Micro-CT evaluation of canal shaping extent with apex locator–integrated endodontic motors: An in vitro study

INTRODUCTION: This in vitro study evaluated the extent accuracy of canal shaping to apical constriction (AC) with electronic apex locator (EAL)-integrated endodontic motors using micro-computed tomography (MicroCT). MATERIALS AND METHODS: Sixty single-canalled teeth were decoronated, scanned using microCT, and distributed into three groups: Rooter Universal, Tri Auto ZX II, and control group (n = 20 each). The working length (WL) was determined simultaneously during canal shaping using the auto-reverse function set at the 0.5 mark in the RU and TA groups. In the control group, the WL was measured before the shaping using Root ZX (J. Morita, Tokyo, Japan). In the three groups, canals were shaped using ProTaper Gold files (Dentsply Sirona, Ballaigues, Switzerland) until the preset mark was reached in the two experimental groups and to the measured WL in the control group. Afterwards, post-shaping microCT scans were taken. Scan superimposition was conducted to assess the absolute distance between the final shaping endpoint and AC. Negative and positive data were recorded when the measurements were found short or beyond the AC, respectively. Analysis of variance (ANOVA) and post-hoc Tukey tests were used to analyze the distance. The chi-square test was applied to analyze the accuracy of the measurements with respect to the AC. The significance level was set at 5%. RESULTS: No over-shaping beyond the apical foramen was found in any sample, and all systems showed similar accuracy when a tolerance of 0.5 mm was set ((V = 0.263, P = 0.126). However, the shaping extent to AC was closer in the control group than the RU (P = 0.001) and TA (P < 0.001), while RU and TA were similar (P = 0.588). CONCLUSIONS: The accuracy of WL determination was not affected by the tested EAL. However, the standalone EAL was more reliable at maintaining AC integrity.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2026-08-27
DOI
https://doi.org/10.1371/journal.pone.0355939
Primary Topic
Endodontics and Root Canal Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
article

Micro-CT evaluation of canal shaping extent with apex locator–integrated endodontic motors: An in vitro study

Ahmed Jamleh, Anas Al‐Jadaa, Zainab Shirazi, Dalya Sharaf et al.
PLoS ONE
Endodontics and Root Canal Treatments
article

Micro-CT evaluation of canal shaping extent with apex locator–integrated endodontic motors: An in vitro study

Ahmed Jamleh, Anas Al‐Jadaa, Zainab Shirazi, Dalya Sharaf, Abdulrahman Mohammed Saleh
article en

Abstract

INTRODUCTION: This in vitro study evaluated the extent accuracy of canal shaping to apical constriction (AC) with electronic apex locator (EAL)-integrated endodontic motors using micro-computed tomography (MicroCT). MATERIALS AND METHODS: Sixty single-canalled teeth were decoronated, scanned using microCT, and distributed into three groups: Rooter Universal, Tri Auto ZX II, and control group (n = 20 each). The working length (WL) was determined simultaneously during canal shaping using the auto-reverse function set at the 0.5 mark in the RU and TA groups. In the control group, the WL was measured before the shaping using Root ZX (J. Morita, Tokyo, Japan). In the three groups, canals were shaped using ProTaper Gold files (Dentsply Sirona, Ballaigues, Switzerland) until the preset mark was reached in the two experimental groups and to the measured WL in the control group. Afterwards, post-shaping microCT scans were taken. Scan superimposition was conducted to assess the absolute distance between the final shaping endpoint and AC. Negative and positive data were recorded when the measurements were found short or beyond the AC, respectively. Analysis of variance (ANOVA) and post-hoc Tukey tests were used to analyze the distance. The chi-square test was applied to analyze the accuracy of the measurements with respect to the AC. The significance level was set at 5%. RESULTS: No over-shaping beyond the apical foramen was found in any sample, and all systems showed similar accuracy when a tolerance of 0.5 mm was set ((V = 0.263, P = 0.126). However, the shaping extent to AC was closer in the control group than the RU (P = 0.001) and TA (P < 0.001), while RU and TA were similar (P = 0.588). CONCLUSIONS: The accuracy of WL determination was not affected by the tested EAL. However, the standalone EAL was more reliable at maintaining AC integrity.

PLoS ONEVol. 21(8)
Ajman University (AE), University of Sharjah (AE)
Openalex Percentile: Top 8%
Endodontics and Root Canal Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.