Log file‐based patient‐specific quality assurance for MR‐guided radiation therapy with comprehensive motion management

BACKGROUND: Patient-Specific Quality Assurance (PSQA) for Elekta Unity MR-Linac treatments delivered under Comprehensive Motion Management (CMM) is challenging because beam gating, unexpected interruptions, and split sub-segments introduce delivery complexities that are not fully addressed by conventional quality assurance workflows. PURPOSE: This study aimed to develop and evaluate a log file-based PSQA program for Elekta Unity MR-Linac treatments delivered under CMM. METHODS: Log files were parsed to extract Multileaf Collimator (MLC) positions, jaw positions, and Monitor Units (MU) for each segment, with dedicated handling of CMM-induced beam gating, unexpected beam interruptions, and split sub-segments. Gantry-resolved fluence maps reconstructed from the log files were compared with planned fluence maps using gamma analysis. The system was evaluated for 66 cases, including 3 completion plans and 2 baseline shift plans. RESULTS: Among the 66 plans, 64 achieved passing gamma results, with a pooled average passing rate of 99.91 ± 0.21%. CONCLUSIONS: The proposed log file-based PSQA framework demonstrates accuracy and clinical practicality for MR-Linac treatments in CMM mode. It may serve as a reliable post-treatment verification tool and reduce dependence on conventional phantom-based QA.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-08-25
DOI
https://doi.org/10.1002/acm2.70773
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Log file‐based patient‐specific quality assurance for MR‐guided radiation therapy with comprehensive motion management

Kai Yuan, Matthew M. Cheung, Louis Lee
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
article

Log file‐based patient‐specific quality assurance for MR‐guided radiation therapy with comprehensive motion management

Kai Yuan, Matthew M. Cheung, Louis Lee
article en

Abstract

BACKGROUND: Patient-Specific Quality Assurance (PSQA) for Elekta Unity MR-Linac treatments delivered under Comprehensive Motion Management (CMM) is challenging because beam gating, unexpected interruptions, and split sub-segments introduce delivery complexities that are not fully addressed by conventional quality assurance workflows. PURPOSE: This study aimed to develop and evaluate a log file-based PSQA program for Elekta Unity MR-Linac treatments delivered under CMM. METHODS: Log files were parsed to extract Multileaf Collimator (MLC) positions, jaw positions, and Monitor Units (MU) for each segment, with dedicated handling of CMM-induced beam gating, unexpected beam interruptions, and split sub-segments. Gantry-resolved fluence maps reconstructed from the log files were compared with planned fluence maps using gamma analysis. The system was evaluated for 66 cases, including 3 completion plans and 2 baseline shift plans. RESULTS: Among the 66 plans, 64 achieved passing gamma results, with a pooled average passing rate of 99.91 ± 0.21%. CONCLUSIONS: The proposed log file-based PSQA framework demonstrates accuracy and clinical practicality for MR-Linac treatments in CMM mode. It may serve as a reliable post-treatment verification tool and reduce dependence on conventional phantom-based QA.

Journal of Applied Clinical Medical PhysicsVol. 27(9)
Chinese University of Hong Kong (HK), Caritas Medical Centre (CN), University of Hong Kong (HK)
Openalex Percentile: Top 11%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.