Fabrication of a 3D-Printed Fixation Device and Quantitative Analysis of Reproducibility in Hip Joint Axiolateral Radiography

This study aimed to evaluate the utility of a 3D-printed, weight-bearing auxiliary device designed to improve reproducibility and facilitate detector fixation during axiolateral radiography of hip joints known to be affected by inter-operator variability. The experiment was conducted using a radiolucent human phantom. Multiple radiographs were taken under identical conditions. Reproducibility and utility were assessed by measuring the α angle and off-set ratio from acquired images. Results showed that the mean α angle was 51.58° (±5.14°) with the auxiliary device and 50.45° (±4.90°) without it, with no statistically significant difference between conditions (Wilcoxon signed-rank test, p = 0.2024), indicating no measurable improvement in reproducibility. The off-set ratio likewise did not differ significantly between conditions (p = 0.1704). However, using the device did not significantly affect the off-set ratio. In conclusion, the auxiliary device developed in this study allows for consistent detector angle maintenance and fixation during hip joint axiolateral projection, thereby improving reproducibility and ease of examination. This suggests its potential usefulness for achieving more accurate imaging in clinical settings.

Authors

Institutions

Publication Details

Journal
Radiation
Published
2026-08-18
DOI
https://doi.org/10.3390/radiation6030032
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fabrication of a 3D-Printed Fixation Device and Quantitative Analysis of Reproducibility in Hip Joint Axiolateral Radiography

Yeongcheol Joo, Eunhye Kim, Donghee Hong
Radiation
Orthopaedic implants and arthroplasty
article

Fabrication of a 3D-Printed Fixation Device and Quantitative Analysis of Reproducibility in Hip Joint Axiolateral Radiography

Yeongcheol Joo, Eunhye Kim, Donghee Hong
article en

Abstract

This study aimed to evaluate the utility of a 3D-printed, weight-bearing auxiliary device designed to improve reproducibility and facilitate detector fixation during axiolateral radiography of hip joints known to be affected by inter-operator variability. The experiment was conducted using a radiolucent human phantom. Multiple radiographs were taken under identical conditions. Reproducibility and utility were assessed by measuring the α angle and off-set ratio from acquired images. Results showed that the mean α angle was 51.58° (±5.14°) with the auxiliary device and 50.45° (±4.90°) without it, with no statistically significant difference between conditions (Wilcoxon signed-rank test, p = 0.2024), indicating no measurable improvement in reproducibility. The off-set ratio likewise did not differ significantly between conditions (p = 0.1704). However, using the device did not significantly affect the off-set ratio. In conclusion, the auxiliary device developed in this study allows for consistent detector angle maintenance and fixation during hip joint axiolateral projection, thereby improving reproducibility and ease of examination. This suggests its potential usefulness for achieving more accurate imaging in clinical settings.

RadiationVol. 6(3)
Hanseo University (KR), Samsung Medical Center (KR), Shinhan University (KR)
Openalex Percentile: Top 7%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.