Endoleak detection with photon-counting detector CT at various radiation doses after endovascular aneurysm repair: a phantom study

Abstract After endovascular aneurysm repair, patients require lifelong CT follow-ups. Photon-counting detector CT (PCD-CT) may improve detection of tiny iodinated structures at low radiation dose. The aim of this experimental study was to find the lowest dose which still provides high detectability of small endoleaks using PCD-CT with 55 keV and 50 keV virtual monoenergetic images (VMIs). An abdominal aneurysm phantom mimicking medium and large patients was scanned at 100%, 50%, 25%, and 12.5% radiation dose using PCD-CT. Besides standard 55 keV images, 50 keV VMI were reconstructed at 25% and 12.5% dose. The ninety-six CT series containing 1728 endoleaks of 2–6 mm were read by three independent observers. Areas under the alternative free-response operational characteristic curve (A 1 ) were analysed with the jackknife-method. In the medium phantom, A 1 for 6 mm endoleaks was similar at 100% and 50% doses (0.93 vs. 0.86; p = 0.23), but performance at 50% dose decreased in 4–6 mm endoleaks (A 1 , 0.81 vs. 0.68; p = 0.02). In the large phantom, A 1 decreased at 50% dose both for 6 mm (0.72 vs. 0.42; p = 0.003) and for 4–6 mm (0.49 vs. 0.29; p = 0.02) endoleaks. Compared with standard reconstruction, 50 keV VMIs did not change A 1 at 25% dose (0.56 vs. 0.52; p = 0.38) and led to worse performance at 12.5% dose (0.30 vs. 0.19; p = 0.01). In summary, diagnostic performance for 6 mm endoleaks remained high with 50%-dose reduction using PCD-CT in simulated medium but not in large patients. 50 keV VMIs at 12.5% and 25% dose did not improve endoleak detection.

Authors

Institutions

Publication Details

Journal
The International Journal of Cardiovascular Imaging
Published
2026-08-13
DOI
https://doi.org/10.1007/s10554-026-03765-y
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Endoleak detection with photon-counting detector CT at various radiation doses after endovascular aneurysm repair: a phantom study

Andreas Christe, Zsolt Szucs‐Farkas, Adrian Huber, Badis M’Hamdi et al.
The International Journal of Cardiovascular Imaging
Advanced X-ray and CT Imaging
article

Endoleak detection with photon-counting detector CT at various radiation doses after endovascular aneurysm repair: a phantom study

Andreas Christe, Zsolt Szucs‐Farkas, Adrian Huber, Badis M’Hamdi, Ludovic G. Perrin, Gergely Szucs-Farkas, Sofia Bespyatova
article en

Abstract

Abstract After endovascular aneurysm repair, patients require lifelong CT follow-ups. Photon-counting detector CT (PCD-CT) may improve detection of tiny iodinated structures at low radiation dose. The aim of this experimental study was to find the lowest dose which still provides high detectability of small endoleaks using PCD-CT with 55 keV and 50 keV virtual monoenergetic images (VMIs). An abdominal aneurysm phantom mimicking medium and large patients was scanned at 100%, 50%, 25%, and 12.5% radiation dose using PCD-CT. Besides standard 55 keV images, 50 keV VMI were reconstructed at 25% and 12.5% dose. The ninety-six CT series containing 1728 endoleaks of 2–6 mm were read by three independent observers. Areas under the alternative free-response operational characteristic curve (A 1 ) were analysed with the jackknife-method. In the medium phantom, A 1 for 6 mm endoleaks was similar at 100% and 50% doses (0.93 vs. 0.86; p = 0.23), but performance at 50% dose decreased in 4–6 mm endoleaks (A 1 , 0.81 vs. 0.68; p = 0.02). In the large phantom, A 1 decreased at 50% dose both for 6 mm (0.72 vs. 0.42; p = 0.003) and for 4–6 mm (0.49 vs. 0.29; p = 0.02) endoleaks. Compared with standard reconstruction, 50 keV VMIs did not change A 1 at 25% dose (0.56 vs. 0.52; p = 0.38) and led to worse performance at 12.5% dose (0.30 vs. 0.19; p = 0.01). In summary, diagnostic performance for 6 mm endoleaks remained high with 50%-dose reduction using PCD-CT in simulated medium but not in large patients. 50 keV VMIs at 12.5% and 25% dose did not improve endoleak detection.

The International Journal of Cardiovascular Imaging
University of Bern (CH), University Hospital of Bern (CH), Luzerner Kantonsspital (CH), Spitalzentrum Centre hospitalier Biel- Bienne (CH)
Openalex Percentile: Top 18%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.