Optimizing Oscillating Diffusion Gradients to Reduce Mechanical Vibration and Acoustic Noise on Ultra‐High Gradient Scanners

PURPOSE: The deployment of oscillating gradient spin echo (OGSE) diffusion encoding sequences on ultra-high gradient scanners enables advanced microstructure imaging but induces intense acoustic noise and mechanical vibrations, causing subject discomfort and compromising imaging stability and quantitative accuracy. This study implements Acoustic noise and Vibration Optimized dIffusion gradient Design (AVOID) to mitigate these side effects, facilitating the high-performance OGSE applications in demanding hardware environments. METHODS: System transfer functions were measured on a 3.0T ultra-high gradient NeuroFrontier MRI system to characterize hardware responses. The optimization algorithm reshaped OGSE waveforms by redistributing spectral energy away from resonant frequencies while maintaining multiple constraints including hardware limits and desired diffusion encoding spectrum. The optimization was validated by comparing non-optimized and optimized waveforms through physical measurements, phantom scans, and in vivo human brain scans. RESULTS: Reductions of up to 10 dBA in predicted sound pressure level and 3.4 dB in vibration power spectral intensity were achieved. In phantom experiments, vibration-induced artifacts were significantly mitigated, resulting in superior repeatability and lower coefficients of variation in ADC maps, approaching the minimum-vibration reference by pulsed gradient sequences. In vivo validation demonstrated enhanced acquisition stability and consistent frequency-dependent ADC trends from 0 to 120 Hz. CONCLUSION: The AVOID framework effectively suppressed intense acoustic noise and mechanical vibration induced by strong oscillating gradients on ultra-high gradient scanners, ensuring a stable and comfortable environment for advanced microstructural imaging.

Authors

Institutions

Publication Details

Journal
Magnetic Resonance in Medicine
Published
2026-08-25
DOI
https://doi.org/10.1002/mrm.70580
Primary Topic
Advanced Neuroimaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimizing Oscillating Diffusion Gradients to Reduce Mechanical Vibration and Acoustic Noise on Ultra‐High Gradient Scanners

Xingzhou Chen, Dan Wu, Liyi Kang, Haotian Li et al.
Magnetic Resonance in Medicine
Advanced Neuroimaging Techniques and Applications
article

Optimizing Oscillating Diffusion Gradients to Reduce Mechanical Vibration and Acoustic Noise on Ultra‐High Gradient Scanners

Xingzhou Chen, Dan Wu, Liyi Kang, Haotian Li, Chengbo Wang, Jianmin Yuan, Jiayu Zhu, Zhuming Shuai
article en

Abstract

PURPOSE: The deployment of oscillating gradient spin echo (OGSE) diffusion encoding sequences on ultra-high gradient scanners enables advanced microstructure imaging but induces intense acoustic noise and mechanical vibrations, causing subject discomfort and compromising imaging stability and quantitative accuracy. This study implements Acoustic noise and Vibration Optimized dIffusion gradient Design (AVOID) to mitigate these side effects, facilitating the high-performance OGSE applications in demanding hardware environments. METHODS: System transfer functions were measured on a 3.0T ultra-high gradient NeuroFrontier MRI system to characterize hardware responses. The optimization algorithm reshaped OGSE waveforms by redistributing spectral energy away from resonant frequencies while maintaining multiple constraints including hardware limits and desired diffusion encoding spectrum. The optimization was validated by comparing non-optimized and optimized waveforms through physical measurements, phantom scans, and in vivo human brain scans. RESULTS: Reductions of up to 10 dBA in predicted sound pressure level and 3.4 dB in vibration power spectral intensity were achieved. In phantom experiments, vibration-induced artifacts were significantly mitigated, resulting in superior repeatability and lower coefficients of variation in ADC maps, approaching the minimum-vibration reference by pulsed gradient sequences. In vivo validation demonstrated enhanced acquisition stability and consistent frequency-dependent ADC trends from 0 to 120 Hz. CONCLUSION: The AVOID framework effectively suppressed intense acoustic noise and mechanical vibration induced by strong oscillating gradients on ultra-high gradient scanners, ensuring a stable and comfortable environment for advanced microstructural imaging.

Magnetic Resonance in Medicine
University of Nottingham Ningbo China (CN), Children's Hospital of Zhejiang University (CN), United Imaging Healthcare (China) (CN), Zhejiang University (CN)
Openalex Percentile: Top 10%
Advanced Neuroimaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.