Development and validation of an automated MTF measurement method for five types of Catphan phantoms using bar pattern objects

PURPOSE: To develop an automated method for measuring spatial resolution using the bar pattern objects in Catphan 500, 503, 504, 600, and 604 phantoms. Method: After identifying the phantom's center coordinates and anchor, we used template matching to estimate the position of each bar pattern object. Subsequently, a circular mask was used to segment the objects locally, yielding the accurate center point of each object. A region of interest (ROI) was then drawn on each object, taking into account its position and angle relative to the phantom center. By plotting the modulation transfer function (MTF) derived from the variance method against the spatial frequency of each object, an MTF curve was constructed. To test the robustness of the proposed method, we evaluated the automated algorithm across phantom model variations, geometric modifications, and reconstruction filters. The measurement results were compared with the visual assessments of 10 human observers. Results: The proposed method successfully identified targets and placed ROIs with 100% accuracy across all phantom models and extreme geometric variations. Statistical validation revealed that the automated 10% MTF limits possess a strong and highly significant positive correlation with the visual detection thresholds of human observers. This robust agreement was consistently maintained across all evaluated phantom models (r = 0.96, p = 0.01) as well as the varying reconstruction filters (r = 0.97, p = 0.03). Under geometric modifications (rotations and flips), the method exhibited excellent measurement stability with a coefficient of variation (CoV) of 4.05%. Conclusion: The proposed automated method accurately measures the MTF of bar pattern objects across Catphan models 500, 503, 504, 600, and 604 with 100% target identification accuracy. It demonstrates geometric robustness against image misalignment, shows high sensitivity in quantifying minor resolution degradation, and correlates strongly with human visual thresholds (r ≥ 0.96). .

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-08-25
DOI
https://doi.org/10.1088/2057-1976/ae9e3a
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and validation of an automated MTF measurement method for five types of Catphan phantoms using bar pattern objects

Evi Setiawati, Ariij Naufal, Choirul Anam, Noor Diyana Osman et al.
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

Development and validation of an automated MTF measurement method for five types of Catphan phantoms using bar pattern objects

Evi Setiawati, Ariij Naufal, Choirul Anam, Noor Diyana Osman, Geoff Dougherty
article en

Abstract

PURPOSE: To develop an automated method for measuring spatial resolution using the bar pattern objects in Catphan 500, 503, 504, 600, and 604 phantoms. Method: After identifying the phantom's center coordinates and anchor, we used template matching to estimate the position of each bar pattern object. Subsequently, a circular mask was used to segment the objects locally, yielding the accurate center point of each object. A region of interest (ROI) was then drawn on each object, taking into account its position and angle relative to the phantom center. By plotting the modulation transfer function (MTF) derived from the variance method against the spatial frequency of each object, an MTF curve was constructed. To test the robustness of the proposed method, we evaluated the automated algorithm across phantom model variations, geometric modifications, and reconstruction filters. The measurement results were compared with the visual assessments of 10 human observers. Results: The proposed method successfully identified targets and placed ROIs with 100% accuracy across all phantom models and extreme geometric variations. Statistical validation revealed that the automated 10% MTF limits possess a strong and highly significant positive correlation with the visual detection thresholds of human observers. This robust agreement was consistently maintained across all evaluated phantom models (r = 0.96, p = 0.01) as well as the varying reconstruction filters (r = 0.97, p = 0.03). Under geometric modifications (rotations and flips), the method exhibited excellent measurement stability with a coefficient of variation (CoV) of 4.05%. Conclusion: The proposed automated method accurately measures the MTF of bar pattern objects across Catphan models 500, 503, 504, 600, and 604 with 100% target identification accuracy. It demonstrates geometric robustness against image misalignment, shows high sensitivity in quantifying minor resolution degradation, and correlates strongly with human visual thresholds (r ≥ 0.96). .

Biomedical Physics & Engineering Express
Diponegoro University (ID), California State University, Channel Islands (US), Hospital Universiti Sains Malaysia (MY)
Openalex Percentile: Top 11%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.