Evaluation of AI‐based segmentation and synthetic CT generation for MRI‐only prostate radiotherapy planning

BACKGROUND: There is an increasing use of magnetic resonance imaging (MRI) for segmentation of target volumes and organs at risk due to superior soft tissue contrast compared to computed tomography (CT). Conventional workflows require CT-MRI registration, introducing geometric uncertainty that can propagate to dose delivery. MRI-only radiotherapy, based on AI-generated synthetic CT (sCT) and segmentations from T2-weighted images, offers a single-modality approach, potentially eliminating registration errors. PURPOSE: Our study aims to evaluate whether an MRI-only workflow, using commercial AI models for segmentation and sCT generation (MVision), demonstrates performance comparable to expert-defined segmentations and reference CT-based calculations within the studied dataset. METHODS: We analyzed five segmentations on 19 male pelvis patients from the Gold Atlas dataset, providing deformably co-registered CT and MRI, sCTs, with multi-observer and consensus segmentations. We investigated segmentation accuracy, image similarity, and evaluated the dosimetric impact by comparing AI- to manual segmentations and CTs to sCTs. Wilcoxon signed-rank tests were used for statistical significance and leave-one-out kappa determined segmentation outliers. RESULTS: Results show that AI-segmentations achieved accuracy comparable to interobserver variability across all organs. Median Dice coefficients exceeded 0.9 for the prostate and bladder, and distance metrics closely matched the manual results. The leave-one-out kappa showed that AI-segmentations were indistinguishable from manual observers for all organs. sCTs showed high similarity to the reference CTs, with a median mean absolute error of 34.0 HU and median structural similarity index of 0.93. Dose recalculations showed excellent agreement; median 3D-gamma pass rate was 99.8% (2%/2 mm), and mean absolute dose difference was 0.036 Gy. All evaluated clinical dose constraints remained fulfilled after recalculation on the sCT. CONCLUSIONS: Our results confirm that an AI-segmentation and MRI-sCT workflow provides geometric, image, and dose accuracy comparable to manual- and CT-based references, supporting their integration into clinical workflows for pelvic radiotherapy.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-08-25
DOI
https://doi.org/10.1002/acm2.70765
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of AI‐based segmentation and synthetic CT generation for MRI‐only prostate radiotherapy planning

Joakim Jonsson, Amanda Östensson
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
article

Evaluation of AI‐based segmentation and synthetic CT generation for MRI‐only prostate radiotherapy planning

Joakim Jonsson, Amanda Östensson
article en

Abstract

BACKGROUND: There is an increasing use of magnetic resonance imaging (MRI) for segmentation of target volumes and organs at risk due to superior soft tissue contrast compared to computed tomography (CT). Conventional workflows require CT-MRI registration, introducing geometric uncertainty that can propagate to dose delivery. MRI-only radiotherapy, based on AI-generated synthetic CT (sCT) and segmentations from T2-weighted images, offers a single-modality approach, potentially eliminating registration errors. PURPOSE: Our study aims to evaluate whether an MRI-only workflow, using commercial AI models for segmentation and sCT generation (MVision), demonstrates performance comparable to expert-defined segmentations and reference CT-based calculations within the studied dataset. METHODS: We analyzed five segmentations on 19 male pelvis patients from the Gold Atlas dataset, providing deformably co-registered CT and MRI, sCTs, with multi-observer and consensus segmentations. We investigated segmentation accuracy, image similarity, and evaluated the dosimetric impact by comparing AI- to manual segmentations and CTs to sCTs. Wilcoxon signed-rank tests were used for statistical significance and leave-one-out kappa determined segmentation outliers. RESULTS: Results show that AI-segmentations achieved accuracy comparable to interobserver variability across all organs. Median Dice coefficients exceeded 0.9 for the prostate and bladder, and distance metrics closely matched the manual results. The leave-one-out kappa showed that AI-segmentations were indistinguishable from manual observers for all organs. sCTs showed high similarity to the reference CTs, with a median mean absolute error of 34.0 HU and median structural similarity index of 0.93. Dose recalculations showed excellent agreement; median 3D-gamma pass rate was 99.8% (2%/2 mm), and mean absolute dose difference was 0.036 Gy. All evaluated clinical dose constraints remained fulfilled after recalculation on the sCT. CONCLUSIONS: Our results confirm that an AI-segmentation and MRI-sCT workflow provides geometric, image, and dose accuracy comparable to manual- and CT-based references, supporting their integration into clinical workflows for pelvic radiotherapy.

Journal of Applied Clinical Medical PhysicsVol. 27(9)
Umeå University (SE)
Cancerforskningsfonden i Norrland
Openalex Percentile: Top 11%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.