Real-World safety, patient-reported outcomes, and glycemic metrics with a 7-Day extended infusion set in adults with Type 1 diabetes using automated insulin delivery systems

BACKGROUND: Extended-wear infusion sets (E-iS) allowing up to 7 days of catheter use have recently become available. Real-world evidence of contemporary automated insulin delivery (AID) systems using E-iS remains limited. METHOD: A prospective non-randomized single-arm real-world study in adults with type 1 diabetes using the MiniMed-780G AID system who transitioned from conventional 3-day infusion sets to E-iS was conducted. Clinical variables, device settings, adverse events, and patient-reported outcomes were assessed at baseline, 3 and 6 months. RESULTS: 65 participants were included, 58 completed 6 months of follow-up.Mean catheter wear time increased from 3.6 ± 1.1 to 6.1 ± 2.3 days,(p < 0.001), and reservoir change interval increased from 3.5 ± 1.0 to 5.2 ± 2.3 days,(p < 0.001). Glycemic control remained overall stable, although small changes were observed in some CGM metrics (Time in range-TIR: 80.0 ± 7.1% to 78.0 ± 10.0%, p = 0.014) and slight increases in time above range, while hypoglycemia exposure remained unchanged. HbA1c decreased from 6.8 ± 0.5% to 6.4 ± 1.0%, (p = 0.03). Total daily insulin dose did not change, but insulin use related to infusion set management decreased significantly. Pain at cannula insertion improved from 2.1 ± 1.5 to 1.2 ± 1.1,(p < 0.001), and treatment satisfaction increased from 8.2 ± 1.9 to 9.0 ± 1.2,(p = 0.012). No episodes of severe hypoglycemia or diabetic ketoacidosis occurred. Cannula wear duration was strongly correlated with reservoir change interval (r = 0.712, p < 0.001). At 6 months, 92.3% of participants continued using the E-iS, with 48.3% maintaining the full 7-day wear duration. CONCLUSIONS: In real-world use with an AID system, the E-iS significantly prolonged catheter wear duration and reservoir change intervals while remaining safe and well tolerated. It was associated with high satisfaction and stable glycemic outcomes, supporting E-iS as a practical strategy to reduce the burden of infusion set changes in routine clinical practice.

Authors

Institutions

Publication Details

Journal
Endocrine
Published
2026-08-21
DOI
https://doi.org/10.1007/s12020-026-04743-z
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Real-World safety, patient-reported outcomes, and glycemic metrics with a 7-Day extended infusion set in adults with Type 1 diabetes using automated insulin delivery systems

Ana Ortolá Buigues, Sofía del Amo Simón, Marta Sánchez Ibáñez, Emilia Gómez Hoyos et al.
Endocrine
Diabetes Management and Research
article

Real-World safety, patient-reported outcomes, and glycemic metrics with a 7-Day extended infusion set in adults with Type 1 diabetes using automated insulin delivery systems

Ana Ortolá Buigues, Sofía del Amo Simón, Marta Sánchez Ibáñez, Emilia Gómez Hoyos, Gonzalo Díaz-Soto, María de la O Nieto de la Marca, Pablo Fernández Velasco, Daniel Antonio de Luis Román, Beatriz Torres, Ana Collantes Matallana
article en

Abstract

BACKGROUND: Extended-wear infusion sets (E-iS) allowing up to 7 days of catheter use have recently become available. Real-world evidence of contemporary automated insulin delivery (AID) systems using E-iS remains limited. METHOD: A prospective non-randomized single-arm real-world study in adults with type 1 diabetes using the MiniMed-780G AID system who transitioned from conventional 3-day infusion sets to E-iS was conducted. Clinical variables, device settings, adverse events, and patient-reported outcomes were assessed at baseline, 3 and 6 months. RESULTS: 65 participants were included, 58 completed 6 months of follow-up.Mean catheter wear time increased from 3.6 ± 1.1 to 6.1 ± 2.3 days,(p < 0.001), and reservoir change interval increased from 3.5 ± 1.0 to 5.2 ± 2.3 days,(p < 0.001). Glycemic control remained overall stable, although small changes were observed in some CGM metrics (Time in range-TIR: 80.0 ± 7.1% to 78.0 ± 10.0%, p = 0.014) and slight increases in time above range, while hypoglycemia exposure remained unchanged. HbA1c decreased from 6.8 ± 0.5% to 6.4 ± 1.0%, (p = 0.03). Total daily insulin dose did not change, but insulin use related to infusion set management decreased significantly. Pain at cannula insertion improved from 2.1 ± 1.5 to 1.2 ± 1.1,(p < 0.001), and treatment satisfaction increased from 8.2 ± 1.9 to 9.0 ± 1.2,(p = 0.012). No episodes of severe hypoglycemia or diabetic ketoacidosis occurred. Cannula wear duration was strongly correlated with reservoir change interval (r = 0.712, p < 0.001). At 6 months, 92.3% of participants continued using the E-iS, with 48.3% maintaining the full 7-day wear duration. CONCLUSIONS: In real-world use with an AID system, the E-iS significantly prolonged catheter wear duration and reservoir change intervals while remaining safe and well tolerated. It was associated with high satisfaction and stable glycemic outcomes, supporting E-iS as a practical strategy to reduce the burden of infusion set changes in routine clinical practice.

EndocrineVol. 91(1)
Universidad de Valladolid (ES), Instituto de Salud Carlos III (ES), Instituto de Biomedicina y Genética Molecular de Valladolid (ES), Hospital Clínico Universitario de Valladolid (ES)
Universidad de Valladolid, Junta de Castilla y León
Good health and well-being
Openalex Percentile: Top 10%
Diabetes Management and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.