Simulation-Based Functional Assessment of a Single Commercial AI-Assisted ECG Workflow: A Pre-Implementation Proof-of-Concept Study

Commercial artificial intelligence (AI)-assisted electrocardiography systems require local assessment of the acquisition-to-review pathway before patient-facing use. We conducted a single-site, simulation-based functional assessment of one commercial 12-lead ECG configuration. A physiological signal simulator, ECG acquisition unit, AI-assisted review platform, multiparameter monitor, and central monitoring system were used to examine 13 predefined rhythm and conduction categories, alarm behavior, local export pathways, a representative five-lead signal-display subset, and formative expert feedback. Expected category-level functional agreement was observed in 12 of 13 predefined categories. In the retained record, the discordant asystole scenario was categorized as lead-off/electrode disconnection, identifying a signal-integrity boundary that requires verification of electrodes, leads, cables, waveforms, and context before clinician adjudication. Signal-to-noise ratio and root-mean-square error values from 90 derived windows in five leads were reported descriptively. These findings cannot be generalized to the seven unmeasured diagnostic leads or to complete 12-lead signal fidelity. Local PDF, HL7 v2.x, DICOM, and platform synchronization were demonstrated without independent conformance or semantic-integrity testing. Formative input from six purposively selected experts comprised 30 ordinal ratings (median 5; observed range 4–5) and qualitative comments. These data do not constitute usability or educational-effectiveness validation. The study provides bounded functional evidence for the tested configuration and identifies requirements for prospective technical, human-factors, multivendor, and patient-level validation before clinical deployment.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2026-08-20
DOI
https://doi.org/10.3390/app16168282
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simulation-Based Functional Assessment of a Single Commercial AI-Assisted ECG Workflow: A Pre-Implementation Proof-of-Concept Study

Ángel Herranz, Leonel Vasquez-Cevallos, Susana Muñoz-Hernández, Pedro Salazar et al.
Applied Sciences
ECG Monitoring and Analysis
article

Simulation-Based Functional Assessment of a Single Commercial AI-Assisted ECG Workflow: A Pre-Implementation Proof-of-Concept Study

Ángel Herranz, Leonel Vasquez-Cevallos, Susana Muñoz-Hernández, Pedro Salazar, Paul E. D. Soto Rodriguez, Ricardo Grunauer-Robalino
article en

Abstract

Commercial artificial intelligence (AI)-assisted electrocardiography systems require local assessment of the acquisition-to-review pathway before patient-facing use. We conducted a single-site, simulation-based functional assessment of one commercial 12-lead ECG configuration. A physiological signal simulator, ECG acquisition unit, AI-assisted review platform, multiparameter monitor, and central monitoring system were used to examine 13 predefined rhythm and conduction categories, alarm behavior, local export pathways, a representative five-lead signal-display subset, and formative expert feedback. Expected category-level functional agreement was observed in 12 of 13 predefined categories. In the retained record, the discordant asystole scenario was categorized as lead-off/electrode disconnection, identifying a signal-integrity boundary that requires verification of electrodes, leads, cables, waveforms, and context before clinician adjudication. Signal-to-noise ratio and root-mean-square error values from 90 derived windows in five leads were reported descriptively. These findings cannot be generalized to the seven unmeasured diagnostic leads or to complete 12-lead signal fidelity. Local PDF, HL7 v2.x, DICOM, and platform synchronization were demonstrated without independent conformance or semantic-integrity testing. Formative input from six purposively selected experts comprised 30 ordinal ratings (median 5; observed range 4–5) and qualitative comments. These data do not constitute usability or educational-effectiveness validation. The study provides bounded functional evidence for the tested configuration and identifies requirements for prospective technical, human-factors, multivendor, and patient-level validation before clinical deployment.

Applied SciencesVol. 16(16)
Central University of Ecuador (EC), Universidad de La Laguna (ES), Universidad Estatal de Bolívar (EC), Universidad Ecotec (EC), Universidad Politécnica de Madrid (ES)
Openalex Percentile: Top 9%
ECG Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.