Effects of pneumatic tube transport on neurofilament light chain, neuron-specific enolase, and cardiac biomarkers

Abstract Objectives Pneumatic tube systems (PTS) accelerate clinical blood sample transport but may introduce preanalytical bias through mechanical stress and hemolysis. Effects on neurofilament light chain (NfL), neuron-specific enolase (NSE), cardiac biomarkers, and hemolysis-sensitive analytes remain incompletely defined. We assessed the impact of routine PTS transport on these analytes under controlled paired transport conditions. Methods Paired blood samples from 20 healthy volunteers were transported simultaneously by PTS or manual courier. Three-axis acceleration and temperature loggers quantified transport-associated mechanical stress and temperature exposure. NfL, NSE, cardiac biomarkers, and hemolysis-sensitive analytes were measured using routine diagnostic platforms. PTS and courier results were compared by paired analyses, and percentage bias was calculated relative to courier-transported samples. Results After Benjamini–Hochberg correction, only NfL and lactate dehydrogenase (LDH) remained significantly elevated in PTS-transported samples. Mean biases were 12.01 % (95 % CI, 7.05–16.97 %) for NfL and 10.08 % (95 % CI, 7.23–12.93 %) for LDH. NSE exhibited the largest percentage bias but did not remain statistically significant after correction. Cardiac biomarkers, including troponin-I, N-terminal pro-B-type natriuretic peptide, creatine kinase (CK), CK-MB, and high-sensitivity C-reactive protein, were not significantly affected. Potassium, bilirubin, aspartate aminotransferase, and alanine aminotransferase were likewise not significantly changed. Conclusions PTS transport significantly increased NfL and LDH concentrations, whereas cardiac biomarkers remained constant under the tested conditions. NSE showed a non-significant trend toward higher concentrations after PTS transport. These findings indicate analyte-specific vulnerability to transport-associated mechanical stress and support analyte-specific PTS route validation before routine clinical implementation.

Authors

Institutions

Publication Details

Journal
Journal of Laboratory Medicine
Published
2026-08-21
DOI
https://doi.org/10.1515/labmed-2026-0112
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effects of pneumatic tube transport on neurofilament light chain, neuron-specific enolase, and cardiac biomarkers

Wen-Chan Huang, Felix Eckelt, Herag Arabian, Moritz Waldmann et al.
Journal of Laboratory Medicine
Clinical Laboratory Practices and Quality Control
article

Effects of pneumatic tube transport on neurofilament light chain, neuron-specific enolase, and cardiac biomarkers

Wen-Chan Huang, Felix Eckelt, Herag Arabian, Moritz Waldmann, David Cetaj, Thomas Renné, Rainer Erler-Zirnwald
article en

Abstract

Abstract Objectives Pneumatic tube systems (PTS) accelerate clinical blood sample transport but may introduce preanalytical bias through mechanical stress and hemolysis. Effects on neurofilament light chain (NfL), neuron-specific enolase (NSE), cardiac biomarkers, and hemolysis-sensitive analytes remain incompletely defined. We assessed the impact of routine PTS transport on these analytes under controlled paired transport conditions. Methods Paired blood samples from 20 healthy volunteers were transported simultaneously by PTS or manual courier. Three-axis acceleration and temperature loggers quantified transport-associated mechanical stress and temperature exposure. NfL, NSE, cardiac biomarkers, and hemolysis-sensitive analytes were measured using routine diagnostic platforms. PTS and courier results were compared by paired analyses, and percentage bias was calculated relative to courier-transported samples. Results After Benjamini–Hochberg correction, only NfL and lactate dehydrogenase (LDH) remained significantly elevated in PTS-transported samples. Mean biases were 12.01 % (95 % CI, 7.05–16.97 %) for NfL and 10.08 % (95 % CI, 7.23–12.93 %) for LDH. NSE exhibited the largest percentage bias but did not remain statistically significant after correction. Cardiac biomarkers, including troponin-I, N-terminal pro-B-type natriuretic peptide, creatine kinase (CK), CK-MB, and high-sensitivity C-reactive protein, were not significantly affected. Potassium, bilirubin, aspartate aminotransferase, and alanine aminotransferase were likewise not significantly changed. Conclusions PTS transport significantly increased NfL and LDH concentrations, whereas cardiac biomarkers remained constant under the tested conditions. NSE showed a non-significant trend toward higher concentrations after PTS transport. These findings indicate analyte-specific vulnerability to transport-associated mechanical stress and support analyte-specific PTS route validation before routine clinical implementation.

Journal of Laboratory Medicine
Royal College of Surgeons in Ireland (IE), Universität Hamburg (DE), Walisongo State Islamic University (ID), University Medical Center Hamburg-Eppendorf (DE)
Good health and well-being
Openalex Percentile: Top 10%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.