Log file‐based patient‐specific quality assurance for MR‐guided radiation therapy with comprehensive motion management
BACKGROUND: Patient-Specific Quality Assurance (PSQA) for Elekta Unity MR-Linac treatments delivered under Comprehensive Motion Management (CMM) is challenging because beam gating, unexpected interruptions, and split sub-segments introduce delivery complexities that are not fully addressed by conventional quality assurance workflows. PURPOSE: This study aimed to develop and evaluate a log file-based PSQA program for Elekta Unity MR-Linac treatments delivered under CMM. METHODS: Log files were parsed to extract Multileaf Collimator (MLC) positions, jaw positions, and Monitor Units (MU) for each segment, with dedicated handling of CMM-induced beam gating, unexpected beam interruptions, and split sub-segments. Gantry-resolved fluence maps reconstructed from the log files were compared with planned fluence maps using gamma analysis. The system was evaluated for 66 cases, including 3 completion plans and 2 baseline shift plans. RESULTS: Among the 66 plans, 64 achieved passing gamma results, with a pooled average passing rate of 99.91 ± 0.21%. CONCLUSIONS: The proposed log file-based PSQA framework demonstrates accuracy and clinical practicality for MR-Linac treatments in CMM mode. It may serve as a reliable post-treatment verification tool and reduce dependence on conventional phantom-based QA.
Authors
- Kai Yuan (ORCID: https://orcid.org/0000-0001-6105-4772)
- Matthew M. Cheung
- Louis Lee
Institutions
- Chinese University of Hong Kong (HK)
- Caritas Medical Centre (CN)
- University of Hong Kong (HK)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2026-08-25
- DOI
- https://doi.org/10.1002/acm2.70773
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00