Evaluation of 7 Tesla <scp>pTx</scp> Body Coils Regarding Transmit and Receive Performance

ABSTRACT Purpose Most sites develop and use custom coils for 7 T body MRI since no standard pTx body coil exists, leading to diverse coil designs differing in transmit element type, layout, and number of receive channels. This study investigates and compares eight existing coils, including one remote body coil array and seven local arrays, regarding their transmit and receive performance. Methods Phantom measurements were conducted with all coils on the same 7 T scanner, using the same phantom and imaging protocol. Transmit performance was compared in terms of B 1 + efficiency and coverage. Receive performance was compared in terms of SNR, coverage, noise correlation, and g‐factors. Results Mean B 1 + efficiency ranged from 2.02 to 4.33 μT/√kW across configurations, lowest for the remote array and highest for an 8Tx8Rx local array. Central SNR ranged from 278 to 1072, increasing with receive element count and peaking for an 8Tx32Rx configuration. HF‐excitation‐coverage ranged from 34 to 384 mm, and HF‐receive‐coverage from 135 to 384 mm, with the remote array combined with a local 32Rx array achieving highest coverage. Acceleration performance improved with increasing receive element count in the corresponding direction. An 8Tx16Rx and an 8Tx32Rx arrays performed best in LR‐/AP‐direction and the 32Rx array in HF‐direction. Conclusion No existing local pTx coil provides universally optimal performance, as each design offers different advantages in either B 1 + efficiency, coverage, SNR, or acceleration. Among the local arrays, the 8Tx32Rx array (C5) might represent a reasonable compromise with respect to the parameters evaluated in this study, as it exhibits the highest central SNR, high coverage, and acceleration.

Authors

Institutions

Publication Details

Journal
Magnetic Resonance in Medicine
Published
2026-08-14
DOI
https://doi.org/10.1002/mrm.70557
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of 7 Tesla pTx Body Coils Regarding Transmit and Receive Performance

M Ladd, Stephan Orzada, Sebastian Schmitter, Thomas M. Fiedler et al.
Magnetic Resonance in Medicine
Advanced MRI Techniques and Applications
article

Evaluation of 7 Tesla pTx Body Coils Regarding Transmit and Receive Performance

M Ladd, Stephan Orzada, Sebastian Schmitter, Thomas M. Fiedler, Max Lutz, Oliver Kraff, Markus May, Harald H. Quick, Johannes A. Grimm, Armin Michael Nagel, Steen Moeller, Simon Schmidt
article en

Abstract

ABSTRACT Purpose Most sites develop and use custom coils for 7 T body MRI since no standard pTx body coil exists, leading to diverse coil designs differing in transmit element type, layout, and number of receive channels. This study investigates and compares eight existing coils, including one remote body coil array and seven local arrays, regarding their transmit and receive performance. Methods Phantom measurements were conducted with all coils on the same 7 T scanner, using the same phantom and imaging protocol. Transmit performance was compared in terms of B 1 + efficiency and coverage. Receive performance was compared in terms of SNR, coverage, noise correlation, and g‐factors. Results Mean B 1 + efficiency ranged from 2.02 to 4.33 μT/√kW across configurations, lowest for the remote array and highest for an 8Tx8Rx local array. Central SNR ranged from 278 to 1072, increasing with receive element count and peaking for an 8Tx32Rx configuration. HF‐excitation‐coverage ranged from 34 to 384 mm, and HF‐receive‐coverage from 135 to 384 mm, with the remote array combined with a local 32Rx array achieving highest coverage. Acceleration performance improved with increasing receive element count in the corresponding direction. An 8Tx16Rx and an 8Tx32Rx arrays performed best in LR‐/AP‐direction and the 32Rx array in HF‐direction. Conclusion No existing local pTx coil provides universally optimal performance, as each design offers different advantages in either B 1 + efficiency, coverage, SNR, or acceleration. Among the local arrays, the 8Tx32Rx array (C5) might represent a reasonable compromise with respect to the parameters evaluated in this study, as it exhibits the highest central SNR, high coverage, and acceleration.

Magnetic Resonance in Medicine
Erwin L. Hahn Institute for Magnetic Resonance Imaging (DE), Physikalisch-Technische Bundesanstalt (DE), University of Minnesota (US), German Cancer Research Center (DE), Friedrich-Alexander-Universität Erlangen-Nürnberg (DE), Heidelberg University (DE), Essen University Hospital (DE), Resonance Research (United States) (US)
Deutsche Forschungsgemeinschaft, National Institutes of Health
Affordable and clean energy
Openalex Percentile: Top 10%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.