Long-term Incidence of Revision Surgery Caused by Electrode Migration in a Large Cochlear Implant Cohort Using a Titanium Fixation Clip Applied by a Novel Reverse Application Technique

OBJECTIVE: To determine the effect on the frequency of revision surgery caused by electrode migration (EM) in cochlear implantation by use of an electrode fixation clip combined with a novel surgical technique. STUDY DESIGN: Retrospective chart review. SETTING: Tertiary referral center. PATIENTS: The investigated institutional cohort included 1052 cases of cochlear implantation with straight electrode arrays from a single manufacturer between 2011 and 2023. Until 2016, the fixation clip was not used, whereas since 2017, the fixation clip and the described novel surgical technique have been applied as an institutional standard in nearly all cases. INTERVENTION: The titanium electrode fixation clip was first introduced and described in 1998 and was CE marked in Europe in 2014. In modification of the original description, a "Reverse Fixation Clip Technique" was applied to optimize the surgical handling and fixation stability. A similar approach has been described recently. MAIN OUTCOME MEASURE: Cumulative incidence of revision surgery caused by EM. RESULTS: The cumulative incidence of revision surgery caused by EM at 3, 5, and 7 years was 0.5%, 1.0%, and 1.9% in the group without the fixation clip (n=461; 43.4%) and 0.0% at all time points in the group with the fixation clip using the "Reverse Fixation Clip Technique" (n=591; 56.6%) (log-rank test P=0.054). No fixation clip-related complications were noted. CONCLUSION: The application of the fixation clip using the "Reverse Fixation Clip Technique" prevents revision surgery caused by EM in cochlear implantation.

Authors

Institutions

Publication Details

Journal
Otology & Neurotology
Published
2026-08-24
DOI
https://doi.org/10.1097/mao.0000000000005005
Primary Topic
Hearing Loss and Rehabilitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Long-term Incidence of Revision Surgery Caused by Electrode Migration in a Large Cochlear Implant Cohort Using a Titanium Fixation Clip Applied by a Novel Reverse Application Technique

Thore Schade‐Mann, Steffen Paasch, Martin Holderried, Philipp Gamerdinger et al.
Otology & Neurotology
Hearing Loss and Rehabilitation
article

Long-term Incidence of Revision Surgery Caused by Electrode Migration in a Large Cochlear Implant Cohort Using a Titanium Fixation Clip Applied by a Novel Reverse Application Technique

Thore Schade‐Mann, Steffen Paasch, Martin Holderried, Philipp Gamerdinger, Hubert Löwenheim, G. Schweizer, Agnes Mühle, Anke Tropitzsch, Lina Maria Serna-Higuita
article en

Abstract

OBJECTIVE: To determine the effect on the frequency of revision surgery caused by electrode migration (EM) in cochlear implantation by use of an electrode fixation clip combined with a novel surgical technique. STUDY DESIGN: Retrospective chart review. SETTING: Tertiary referral center. PATIENTS: The investigated institutional cohort included 1052 cases of cochlear implantation with straight electrode arrays from a single manufacturer between 2011 and 2023. Until 2016, the fixation clip was not used, whereas since 2017, the fixation clip and the described novel surgical technique have been applied as an institutional standard in nearly all cases. INTERVENTION: The titanium electrode fixation clip was first introduced and described in 1998 and was CE marked in Europe in 2014. In modification of the original description, a "Reverse Fixation Clip Technique" was applied to optimize the surgical handling and fixation stability. A similar approach has been described recently. MAIN OUTCOME MEASURE: Cumulative incidence of revision surgery caused by EM. RESULTS: The cumulative incidence of revision surgery caused by EM at 3, 5, and 7 years was 0.5%, 1.0%, and 1.9% in the group without the fixation clip (n=461; 43.4%) and 0.0% at all time points in the group with the fixation clip using the "Reverse Fixation Clip Technique" (n=591; 56.6%) (log-rank test P=0.054). No fixation clip-related complications were noted. CONCLUSION: The application of the fixation clip using the "Reverse Fixation Clip Technique" prevents revision surgery caused by EM in cochlear implantation.

Otology & Neurotology
Boston Children's Hospital (US), Harvard University (US), Zimmer Biomet (Netherlands) (NL), Bernstein Center for Computational Neuroscience Tübingen (DE), Boston Children's Museum (US), Dana-Farber/Boston Children's Cancer and Blood Disorders Center (US), University of Tübingen (DE)
Reduced inequalities
Openalex Percentile: Top 8%
Hearing Loss and Rehabilitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.