PURPOSE: Patellar resurfacing in total knee arthroplasty requires stable cement fixation of the patellar button. Unlike femoral and tibial components, the patellar button is held under manual compression until cement curing, potentially prolonging operative time. Since the patella experiences limited immediate loading after joint closure, a shorter compression time might suffice for primary fixation. This study investigated the effect of reduced manual compression time on cement morphology, micromotion and load-to-shear failure of cemented all-polyethylene patellar buttons. METHODS: Twelve human patellae from six donors were implanted with cemented all-polyethylene patellar buttons and assigned to either 30 or 600-s compression group. Micro-computed tomography (µCT) analysis assessed cement volume, penetration depth, expansion and height. Biomechanical testing included cyclic loading to determine micromotion, followed by shear testing to measure maximum failure force. Statistical comparisons and correlation analyses evaluated group differences and associations between cement morphology and biomechanical parameters. RESULTS: Cement morphology did not differ significantly between groups; volume, penetration depth, expansion and height were comparable. However, micromotion was significantly higher in the 30-s group (0.015 ± 0.009 mm) compared to the 600-s group (0.006 ± 0.003 mm, p = 0.028). The larger standard deviation in the 30-s group indicated greater variability. Maximum shear force did not differ significantly between groups (2450 ± 971 vs. 2383 ± 232 N, p = 0.857). Correlation analyses showed only limited associations between cement morphology and biomechanical outcomes. CONCLUSIONS: Reduced compression time did not significantly affect cement morphology or ultimate shear strength of cemented patellar buttons. However, shortened compression led to significantly increased micromotion, indicating impaired early interface stability. These findings suggest that micromotion under cyclic loading may be a more sensitive endpoint than maximum shear force for evaluating patellar button cementing strategies. Maintaining compression until full cement curing may optimize primary fixation stability. LEVEL OF EVIDENCE: Level V, biomechanical study.

Authors

Institutions

Publication Details

Journal
Knee Surgery Sports Traumatology Arthroscopy
Published
2026-08-03
DOI
https://doi.org/10.1002/ksa.70561
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reduced compression time is not sufficient for stability in patellar buttons

Benjamin Bartz, Leandra Bauer, Matthias Woiczinski, Matthias Brensing et al.
Knee Surgery Sports Traumatology Arthroscopy
Total Knee Arthroplasty Outcomes
article

Reduced compression time is not sufficient for stability in patellar buttons

Benjamin Bartz, Leandra Bauer, Matthias Woiczinski, Matthias Brensing, Johannes Gramzow, Georg Matziolis
article en

Abstract

PURPOSE: Patellar resurfacing in total knee arthroplasty requires stable cement fixation of the patellar button. Unlike femoral and tibial components, the patellar button is held under manual compression until cement curing, potentially prolonging operative time. Since the patella experiences limited immediate loading after joint closure, a shorter compression time might suffice for primary fixation. This study investigated the effect of reduced manual compression time on cement morphology, micromotion and load-to-shear failure of cemented all-polyethylene patellar buttons. METHODS: Twelve human patellae from six donors were implanted with cemented all-polyethylene patellar buttons and assigned to either 30 or 600-s compression group. Micro-computed tomography (µCT) analysis assessed cement volume, penetration depth, expansion and height. Biomechanical testing included cyclic loading to determine micromotion, followed by shear testing to measure maximum failure force. Statistical comparisons and correlation analyses evaluated group differences and associations between cement morphology and biomechanical parameters. RESULTS: Cement morphology did not differ significantly between groups; volume, penetration depth, expansion and height were comparable. However, micromotion was significantly higher in the 30-s group (0.015 ± 0.009 mm) compared to the 600-s group (0.006 ± 0.003 mm, p = 0.028). The larger standard deviation in the 30-s group indicated greater variability. Maximum shear force did not differ significantly between groups (2450 ± 971 vs. 2383 ± 232 N, p = 0.857). Correlation analyses showed only limited associations between cement morphology and biomechanical outcomes. CONCLUSIONS: Reduced compression time did not significantly affect cement morphology or ultimate shear strength of cemented patellar buttons. However, shortened compression led to significantly increased micromotion, indicating impaired early interface stability. These findings suggest that micromotion under cyclic loading may be a more sensitive endpoint than maximum shear force for evaluating patellar button cementing strategies. Maintaining compression until full cement curing may optimize primary fixation stability. LEVEL OF EVIDENCE: Level V, biomechanical study.

Knee Surgery Sports Traumatology Arthroscopy
Waldkrankenhaus Rudolf Elle (DE), Friedrich Schiller University Jena (DE)
Deutsche Kniegesellschaft
Openalex Percentile: Top 7%
Total Knee Arthroplasty Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.