Dose-dependent refractive index changes in megavoltage X-ray-irradiated high-density polyethylene measured by terahertz time-domain spectroscopy

This note investigates terahertz time-domain spectroscopy (THz-TDS) as a non-destructive technique for detecting radiation-induced changes in high-density polyethylene (HDPE). HDPE samples were irradiated with megavoltage X-rays at doses from 100 Gy to 800 Gy using a medical linear accelerator at a dose rate of 6.00 Gy/min. The refractive index, measured with THz-TDS in the range 0.3-2.5 THz, showed an overall positive linear association with absorbed dose (R2=0.83), corresponding to a sensitivity of 7×10⁻⁶ Gy⁻¹ with a 95 % confidence interval of 4.10-9.84 × 10⁻⁶ Gy⁻¹. The absorption coefficient shows no systematic dependence on dose within the measurement uncertainty. These results provide preliminary evidence that THz-TDS can detect dose-dependent changes in the refractive index of HDPE and may provide a basis for further investigation of polymer-based dosimetry.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-08-25
DOI
https://doi.org/10.1088/2057-1976/ae9e38
Primary Topic
Terahertz technology and applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dose-dependent refractive index changes in megavoltage X-ray-irradiated high-density polyethylene measured by terahertz time-domain spectroscopy

Mads Brincker, Katrine Gerassimovitch Eskildsen, Esben Pilgaard Skovsen
Biomedical Physics & Engineering Express
Terahertz technology and applications
article

Dose-dependent refractive index changes in megavoltage X-ray-irradiated high-density polyethylene measured by terahertz time-domain spectroscopy

Mads Brincker, Katrine Gerassimovitch Eskildsen, Esben Pilgaard Skovsen
article en

Abstract

This note investigates terahertz time-domain spectroscopy (THz-TDS) as a non-destructive technique for detecting radiation-induced changes in high-density polyethylene (HDPE). HDPE samples were irradiated with megavoltage X-rays at doses from 100 Gy to 800 Gy using a medical linear accelerator at a dose rate of 6.00 Gy/min. The refractive index, measured with THz-TDS in the range 0.3-2.5 THz, showed an overall positive linear association with absorbed dose (R2=0.83), corresponding to a sensitivity of 7×10⁻⁶ Gy⁻¹ with a 95 % confidence interval of 4.10-9.84 × 10⁻⁶ Gy⁻¹. The absorption coefficient shows no systematic dependence on dose within the measurement uncertainty. These results provide preliminary evidence that THz-TDS can detect dose-dependent changes in the refractive index of HDPE and may provide a basis for further investigation of polymer-based dosimetry.

Biomedical Physics & Engineering Express
Aalborg University Hospital (DK), Aalborg University (DK)
Aalborg Universitet
Affordable and clean energy
Openalex Percentile: Top 19%
Terahertz technology and applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.