Development and Finite Element Analysis of a Titanium Bone Plate with a Localized Porous Structure for Osteosynthesis of the Radial Shaft

Traditional bone plates provide reliable fracture fixation; however, their high stiffness can cause stress shielding, thereby reducing mechanical stimulation of the bone tissue and slowing its regeneration. The biomechanical performance of a titanium plate with a localized porous structure for osteosynthesis of the radial shaft was investigated in this study. Three designs were considered: a solid plate and two plates with localized porous regions measuring 10 × 10 mm and 10 × 15 mm. The finite element analysis of the bone–plate system was performed using MSC Patran/Nastran, with rigid fixation of the proximal end of the bone and sequential application of an axial compressive load of 100 N, bending, and torsion with a moment of 1 N·m. Biomechanical performance was evaluated based on von Mises equivalent stress, fragment displacement (FD), interfragmentary movement (IFM), interfragmentary strain (IFS), and strain energy density (SED). To confirm the manufacturability of the design, the plate was fabricated from Ti–6Al–4V alloy using laser powder bed fusion (LPBF), and the geometry of the porous structure was verified by scanning electron microscopy. The results showed that the localized porous structure altered the load distribution between the plate and the bone, resulting in an increase in local stresses in the bone under the investigated loading conditions. These changes indicate an alteration in the mechanical environment within the bone, which may potentially affect conditions related to fracture healing. Among the investigated configurations, the plate with a 10 × 10 mm porous insert demonstrated the most balanced mechanical characteristics in terms of stresses in the implant, stress distribution in the bone, and structural stability.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2026-08-24
DOI
https://doi.org/10.3390/app16178405
Primary Topic
Cellular and Composite Structures
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and Finite Element Analysis of a Titanium Bone Plate with a Localized Porous Structure for Osteosynthesis of the Radial Shaft

Мадина Исаметова, Aaron Vance, Arun Arjunan, Fuad Khoshnaw et al.
Applied Sciences
Cellular and Composite Structures
article

Development and Finite Element Analysis of a Titanium Bone Plate with a Localized Porous Structure for Osteosynthesis of the Radial Shaft

Мадина Исаметова, Aaron Vance, Arun Arjunan, Fuad Khoshnaw, Denis Tkachenko, Yeszhan Ilyassov, Yersin Zhunussov
article en

Abstract

Traditional bone plates provide reliable fracture fixation; however, their high stiffness can cause stress shielding, thereby reducing mechanical stimulation of the bone tissue and slowing its regeneration. The biomechanical performance of a titanium plate with a localized porous structure for osteosynthesis of the radial shaft was investigated in this study. Three designs were considered: a solid plate and two plates with localized porous regions measuring 10 × 10 mm and 10 × 15 mm. The finite element analysis of the bone–plate system was performed using MSC Patran/Nastran, with rigid fixation of the proximal end of the bone and sequential application of an axial compressive load of 100 N, bending, and torsion with a moment of 1 N·m. Biomechanical performance was evaluated based on von Mises equivalent stress, fragment displacement (FD), interfragmentary movement (IFM), interfragmentary strain (IFS), and strain energy density (SED). To confirm the manufacturability of the design, the plate was fabricated from Ti–6Al–4V alloy using laser powder bed fusion (LPBF), and the geometry of the porous structure was verified by scanning electron microscopy. The results showed that the localized porous structure altered the load distribution between the plate and the bone, resulting in an increase in local stresses in the bone under the investigated loading conditions. These changes indicate an alteration in the mechanical environment within the bone, which may potentially affect conditions related to fracture healing. Among the investigated configurations, the plate with a 10 × 10 mm porous insert demonstrated the most balanced mechanical characteristics in terms of stresses in the implant, stress distribution in the bone, and structural stability.

Applied SciencesVol. 16(17)
University of Wolverhampton (GB), Satbayev University (KZ), International Information Technologies University (KZ), De Montfort University (GB)
Affordable and clean energy
Openalex Percentile: Top 18%
Cellular and Composite Structures
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.