First-in-human use of SmartGUIDETM steerable 0.014’’ microguidewire for the treatment of intracranial aneurysms: a prospective single-center case series

Abstract Purpose To report the first-in-human, prospective clinical evaluation of a steerable 0.014″ microguidewire SmartGUIDE™ (Artiria Medical, Switzerland) for endovascular treatment of intracranial aneurysms, assessing technical feasibility, safety, and operator-reported usability. Methods Adult patients undergoing elective treatment of unruptured intracranial aneurysms were prospectively enrolled between August 2023 and July 2024 at a single center. All procedures were performed by a senior neurointerventionalist. The primary endpoint was the device technical success, defined as successful SmartGUIDE™ navigation to a predefined anatomical target. Successful microcatheterization over the wire was a clinically meaningful secondary endpoint, together with safety, device-related adverse events, navigation time, and operator-reported usability (5-point Likert scale). Patient-specific Flow Twin ® vascular replicas were used for pre-therapeutic training to refine strategy and mitigate risks inherent to first-in-human use. Clinical and imaging follow-up was performed at 1 year. Results Five patients were included. SmartGUIDE™ successfully reached the predefined target in all procedures (5/5). Microcatheterization was achieved in four cases (4/5). Median SmartGUIDE navigation time was 7 min (IQR 4–10). No wire-related complications or device malfunctions were observed. Operators reported high device usability, with median Likert scores of 4–5/5 across all parameters. At 1-year follow-up, all aneurysms demonstrated complete and stable occlusion without neurological morbidity. Conclusion This first-in-human prospective series supports the feasibility and safety of the SmartGUIDE™ steerable 0.014″ microguidewire for the treatment of intracranial aneurysms. The study reports reliable navigation to the target without device-related malfunction or complication, favorable operator-reported usability, and highlights the value of incorporating patient-specific flow-models for pre-procedural training. Comparative studies are required to establish efficiency advantage over conventional microguidewires.

Authors

Institutions

Publication Details

Journal
Neuroradiology
Published
2026-08-18
DOI
https://doi.org/10.1007/s00234-026-04143-0
Primary Topic
Intracranial Aneurysms: Treatment and Complications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

First-in-human use of SmartGUIDETM steerable 0.014’’ microguidewire for the treatment of intracranial aneurysms: a prospective single-center case series

Karl‐Olof Lövblad, Andrea Rosi, Philippe Reymond, Priya Gaiha et al.
Neuroradiology
Intracranial Aneurysms: Treatment and Complications
article

First-in-human use of SmartGUIDETM steerable 0.014’’ microguidewire for the treatment of intracranial aneurysms: a prospective single-center case series

Karl‐Olof Lövblad, Andrea Rosi, Philippe Reymond, Priya Gaiha, Olivier Brina, Gianmarco Bernava, Michel Muster, Jérémy Hofmeister, Paolo Machi, Hasan Yilmaz, Marc Boers, Guillaume Petitpierre
article en

Abstract

Abstract Purpose To report the first-in-human, prospective clinical evaluation of a steerable 0.014″ microguidewire SmartGUIDE™ (Artiria Medical, Switzerland) for endovascular treatment of intracranial aneurysms, assessing technical feasibility, safety, and operator-reported usability. Methods Adult patients undergoing elective treatment of unruptured intracranial aneurysms were prospectively enrolled between August 2023 and July 2024 at a single center. All procedures were performed by a senior neurointerventionalist. The primary endpoint was the device technical success, defined as successful SmartGUIDE™ navigation to a predefined anatomical target. Successful microcatheterization over the wire was a clinically meaningful secondary endpoint, together with safety, device-related adverse events, navigation time, and operator-reported usability (5-point Likert scale). Patient-specific Flow Twin ® vascular replicas were used for pre-therapeutic training to refine strategy and mitigate risks inherent to first-in-human use. Clinical and imaging follow-up was performed at 1 year. Results Five patients were included. SmartGUIDE™ successfully reached the predefined target in all procedures (5/5). Microcatheterization was achieved in four cases (4/5). Median SmartGUIDE navigation time was 7 min (IQR 4–10). No wire-related complications or device malfunctions were observed. Operators reported high device usability, with median Likert scores of 4–5/5 across all parameters. At 1-year follow-up, all aneurysms demonstrated complete and stable occlusion without neurological morbidity. Conclusion This first-in-human prospective series supports the feasibility and safety of the SmartGUIDE™ steerable 0.014″ microguidewire for the treatment of intracranial aneurysms. The study reports reliable navigation to the target without device-related malfunction or complication, favorable operator-reported usability, and highlights the value of incorporating patient-specific flow-models for pre-procedural training. Comparative studies are required to establish efficiency advantage over conventional microguidewires.

Neuroradiology
University of Geneva (CH), University Hospital of Geneva (CH), University Hospitals Geneva Medical Center (US)
Good health and well-being
Openalex Percentile: Top 10%
Intracranial Aneurysms: Treatment and Complications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.