Compared with conventional needle-based injections, needle-free injections offer several advantages, including ease of operation and elimination of accidental needlestick injury. To investigate the influence of piston structure on the performance of a needle-free injection system, this study combined experiments with fluid-structure interaction numerical simulations to systematically examine the effects of groove length (L), groove height (H), and the spacing (L0) between the two grooves on jet performance and piston stress. The results showed that during the initial stage of needle-free injection, both the stagnation pressure and jet velocity exhibited pronounced oscillatory behavior, and the driving structure played a significant role in modulating this oscillation pattern. For the single-groove piston, as the groove volume increased from L = 0 mm, H = 0 mm to L = 3 mm, H = 0.13 mm, the peak stagnation pressure increased by 22.3%, whereas the maximum Von Mises stress in the piston rose from 0.53 MPa to 7.21 MPa. Based on the optimal single-groove configuration, the groove was further divided into a dual-groove structure with identical groove length and height. The results indicated that, as the dualgroove spacing L0 increased from 0 to 1.5 mm, the maximum Von Mises stress first rose sharply to a peak, then decreased to approximately 3 MPa at L₀ = 0.25 mm, and remained nearly stable for L₀ > 0.25 mm. Meanwhile, the peak stagnation pressure gradually decreased with increasing groove spacing.Considering the combined variation in peak stagnation pressure and structural stress, L0 = 0.25 mm was identified as the optimal groove spacing. Compared with the optimal single-groove structure, the corresponding dual-groove configuration resulted in a reduction of less than 1% in peak stagnation pressure, while reducing the maximum Von Mises stress by 58.4%.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2026-08-03
DOI
https://doi.org/10.1088/1873-4030/ae93b5
Primary Topic
Fluid Dynamics and Heat Transfer
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of piston groove structure on the fluid dynamics of needle-free jet injectors

Hua Wen, Siming Zhou, Hongqiang Xiong, Wei Feng
Medical Engineering & Physics
Fluid Dynamics and Heat Transfer
article

Effect of piston groove structure on the fluid dynamics of needle-free jet injectors

Hua Wen, Siming Zhou, Hongqiang Xiong, Wei Feng
article en

Abstract

Compared with conventional needle-based injections, needle-free injections offer several advantages, including ease of operation and elimination of accidental needlestick injury. To investigate the influence of piston structure on the performance of a needle-free injection system, this study combined experiments with fluid-structure interaction numerical simulations to systematically examine the effects of groove length (L), groove height (H), and the spacing (L0) between the two grooves on jet performance and piston stress. The results showed that during the initial stage of needle-free injection, both the stagnation pressure and jet velocity exhibited pronounced oscillatory behavior, and the driving structure played a significant role in modulating this oscillation pattern. For the single-groove piston, as the groove volume increased from L = 0 mm, H = 0 mm to L = 3 mm, H = 0.13 mm, the peak stagnation pressure increased by 22.3%, whereas the maximum Von Mises stress in the piston rose from 0.53 MPa to 7.21 MPa. Based on the optimal single-groove configuration, the groove was further divided into a dual-groove structure with identical groove length and height. The results indicated that, as the dualgroove spacing L0 increased from 0 to 1.5 mm, the maximum Von Mises stress first rose sharply to a peak, then decreased to approximately 3 MPa at L₀ = 0.25 mm, and remained nearly stable for L₀ > 0.25 mm. Meanwhile, the peak stagnation pressure gradually decreased with increasing groove spacing.Considering the combined variation in peak stagnation pressure and structural stress, L0 = 0.25 mm was identified as the optimal groove spacing. Compared with the optimal single-groove structure, the corresponding dual-groove configuration resulted in a reduction of less than 1% in peak stagnation pressure, while reducing the maximum Von Mises stress by 58.4%.

Medical Engineering & Physics
Tongji University (CN), Nanchang University (CN), East China Normal University (CN)
Affordable and clean energy
Openalex Percentile: Top 11%
Fluid Dynamics and Heat Transfer
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.