Spot size variation and skin interaction of long-pulse 1064nm Nd: YAG laser used for varicose veins treatment: A prospective, randomized, assessor-blinded ex vivo study

Background Long-pulsed Nd: YAG 1064 lasers are the standard of care for superficial vein transcutaneous laser treatment. The current concept suggests that the larger the laser spot size, the deeper the penetration into the skin for two reasons: First, wider spot sizes allow the use of higher pulse energy, as energy density (Fluence) decreases exponentially as the beam area widens. Second, the same pulse energy delivered by a smaller beam will be concentrated in the surface owing to the scattering effect. In this study, we examine the effect on skin penetration by changing the spot size of a long-pulsed Nd: YAG 1064 nm laser. Methods 100 different skin samples from 5 human dermolipectomy flaps were shot with 3 consecutive long-pulsed Nd: YAG 1064 nm pulses, one of each with a different spot size (3, 5, and 7 mm), while an energy meter was placed underneath to measure energy. In the first 80 samples, the same total energy (10J) was used for all 240 shots. In the last 20 samples, the same energy density (Fluence) of 100 J/cm2 was used in all 60 shots. Results No variations in mean energy across all spot sizes were observed (3 mm = 3.85 ± 0.89 J, 5 mm = 4.15 ± 0.83 J, 7 mm = 3.96 ± 0.73 J, p = 0.072). Regardless of the spot size, the average pulse energy attenuation was 60.1% with a standard deviation of 8.3%. The average fluence attenuation was 52.3% with a standard deviation of 5.7%. A strong correlation was found in the final energy measured in the same area, regardless of the spot size used. Conclusion In an ex vivo human skin model, varying the spot size of a long-pulsed Nd: YAG 1064 nm laser did not significantly affect energy transmission to the subdermal tissue.

Authors

Institutions

Publication Details

Journal
Phlebology The Journal of Venous Disease
Published
2026-08-20
DOI
https://doi.org/10.1177/02683555261478721
Primary Topic
Dermatologic Treatments and Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Spot size variation and skin interaction of long-pulse 1064nm Nd: YAG laser used for varicose veins treatment: A prospective, randomized, assessor-blinded ex vivo study

Rodrigo Kikuchi, Marcelo Halfen Grill, Viviane Santana da Silva, Roberto Augusto Caffaro et al.
Phlebology The Journal of Venous Disease
Dermatologic Treatments and Research
article

Spot size variation and skin interaction of long-pulse 1064nm Nd: YAG laser used for varicose veins treatment: A prospective, randomized, assessor-blinded ex vivo study

Rodrigo Kikuchi, Marcelo Halfen Grill, Viviane Santana da Silva, Roberto Augusto Caffaro, Juliana Amorim Teixeira Grill, Valéria Cristina Resende Aguiar, Eduardo Ramacciotti, Maria Fernanda Cassino Portugal, Nuno Vasconcelos, Samantha Neves, Thiago Alvarez Grill, Isabela Leal Zampirolli
article en

Abstract

Background Long-pulsed Nd: YAG 1064 lasers are the standard of care for superficial vein transcutaneous laser treatment. The current concept suggests that the larger the laser spot size, the deeper the penetration into the skin for two reasons: First, wider spot sizes allow the use of higher pulse energy, as energy density (Fluence) decreases exponentially as the beam area widens. Second, the same pulse energy delivered by a smaller beam will be concentrated in the surface owing to the scattering effect. In this study, we examine the effect on skin penetration by changing the spot size of a long-pulsed Nd: YAG 1064 nm laser. Methods 100 different skin samples from 5 human dermolipectomy flaps were shot with 3 consecutive long-pulsed Nd: YAG 1064 nm pulses, one of each with a different spot size (3, 5, and 7 mm), while an energy meter was placed underneath to measure energy. In the first 80 samples, the same total energy (10J) was used for all 240 shots. In the last 20 samples, the same energy density (Fluence) of 100 J/cm2 was used in all 60 shots. Results No variations in mean energy across all spot sizes were observed (3 mm = 3.85 ± 0.89 J, 5 mm = 4.15 ± 0.83 J, 7 mm = 3.96 ± 0.73 J, p = 0.072). Regardless of the spot size, the average pulse energy attenuation was 60.1% with a standard deviation of 8.3%. The average fluence attenuation was 52.3% with a standard deviation of 5.7%. A strong correlation was found in the final energy measured in the same area, regardless of the spot size used. Conclusion In an ex vivo human skin model, varying the spot size of a long-pulsed Nd: YAG 1064 nm laser did not significantly affect energy transmission to the subdermal tissue.

Phlebology The Journal of Venous Disease
Hospital Israelita Albert Einstein (BR), Clínica Rementería (ES), Faculdade de Ciências Médicas da Santa Casa de São Paulo (BR), Clínica Dávila (CL)
Affordable and clean energy
Openalex Percentile: Top 8%
Dermatologic Treatments and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.