Privacy-Preserving Head Pose Estimation System for Measuring Cervical Range of Motion

Cervical range of motion (CROM) has been used in research and clinically for assessing cervical health. Gold-standard goniometers tend to be cumbersome. However, Inertial Measurement Units (IMUs) or vision-based alternatives demand frequent calibration and/or costly hardware; moreover, the subject is aware of being measured and there is a risk of breaching privacy. In response, we have developed a non-contact HPNet system for head pose estimation (HPE) that can use a rear-facing camera to quantify CROM accurately. A Re-parameterized Visual Geometry Group (RepVGG)-D2se model is employed as the backbone of the network, and a Spatial Feature Enhancement (SCFE) module is incorporated to improve feature extraction. HPNet was evaluated on the large-scale Carnegie Mellon University (CMU) Panoptic dataset, achieving a mean absolute error (MAE) of 3.48°, 3.22°and 3.34° for yaw, pitch and roll respectively. Inter-instrument reliability was excellent for all six cervical movements when compared with the research/clinical-grade CROM device, with intraclass correlation coefficients (ICCs) averaging 0.939. Bland–Altman plots confirmed close agreement between the two methods. Cervical movement trajectory curves further confirmed the concordance between the clinical device and our method. The system is fully automatic, requires only a rear-facing camera, effectively preserves patient privacy, and provides accurate cervical posture estimation. This technology may provide a basis for future applications in neck-disorder screening, remote health monitoring, and personalized musculoskeletal wellness management, although further task-specific clinical validation will be required. To date, HPNet has been validated primarily on a computer-based platform and has not yet been deployed on smartphones. Future work will focus on model lightweighting, mobile deployment, and cross-device adaptation to facilitate its practical implementation on mobile devices.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2026-08-21
DOI
https://doi.org/10.3390/s26165310
Primary Topic
Cervical and Thoracic Myelopathy
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Privacy-Preserving Head Pose Estimation System for Measuring Cervical Range of Motion

Zhuofu Liu, G. Li, Peter W. McCarthy, Lichao Zhang
Sensors
Cervical and Thoracic Myelopathy
article

Privacy-Preserving Head Pose Estimation System for Measuring Cervical Range of Motion

Zhuofu Liu, G. Li, Peter W. McCarthy, Lichao Zhang
article en

Abstract

Cervical range of motion (CROM) has been used in research and clinically for assessing cervical health. Gold-standard goniometers tend to be cumbersome. However, Inertial Measurement Units (IMUs) or vision-based alternatives demand frequent calibration and/or costly hardware; moreover, the subject is aware of being measured and there is a risk of breaching privacy. In response, we have developed a non-contact HPNet system for head pose estimation (HPE) that can use a rear-facing camera to quantify CROM accurately. A Re-parameterized Visual Geometry Group (RepVGG)-D2se model is employed as the backbone of the network, and a Spatial Feature Enhancement (SCFE) module is incorporated to improve feature extraction. HPNet was evaluated on the large-scale Carnegie Mellon University (CMU) Panoptic dataset, achieving a mean absolute error (MAE) of 3.48°, 3.22°and 3.34° for yaw, pitch and roll respectively. Inter-instrument reliability was excellent for all six cervical movements when compared with the research/clinical-grade CROM device, with intraclass correlation coefficients (ICCs) averaging 0.939. Bland–Altman plots confirmed close agreement between the two methods. Cervical movement trajectory curves further confirmed the concordance between the clinical device and our method. The system is fully automatic, requires only a rear-facing camera, effectively preserves patient privacy, and provides accurate cervical posture estimation. This technology may provide a basis for future applications in neck-disorder screening, remote health monitoring, and personalized musculoskeletal wellness management, although further task-specific clinical validation will be required. To date, HPNet has been validated primarily on a computer-based platform and has not yet been deployed on smartphones. Future work will focus on model lightweighting, mobile deployment, and cross-device adaptation to facilitate its practical implementation on mobile devices.

SensorsVol. 26(16)
Harbin University of Science and Technology (CN), University of South Wales (GB), Durban University of Technology (ZA)
Openalex Percentile: Top 8%
Cervical and Thoracic Myelopathy
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.