Practical Optimization of Deep Learning Models for Cuffless Blood Pressure Estimation From Photoplethysmography

OBJECTIVE: 
To improve deep learning-based cuffless blood pressure (BP) estimation from photoplethysmography (PPG) for continuous, non-invasive monitoring when cuff measurements are impractical, by assessing sampling frequency, systolic blood pressure (SBP) and diastolic blood pressure (DBP) derivation, respiratory rate (RR) effects, and errors at BP extremes.
Approach:
We analysed 205,850 ten-second paired PPG-arterial BP segments from the Medical Information Mart for Intensive Care (MIMIC) waveform database. A convolutional model with a multiresolution U-Net (MultiResUNet) translated PPG to arterial BP waveforms. Inputs at 125, 62.5, and 31.25 Hz were compared. SBP and DBP were derived by minimum-maximum extraction or by averaging the top 4 systolic peaks and bottom 4 diastolic troughs per segment. Performance was summarized using mean absolute error (MAE).
Main Results:
Inputs at 62.5 Hz reduced MAE versus 125 Hz (SBP 5.24 vs 6.34 mmHg; DBP 2.87 vs 3.27 mmHg), whereas 31.25 Hz increased error (SBP 16.09 mmHg; DBP 9.10 mmHg). Multi-peak averaging outperformed minimum-maximum extraction. MAE varied across respiratory rates, with generally lower errors observed at moderate respiratory rates, and higher errors observed at lower and higher SBP ranges.
Significance:
Sampling frequency, SBP/DBP derivation, and RR shape error patterns in PPG-only cuffless BP estimation, with higher errors at low and high SBP. These results inform device design for bedside and ambulatory monitoring. However, further validation using subject-level datasets is required to assess clinical applicability and compliance with established validation standards.

Authors

Institutions

Publication Details

Journal
Physiological Measurement
Published
2026-08-03
DOI
https://doi.org/10.1088/1361-6579/ae944f
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Practical Optimization of Deep Learning Models for Cuffless Blood Pressure Estimation From Photoplethysmography

Vikas Khurana, Akram Khan, Ratan Bajpai, Rajesh Reddy Reddy Goutha et al.
Physiological Measurement
Non-Invasive Vital Sign Monitoring
article

Practical Optimization of Deep Learning Models for Cuffless Blood Pressure Estimation From Photoplethysmography

Vikas Khurana, Akram Khan, Ratan Bajpai, Rajesh Reddy Reddy Goutha, Mohammad Aejaz Pamidi, Bharat Bhushan Joshi Bhushan Joshi, Subodha Kumar, Ganesh Belde
article en

Abstract

OBJECTIVE: 
To improve deep learning-based cuffless blood pressure (BP) estimation from photoplethysmography (PPG) for continuous, non-invasive monitoring when cuff measurements are impractical, by assessing sampling frequency, systolic blood pressure (SBP) and diastolic blood pressure (DBP) derivation, respiratory rate (RR) effects, and errors at BP extremes.
Approach:
We analysed 205,850 ten-second paired PPG-arterial BP segments from the Medical Information Mart for Intensive Care (MIMIC) waveform database. A convolutional model with a multiresolution U-Net (MultiResUNet) translated PPG to arterial BP waveforms. Inputs at 125, 62.5, and 31.25 Hz were compared. SBP and DBP were derived by minimum-maximum extraction or by averaging the top 4 systolic peaks and bottom 4 diastolic troughs per segment. Performance was summarized using mean absolute error (MAE).
Main Results:
Inputs at 62.5 Hz reduced MAE versus 125 Hz (SBP 5.24 vs 6.34 mmHg; DBP 2.87 vs 3.27 mmHg), whereas 31.25 Hz increased error (SBP 16.09 mmHg; DBP 9.10 mmHg). Multi-peak averaging outperformed minimum-maximum extraction. MAE varied across respiratory rates, with generally lower errors observed at moderate respiratory rates, and higher errors observed at lower and higher SBP ranges.
Significance:
Sampling frequency, SBP/DBP derivation, and RR shape error patterns in PPG-only cuffless BP estimation, with higher errors at low and high SBP. These results inform device design for bedside and ambulatory monitoring. However, further validation using subject-level datasets is required to assess clinical applicability and compliance with established validation standards.

Physiological Measurement
Indian School of Business (IN), IIT Research Institute (US), Oregon Health & Science University (US), Temple College (US), The Wright Center for Graduate Medical Education (US), Temple University (US)
Openalex Percentile: Top 17%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.