Comparative In Vitro Evaluation of Tube Resistance in Glaucoma Drainage Devices with Intraluminal Polypropylene Threads

(1) Background: To compare tube resistance among glaucoma drainage devices (GDDs) with different tube geometries under various intraluminal polypropylene (PP) thread conditions and assess agreement between measured and formula-derived pressure–flow profiles. (2) Methods: Four GDDs were evaluated: Ahmed ClearPath Small Tube (ACP-ST), Paul Glaucoma Implant (PGI), Ahmed ClearPath (ACP), and Virna Glaucoma Implant (VGI) (n = 4/device). ACP-ST and PGI were tested under no-thread, 6-0 PP Corza, and 6-0 PP Ethicon conditions, whereas ACP and VGI were tested under no-thread, 4-0 PP Corza, and 3-0 PP Ethicon conditions. Saline was infused at 1.0–5.0 µL/min, and pressure was recorded using a transducer system. Tube and thread dimensions were measured using ImageJ and used to calculate formula-derived pressure from fluid dynamic equations. Comparisons used linear mixed-effects regression. (3) Results: In ACP-ST and PGI, both 6-0 PP threads significantly increased pressure compared with the no-thread condition, with no significant thread manufacturer-related differences. In ACP and VGI, all threaded conditions significantly increased pressure, and the 3-0 PP Ethicon condition produced higher pressure than the 4-0 PP Corza condition. Under threaded conditions, measured and formula-derived pressure values did not differ significantly in ACP-ST but differed significantly in PGI, ACP, and VGI. (4) Conclusions: Intraluminal PP threads increased GDD tube resistance. Manufacturer-related differences had a limited influence on resistance in small-lumen devices. Agreement between measured and formula-derived pressure varied among devices, highlighting the importance of experimental validation. These findings may help guide intraluminal thread selection and flow-restriction strategies during GDD implantation.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-08-14
DOI
https://doi.org/10.3390/bioengineering13080923
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative In Vitro Evaluation of Tube Resistance in Glaucoma Drainage Devices with Intraluminal Polypropylene Threads

Virna Dwi Oktariana Asrory, Andi Masdipa, Sachiko Kaidzu, Masaki Tanito et al.
Bioengineering
Glaucoma and retinal disorders
article

Comparative In Vitro Evaluation of Tube Resistance in Glaucoma Drainage Devices with Intraluminal Polypropylene Threads

Virna Dwi Oktariana Asrory, Andi Masdipa, Sachiko Kaidzu, Masaki Tanito, Arief Abdurrazaq Dharma
article en

Abstract

(1) Background: To compare tube resistance among glaucoma drainage devices (GDDs) with different tube geometries under various intraluminal polypropylene (PP) thread conditions and assess agreement between measured and formula-derived pressure–flow profiles. (2) Methods: Four GDDs were evaluated: Ahmed ClearPath Small Tube (ACP-ST), Paul Glaucoma Implant (PGI), Ahmed ClearPath (ACP), and Virna Glaucoma Implant (VGI) (n = 4/device). ACP-ST and PGI were tested under no-thread, 6-0 PP Corza, and 6-0 PP Ethicon conditions, whereas ACP and VGI were tested under no-thread, 4-0 PP Corza, and 3-0 PP Ethicon conditions. Saline was infused at 1.0–5.0 µL/min, and pressure was recorded using a transducer system. Tube and thread dimensions were measured using ImageJ and used to calculate formula-derived pressure from fluid dynamic equations. Comparisons used linear mixed-effects regression. (3) Results: In ACP-ST and PGI, both 6-0 PP threads significantly increased pressure compared with the no-thread condition, with no significant thread manufacturer-related differences. In ACP and VGI, all threaded conditions significantly increased pressure, and the 3-0 PP Ethicon condition produced higher pressure than the 4-0 PP Corza condition. Under threaded conditions, measured and formula-derived pressure values did not differ significantly in ACP-ST but differed significantly in PGI, ACP, and VGI. (4) Conclusions: Intraluminal PP threads increased GDD tube resistance. Manufacturer-related differences had a limited influence on resistance in small-lumen devices. Agreement between measured and formula-derived pressure varied among devices, highlighting the importance of experimental validation. These findings may help guide intraluminal thread selection and flow-restriction strategies during GDD implantation.

BioengineeringVol. 13(8)
Shimane University (JP), University of Indonesia (ID), Universitas Muslim Indonesia (ID), Rumah Sakit Umum Pusat Nasional Dr. Cipto Mangunkusumo (ID)
Japan Science and Technology Agency
Openalex Percentile: Top 7%
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.