The impact of abutment manufacturing techniques on reverse torque after cyclic loading and implant-abutment interface fitting: an in vitro study

BACKGROUND: Screw loosening and implant-abutment interface (IAI) are critical for the long-term success of implant-supported restorations. This study aimed to evaluate the effect of abutment manufacturing technique on reverse torque values and micro-gap measurements at the implant-abutment interface. METHODS: Thirty dental implant-abutment-crown complexes (Implance, Verdent, Turkey) were randomly allocated into three groups (n = 10 each) based on the abutment manufacturing technique: Group I-prefabricated titanium abutments (manufacturer-supplied, Implance, Verdent, Turkey; titanium grade 5); Group II-CAD-CAM-milled titanium abutments; Group III-3D-printed titanium abutments (Selective Laser Melting, SLM). All abutments featured an internal hexagonal connection with platform switching. All crown abutments were tightened to 25 Ncm and subjected to cyclic loading (100 N, 1.6 Hz, 250,000 cycles). Removal torque was measured before and after loading. Micro-gaps at the implant-abutment and abutment-screw interfaces were evaluated using scanning electron microscopy (SEM). Statistical analysis was performed using One-Way ANOVA with Tukey's Post Hoc test for reverse torque and the Kruskal-Wallis test for gap measurements (α = 0.05). RESULTS: After cyclic loading, the 3D-printed (SLM) abutments exhibited significantly higher reverse torque values (11.08 ± 0.11 Ncm) compared to CAD-CAM-milled abutments (6.40 ± 0.79 Ncm) (P < 0.001). Prefabricated abutments demonstrated the highest post-loading torque values (13.31 ± 0.31 Ncm). All intragroup comparisons showed statistically significant reductions from baseline (P < 0.0001). No significant difference was found in implant-abutment interface gaps among groups (P = 0.105). CONCLUSION: Within the limitations of this study, prefabricated abutments demonstrated the highest reverse torque values after cyclic loading. 3D-printed abutments showed significantly greater resistance to torque loss compared to CAD-CAM-milled abutments, approaching the performance of prefabricated abutments. The implant-abutment fit was comparable across groups, suggesting that connection design may be more influential than manufacturing method for macroscopic fit. 3D-printed titanium abutments demonstrated significantly greater mechanical stability than CAD-CAM-milled abutments, approaching the performance of the gold-standard prefabricated abutments. This suggests that additive manufacturing may offer a clinically viable alternative to milling when both customization and mechanical integrity are required. However, clinical studies are needed to confirm these in vitro findings.

Authors

Institutions

Publication Details

Journal
BMC Oral Health
Published
2026-08-20
DOI
https://doi.org/10.1186/s12903-026-09530-w
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The impact of abutment manufacturing techniques on reverse torque after cyclic loading and implant-abutment interface fitting: an in vitro study

Ahmed Ismail Taha, Nourhan Ahmed Ragheb, Alaa Mostafa AboRas
BMC Oral Health
Dental Implant Techniques and Outcomes
article

The impact of abutment manufacturing techniques on reverse torque after cyclic loading and implant-abutment interface fitting: an in vitro study

Ahmed Ismail Taha, Nourhan Ahmed Ragheb, Alaa Mostafa AboRas
article en

Abstract

BACKGROUND: Screw loosening and implant-abutment interface (IAI) are critical for the long-term success of implant-supported restorations. This study aimed to evaluate the effect of abutment manufacturing technique on reverse torque values and micro-gap measurements at the implant-abutment interface. METHODS: Thirty dental implant-abutment-crown complexes (Implance, Verdent, Turkey) were randomly allocated into three groups (n = 10 each) based on the abutment manufacturing technique: Group I-prefabricated titanium abutments (manufacturer-supplied, Implance, Verdent, Turkey; titanium grade 5); Group II-CAD-CAM-milled titanium abutments; Group III-3D-printed titanium abutments (Selective Laser Melting, SLM). All abutments featured an internal hexagonal connection with platform switching. All crown abutments were tightened to 25 Ncm and subjected to cyclic loading (100 N, 1.6 Hz, 250,000 cycles). Removal torque was measured before and after loading. Micro-gaps at the implant-abutment and abutment-screw interfaces were evaluated using scanning electron microscopy (SEM). Statistical analysis was performed using One-Way ANOVA with Tukey's Post Hoc test for reverse torque and the Kruskal-Wallis test for gap measurements (α = 0.05). RESULTS: After cyclic loading, the 3D-printed (SLM) abutments exhibited significantly higher reverse torque values (11.08 ± 0.11 Ncm) compared to CAD-CAM-milled abutments (6.40 ± 0.79 Ncm) (P < 0.001). Prefabricated abutments demonstrated the highest post-loading torque values (13.31 ± 0.31 Ncm). All intragroup comparisons showed statistically significant reductions from baseline (P < 0.0001). No significant difference was found in implant-abutment interface gaps among groups (P = 0.105). CONCLUSION: Within the limitations of this study, prefabricated abutments demonstrated the highest reverse torque values after cyclic loading. 3D-printed abutments showed significantly greater resistance to torque loss compared to CAD-CAM-milled abutments, approaching the performance of prefabricated abutments. The implant-abutment fit was comparable across groups, suggesting that connection design may be more influential than manufacturing method for macroscopic fit. 3D-printed titanium abutments demonstrated significantly greater mechanical stability than CAD-CAM-milled abutments, approaching the performance of the gold-standard prefabricated abutments. This suggests that additive manufacturing may offer a clinically viable alternative to milling when both customization and mechanical integrity are required. However, clinical studies are needed to confirm these in vitro findings.

BMC Oral HealthVol. 26(1)
Kafrelsheikh University (EG)
Openalex Percentile: Top 9%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.