Dosimetric and LETd comparison of horizontal port and gantry carbon-ion radiotherapy for early-stage non–small cell lung cancer: an exploratory study

Purpose: We aimed to evaluate the feasibility of using a horizontal port (HP) system for carbon ion radiotherapy (CIRT) in non-small cell lung cancer (NSCLC) by comparing the dosimetric characteristics and dose-averaged linear energy transfer (LETd) with those of a gantry-based plan. Materials and Methods: Five patients with NSCLC treated with CIRT were included in this study. Both gantry-based and HP plans were generated using a single-field uniform dose optimization in RayStation 2024A with a pencil beam algorithm. Gantry plan beam angles followed institutional clinical protocols, whereas HP plans used beam angles of 90° or 270° with a couch roll of ±15° and yaw of ±30°. Plan evaluation was conducted using dose-volume histograms, LETd-volume histograms, and beam range analysis. Results: Dose distributions for the clinical target volume (CTV) and ipsilateral lung were similar between plans. The CTV V95% was 100% in both plans. For the ipsilateral lung, mean V5 Gy (relative biological effectiveness [RBE]) and V20 Gy (RBE) were slightly lower in the gantry plan (12.40% and 5.77%, respectively, compared to 13.54% and 6.15%, respectively). Mean LETd98% within the CTV was higher in the gantry plan (65.27 keV/μm compared to 57.07 keV/μm). Beam energy and beam range were consistently lower and shorter, respectively, in the gantry plan. Conclusion: The preliminary dosimetric performance of the HP plan was generally comparable to that of the gantry plan. However, these findings should be interpreted with caution. The gantry system may provide greater reliability and clinical efficiency owing to the simpler workflow and reduced sensitivity to setup variability and organ motion.

Authors

Institutions

Publication Details

Journal
Radiation Oncology Journal
Published
2026-08-24
DOI
https://doi.org/10.3857/roj.2026.00080
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dosimetric and LETd comparison of horizontal port and gantry carbon-ion radiotherapy for early-stage non–small cell lung cancer: an exploratory study

Seung-Hyeop Baek, Seyjoon Park, Justin C. Park, Keith M. Furutani et al.
Radiation Oncology Journal
Radiation Therapy and Dosimetry
article

Dosimetric and LETd comparison of horizontal port and gantry carbon-ion radiotherapy for early-stage non–small cell lung cancer: an exploratory study

Seung-Hyeop Baek, Seyjoon Park, Justin C. Park, Keith M. Furutani, Kyung Hwan Kim, Chris J Beltran, Jin Sung Kim, Changhwan Kim, Min Cheol Han, Chae-Seon Hong
article en

Abstract

Purpose: We aimed to evaluate the feasibility of using a horizontal port (HP) system for carbon ion radiotherapy (CIRT) in non-small cell lung cancer (NSCLC) by comparing the dosimetric characteristics and dose-averaged linear energy transfer (LETd) with those of a gantry-based plan. Materials and Methods: Five patients with NSCLC treated with CIRT were included in this study. Both gantry-based and HP plans were generated using a single-field uniform dose optimization in RayStation 2024A with a pencil beam algorithm. Gantry plan beam angles followed institutional clinical protocols, whereas HP plans used beam angles of 90° or 270° with a couch roll of ±15° and yaw of ±30°. Plan evaluation was conducted using dose-volume histograms, LETd-volume histograms, and beam range analysis. Results: Dose distributions for the clinical target volume (CTV) and ipsilateral lung were similar between plans. The CTV V95% was 100% in both plans. For the ipsilateral lung, mean V5 Gy (relative biological effectiveness [RBE]) and V20 Gy (RBE) were slightly lower in the gantry plan (12.40% and 5.77%, respectively, compared to 13.54% and 6.15%, respectively). Mean LETd98% within the CTV was higher in the gantry plan (65.27 keV/μm compared to 57.07 keV/μm). Beam energy and beam range were consistently lower and shorter, respectively, in the gantry plan. Conclusion: The preliminary dosimetric performance of the HP plan was generally comparable to that of the gantry plan. However, these findings should be interpreted with caution. The gantry system may provide greater reliability and clinical efficiency owing to the simpler workflow and reduced sensitivity to setup variability and organ motion.

Radiation Oncology Journal
Yonsei University (KR), WinnMed (US), Mayo Clinic in Florida (US)
Affordable and clean energy
Openalex Percentile: Top 10%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.