Determination of institute specific CTV to PTV prostate margins using the 2020 BIR IGRT geometric uncertainty guidelines for a conventional linear accelerator

OBJECTIVES: A multitude of differing clinical practice exists for prostate radiotherapy, this includes dose prescription, planning methods, asymmetric margins and the presence, or not, of fiducial markers. Institute specific CTV to PTV margins expansion is desirable for optimum treatment of patients. Margins can be determined by assessing geometric errors inherent in radiotherapy. METHODS: The institutional CTV to PTV margins for fiducial free prostate treatments are calculated based on the BIR 2020 methodology derived from the van Herk formulism. Systematic (Σ_tot) and random (σ_tot) errors including delineation, matching and technical accuracy were measured locally while rotational, deformation and intra-fraction errors were quantified from the literature.Σtot and σtot and the penumbral width (σp) are combined with statistical parameters (α) and (β) to calculate the institutional CTV to PTV margin expansion (M) for prostate treatments. RESULTS: An anisotropic CTV to PTV margin expansion for prostate EBRT was calculated. The resulting CTV to PTV margin was 4 mm left/right, 7.5 mm superior, 7 mm inferior, 6 mm anterior and 5.5 mm posterior, rounded to the nearest 0.5 mm. CONCLUSIONS: Systematic errors contributed substantially to the overall margin particularly due to the magnitude of delineation errors. The margin calculated was smaller than the previously set institutional margin and ensures coverage of the PTV by the 95% isodose for 90% of prostate treatments at our centre. ADVANCES IN KNOWLEDGE: Institute specific margins for prostate radiotherapy on a linear accelerator are provided. These are calculated using the BIR 2020 methodology of CTV-PTV margin calculation.

Authors

Institutions

Publication Details

Journal
British Journal of Radiology
Published
2026-08-13
DOI
https://doi.org/10.1093/bjr/tqag187
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determination of institute specific CTV to PTV prostate margins using the 2020 BIR IGRT geometric uncertainty guidelines for a conventional linear accelerator

Séan M. O’Cathail, Michael Roche, C. Lyons, Orla McKivitt et al.
British Journal of Radiology
Advanced Radiotherapy Techniques
article

Determination of institute specific CTV to PTV prostate margins using the 2020 BIR IGRT geometric uncertainty guidelines for a conventional linear accelerator

Séan M. O’Cathail, Michael Roche, C. Lyons, Orla McKivitt, Christopher Walker, Michael Moran, Mohammed F Jamaluddin
article en

Abstract

OBJECTIVES: A multitude of differing clinical practice exists for prostate radiotherapy, this includes dose prescription, planning methods, asymmetric margins and the presence, or not, of fiducial markers. Institute specific CTV to PTV margins expansion is desirable for optimum treatment of patients. Margins can be determined by assessing geometric errors inherent in radiotherapy. METHODS: The institutional CTV to PTV margins for fiducial free prostate treatments are calculated based on the BIR 2020 methodology derived from the van Herk formulism. Systematic (Σ_tot) and random (σ_tot) errors including delineation, matching and technical accuracy were measured locally while rotational, deformation and intra-fraction errors were quantified from the literature.Σtot and σtot and the penumbral width (σp) are combined with statistical parameters (α) and (β) to calculate the institutional CTV to PTV margin expansion (M) for prostate treatments. RESULTS: An anisotropic CTV to PTV margin expansion for prostate EBRT was calculated. The resulting CTV to PTV margin was 4 mm left/right, 7.5 mm superior, 7 mm inferior, 6 mm anterior and 5.5 mm posterior, rounded to the nearest 0.5 mm. CONCLUSIONS: Systematic errors contributed substantially to the overall margin particularly due to the magnitude of delineation errors. The margin calculated was smaller than the previously set institutional margin and ensures coverage of the PTV by the 95% isodose for 90% of prostate treatments at our centre. ADVANCES IN KNOWLEDGE: Institute specific margins for prostate radiotherapy on a linear accelerator are provided. These are calculated using the BIR 2020 methodology of CTV-PTV margin calculation.

British Journal of Radiology
University Hospital Waterford (IE), Cork University Hospital (IE)
Sustainable cities and communities
Openalex Percentile: Top 10%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.