Ultrasound measurement of skin-to-septum distance in totally implantable venous access ports: Considerations for Huber needle length

Background: Objective criteria for selecting Huber needle length for totally implantable venous access ports (TIVAPs) are limited. This study primarily described skin-to-septum distance using ultrasonography and its correlation with body habitus. Methods: Thirty-three patients with TIVAPs for chemotherapy were included. Skin-to-septum distance was measured by ultrasonography in two positions: supine with arms at the sides and with shoulder flexion. Pearson’s correlation coefficient assessed correlation between body mass index (BMI) and septal depth. Excess needle length beyond the port base was calculated. Results: Mean skin-to-septum distance was 1.7 ± 0.4 mm (arms at sides) and 1.8 ± 0.5 mm (shoulder flexion). A 16-mm Huber needle was used in 97.0%. The median excess needle length was 2.4 mm. BMI showed weak correlation with septal depth ( r = 0.33, p = 0.062). Conclusions: Septal depth was shallow, and BMI alone was not a reliable predictor. Ultrasound may support Huber needle selection.

Authors

Institutions

Publication Details

Journal
The Journal of Vascular Access
Published
2026-08-18
DOI
https://doi.org/10.1177/11297298261472621
Primary Topic
Central Venous Catheters and Hemodialysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Ultrasound measurement of skin-to-septum distance in totally implantable venous access ports: Considerations for Huber needle length

Takatoshi Kubo, Kazunaga Ishigaki, Gojiro Nakagami, Mari Abe‐Doi et al.
The Journal of Vascular Access
Central Venous Catheters and Hemodialysis
article

Ultrasound measurement of skin-to-septum distance in totally implantable venous access ports: Considerations for Huber needle length

Takatoshi Kubo, Kazunaga Ishigaki, Gojiro Nakagami, Mari Abe‐Doi, Hironori Ishigami, Hikaru Haruki, Ayako Suzuki
article en

Abstract

Background: Objective criteria for selecting Huber needle length for totally implantable venous access ports (TIVAPs) are limited. This study primarily described skin-to-septum distance using ultrasonography and its correlation with body habitus. Methods: Thirty-three patients with TIVAPs for chemotherapy were included. Skin-to-septum distance was measured by ultrasonography in two positions: supine with arms at the sides and with shoulder flexion. Pearson’s correlation coefficient assessed correlation between body mass index (BMI) and septal depth. Excess needle length beyond the port base was calculated. Results: Mean skin-to-septum distance was 1.7 ± 0.4 mm (arms at sides) and 1.8 ± 0.5 mm (shoulder flexion). A 16-mm Huber needle was used in 97.0%. The median excess needle length was 2.4 mm. BMI showed weak correlation with septal depth ( r = 0.33, p = 0.062). Conclusions: Septal depth was shallow, and BMI alone was not a reliable predictor. Ultrasound may support Huber needle selection.

The Journal of Vascular Access
University of Tokyo Hospital (JP), Next Generation Technology (United States) (US), Feminist Center for Studies and Advisory Services (BR), The University of Tokyo (JP)
Openalex Percentile: Top 6%
Central Venous Catheters and Hemodialysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Ultrasound measurement of skin-to-septum distance in totally implantable venous access ports: Considerations for Huber needle length — Takatoshi Kubo, Kazunaga Ishigaki, et al. · The Journal of Vascular Access (2026) | TGRS Research Map | TGRS