reveal About Intraocular Pressure Profiles After Nonpenetrating Glaucoma Surgery?

Marc J. Mackert, Karl T. Boden, Colya N. Englisch, André M. Trouvain, H. Burkhard Dick, Philip Wakili, Esther M. Hoffmann, Kaweh Mansouri, Peter Szurman
article en

Abstract

Purpose: To evaluate how reliably single Goldmann applanation tonometry (GAT) measurements reflect the preceding month's intraocular pressure (IOP) profile in operated open-angle glaucoma. Methods: This longitudinal study included 24 patients with a suprachoroidal telemetric IOP sensor analyzing 156 in-hospital GAT measurements, 154 simultaneous telemetric readings, and 13,734 at-home measurements across follow-up visits. Telemetric data from the 30 days prior to each visit were stratified by time of day and matched to corresponding GAT measurements, which, in turn, were ranked relative to the interquartile range (IQR) of the preceding at-home measurements. Peak-and-trough analyses were performed using hourly aggregated data. Results: The GAT and telemetric IOP measurements showed good agreement (mean ΔIOP = -0.1 ± 3.0 mm Hg). However, 59.7% of the GAT-measured IOP values exceeded the IQR, and 14.3% fell below the IQR of the corresponding telemetric distributions; only 26.0% were within the IQR. Deviations from the telemetric median were 4.2 ± 2.4 mm Hg (above IQR), -2.3 ± 1.5 mm Hg (below IQR), and 0.5 ± 1.2 mm Hg (within IQR). Of the 99 statistically significant extrema (45 peaks and 54 troughs), most peaks occurred within 3 hours before GAT measurements. Troughs were observed both before and after GAT, but with larger nyctohemeral misalignment to GAT. Conclusions: Single GAT measurements frequently failed to represent the underlying 30-day IOP profile and often missed clinically relevant peaks and troughs, despite good pointwise agreement with telemetric measurements. Translational Relevance: High-frequency telemetric monitoring provides a more comprehensive assessment of IOP dynamics and may improve individualized glaucoma management compared with infrequent GAT snapshot measurements.

Translational Vision Science & TechnologyVol. 15(8)
Johannes Gutenberg University Mainz (DE), LMU Klinikum (DE), Knappschaftsklinikum Saar (DE), University Medical Center of the Johannes Gutenberg University Mainz (DE), Universitätsklinikum Knappschaftskrankenhaus Bochum (DE), Ludwig-Maximilians-Universität München (DE), Ruhr University Bochum (DE), Saarland University (DE), University of Colorado Denver (US)
Quality Education
Openalex Percentile: Top 7%
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.