Multicenter Evaluation of Noise Levels and Staff Perceptions Across Laser Technologies During Holmium Laser Enucleation of the Prostate: A Prostate Enucleation Collaborative Initiative

INTRODUCTION: Operating room (OR) noise during holmium laser enucleation of the prostate (HoLEP), particularly from laser activation, may impair surgical workflow and staff performance. METHODS: 84 HoLEP procedures were recorded at three academic institutions between June 2024 and January 2026. Laser platforms included Lumenis Moses 2.0, Quanta Litho 150, Quanta Magneto, and Dornier Thulio. Sound measurements were obtained using a standardized smartphone-based system during three phases: laser-off (baseline), idle (not firing), and active lasing. Acoustic metrics included average sound level (LAeq), peak sound level (Lcpeak), and sound levels exceeded for 10%, 50%, and 90% of the recording time (L10, L50, and L90). Background noise correction was performed using baseline measurements to account for ambient noise and isolate laser-related sound. OR staff completed a validated survey assessing perceived noise burden (0-10 scale). RESULTS: Background-corrected noise during active lasing differed significantly by platform. Litho demonstrated lower mean noise levels compared to Thulio for LAeq (60.6 vs. 68.2 dB, p = 0.023) and L10 (61.5 vs. 70.6 dB, p = 0.023), while Moses had higher LAeq compared to Litho (63.1 vs. 60.6 dB, p = 0.043). Overall, Litho had the lowest noise levels, whereas Thulio had the highest. Survey responses paralleled these findings. Litho was associated with lower perceived overall effort (2.1 ± 0.9), while Moses demonstrated higher overall effort (3.6 ± 1.2) for communication and concentration (Figure 2). Significant pairwise differences were observed between Moses and Litho (all p < 0.05). Trends were consistent across sites. CONCLUSIONS: Noise levels during HoLEP varied by laser technology and correlated with differences in perceived communication and cognitive effort. Litho demonstrated the lowest acoustic burden, while Moses and Thulio had higher noise exposure. These findings highlight the impact of laser selection on the OR environment and intraoperative team performance.

Authors

Institutions

Publication Details

Journal
Lasers in Surgery and Medicine
Published
2026-08-24
DOI
https://doi.org/10.1002/lsm.70197
Primary Topic
Intraocular Surgery and Lenses
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Multicenter Evaluation of Noise Levels and Staff Perceptions Across Laser Technologies During Holmium Laser Enucleation of the Prostate: A Prostate Enucleation Collaborative Initiative

Bryce Baird, Karen Stern, Smita De, Amanda E. Kahn et al.
Lasers in Surgery and Medicine
Intraocular Surgery and Lenses
article

Multicenter Evaluation of Noise Levels and Staff Perceptions Across Laser Technologies During Holmium Laser Enucleation of the Prostate: A Prostate Enucleation Collaborative Initiative

Bryce Baird, Karen Stern, Smita De, Amanda E. Kahn, Adam B. Cohen, Charles Klose, Scott Quarrier, Rajat Jain, Chandler Dora, Eric Qualkenbush, Benjamin Dalkin, Jessica Wu, Satvik Kuravi, Dane Klett
article en

Abstract

INTRODUCTION: Operating room (OR) noise during holmium laser enucleation of the prostate (HoLEP), particularly from laser activation, may impair surgical workflow and staff performance. METHODS: 84 HoLEP procedures were recorded at three academic institutions between June 2024 and January 2026. Laser platforms included Lumenis Moses 2.0, Quanta Litho 150, Quanta Magneto, and Dornier Thulio. Sound measurements were obtained using a standardized smartphone-based system during three phases: laser-off (baseline), idle (not firing), and active lasing. Acoustic metrics included average sound level (LAeq), peak sound level (Lcpeak), and sound levels exceeded for 10%, 50%, and 90% of the recording time (L10, L50, and L90). Background noise correction was performed using baseline measurements to account for ambient noise and isolate laser-related sound. OR staff completed a validated survey assessing perceived noise burden (0-10 scale). RESULTS: Background-corrected noise during active lasing differed significantly by platform. Litho demonstrated lower mean noise levels compared to Thulio for LAeq (60.6 vs. 68.2 dB, p = 0.023) and L10 (61.5 vs. 70.6 dB, p = 0.023), while Moses had higher LAeq compared to Litho (63.1 vs. 60.6 dB, p = 0.043). Overall, Litho had the lowest noise levels, whereas Thulio had the highest. Survey responses paralleled these findings. Litho was associated with lower perceived overall effort (2.1 ± 0.9), while Moses demonstrated higher overall effort (3.6 ± 1.2) for communication and concentration (Figure 2). Significant pairwise differences were observed between Moses and Litho (all p < 0.05). Trends were consistent across sites. CONCLUSIONS: Noise levels during HoLEP varied by laser technology and correlated with differences in perceived communication and cognitive effort. Litho demonstrated the lowest acoustic burden, while Moses and Thulio had higher noise exposure. These findings highlight the impact of laser selection on the OR environment and intraoperative team performance.

Lasers in Surgery and Medicine
Cleveland Clinic (US), Jacksonville College (US), University of Rochester Medical Center (US), Mayo Clinic in Arizona (US), Cambridge Collaborative (United States) (US), Mayo Clinic in Florida (US), University of Rochester (US)
Openalex Percentile: Top 7%
Intraocular Surgery and Lenses
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.