Leveling the playing field: A pressure-sensing trocar improves speed and accuracy in tube thoracostomy

BACKGROUND: Tube thoracostomy is performed over one million times annually in the United States, including ~300,000 emergent procedures often without imaging guidance. Conventional open techniques carry complication rates of 25% to 30%, largely due to malpositioning. ThoraSET (Thoracic Safety Enhancing Trocar) is a pressure-sensing trocar system that differentiates tissue layers and provides real-time visual feedback. We hypothesized that ThoraSET would improve placement accuracy and reduce placement time, narrowing the gap between novice and expert users. METHODS: This prospective study combined iterative device development with stakeholder feedback and comparative evaluation on porcine cadaver and mannequin models. Participants across four experience levels (novices, residents, APPs, attending surgeons) completed paired open and ThoraSET-assisted tube thoracostomy trials. Primary outcomes were placement time and anatomical accuracy. Paired t tests and McNemar or exact McNemar tests were used for analysis, with sensitivity analyses by model type. RESULTS: Among 25 participants, ThoraSET reduced placement time for novices (-62%), residents (-26%), and APPs (-43%), but increased time for attendings (+33%). Large effect sizes were observed for novices and APPs. Overall accuracy was higher with ThoraSET (79% vs. 54%; P=0.07). Residents' accuracy improved from 10% to 80%; APPs and attendings remained 100%; novices decreased from 50% to 25%. Sensitivity analyses confirmed similar trends, with significant time reductions on pig (p=0.038, d=-0.57) and mannequin models (P=0.028, d=-0.73). CONCLUSIONS: ThoraSET improved placement speed and accuracy, particularly among less experienced users, and may reduce procedural variability and complications. Its mechanically responsive, blunt trocar provides tactile feedback, enhanced control, and intrathoracic access, supporting safer placement and use of smaller, softer tubes. These pilot results suggest ThoraSET has the potential to elevate safety and standardization in emergency and trauma care, warranting further clinical validation. (J Trauma Acute Care Surg. 2026;00: 000-000.). LEVEL OF EVIDENCE: Therapeutic/Care Management; Level V.

Authors

Institutions

Publication Details

Journal
The Journal of Trauma: Injury, Infection, and Critical Care
Published
2026-08-20
DOI
https://doi.org/10.1097/ta.0000000000005134
Primary Topic
Pleural and Pulmonary Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Leveling the playing field: A pressure-sensing trocar improves speed and accuracy in tube thoracostomy

Adrian Camarena, Jacob Peloquin, Esther Park, Joseph Fernandez-Moure et al.
The Journal of Trauma: Injury, Infection, and Critical Care
Pleural and Pulmonary Diseases
article

Leveling the playing field: A pressure-sensing trocar improves speed and accuracy in tube thoracostomy

Adrian Camarena, Jacob Peloquin, Esther Park, Joseph Fernandez-Moure, Anna Medrano, Parker Frankiewicz
article en

Abstract

BACKGROUND: Tube thoracostomy is performed over one million times annually in the United States, including ~300,000 emergent procedures often without imaging guidance. Conventional open techniques carry complication rates of 25% to 30%, largely due to malpositioning. ThoraSET (Thoracic Safety Enhancing Trocar) is a pressure-sensing trocar system that differentiates tissue layers and provides real-time visual feedback. We hypothesized that ThoraSET would improve placement accuracy and reduce placement time, narrowing the gap between novice and expert users. METHODS: This prospective study combined iterative device development with stakeholder feedback and comparative evaluation on porcine cadaver and mannequin models. Participants across four experience levels (novices, residents, APPs, attending surgeons) completed paired open and ThoraSET-assisted tube thoracostomy trials. Primary outcomes were placement time and anatomical accuracy. Paired t tests and McNemar or exact McNemar tests were used for analysis, with sensitivity analyses by model type. RESULTS: Among 25 participants, ThoraSET reduced placement time for novices (-62%), residents (-26%), and APPs (-43%), but increased time for attendings (+33%). Large effect sizes were observed for novices and APPs. Overall accuracy was higher with ThoraSET (79% vs. 54%; P=0.07). Residents' accuracy improved from 10% to 80%; APPs and attendings remained 100%; novices decreased from 50% to 25%. Sensitivity analyses confirmed similar trends, with significant time reductions on pig (p=0.038, d=-0.57) and mannequin models (P=0.028, d=-0.73). CONCLUSIONS: ThoraSET improved placement speed and accuracy, particularly among less experienced users, and may reduce procedural variability and complications. Its mechanically responsive, blunt trocar provides tactile feedback, enhanced control, and intrathoracic access, supporting safer placement and use of smaller, softer tubes. These pilot results suggest ThoraSET has the potential to elevate safety and standardization in emergency and trauma care, warranting further clinical validation. (J Trauma Acute Care Surg. 2026;00: 000-000.). LEVEL OF EVIDENCE: Therapeutic/Care Management; Level V.

The Journal of Trauma: Injury, Infection, and Critical Care
Duke University (US)
Good health and well-being
Openalex Percentile: Top 10%
Pleural and Pulmonary Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.