A comparative analysis of predicted spherical refractive outcomes between two swept-source optical biometers

Abstract Purpose To compare the predicted spherical equivalent (PSEq) between the ANTERION ® anterior segment OCT device (Heidelberg Engineering GmbH, Heidelberg, Germany) and the IOLMaster 700 (Carl Zeiss Meditec AG, Jena, Germany) in patient undergoing cataract surgery. Methods Ocular biometry was performed by a single investigator for each patient scheduled for cataract surgery, with ANTERION and IOLMaster 700. Intraocular lens (IOL) power for TECNIS ® 1-Piece Aspheric IOL (Model ZCB00) and corresponding PSEq were calculated using IOL formulas including Barrett Universal II, Cooke K6, EVO 2.0, Hill-RBF 3.0, Hoffer ® QST, Kane, Pearl-DGS and SRK/T. The main outcomes assessed were the mean difference in PSEq between the two devices and the proportion of eyes where the PSEq difference exceeded 0.35 dioptres (D), considered clinically significant as it would likely result in a different IOL power being implanted. Results A total of 155 right eyes from 155 patients were analysed. The ANTERION consistently produced more hyperopic PSEq compared to the IOLMaster, with mean differences across all IOL formulas ranging from + 0.11 to + 0.18 D. These differences were statistically significant ( P < 0.05). Agreement analysis demonstrated that the majority of data points between the devices were within the 95% limits of agreement (− 0.39 D to 0.79 D), regardless of the IOL power calculation formula applied. However, 15–25% of eyes showed a PSEq difference greater than 0.35 D. Conclusions PSEq derived from ANTERION is slightly hyperopic compared to IOLMaster, with discrepancies significant enough to affect IOL power selection in up to 25% of eyes.

Authors

Institutions

Publication Details

Journal
Graefe s Archive for Clinical and Experimental Ophthalmology
Published
2026-08-03
DOI
https://doi.org/10.1007/s00417-026-07411-0
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A comparative analysis of predicted spherical refractive outcomes between two swept-source optical biometers

James McKelvie, Samir Uprety, Cameron McLintock, Saran Takemura
Graefe s Archive for Clinical and Experimental Ophthalmology
Ophthalmology and Visual Impairment Studies
article

A comparative analysis of predicted spherical refractive outcomes between two swept-source optical biometers

James McKelvie, Samir Uprety, Cameron McLintock, Saran Takemura
article en

Abstract

Abstract Purpose To compare the predicted spherical equivalent (PSEq) between the ANTERION ® anterior segment OCT device (Heidelberg Engineering GmbH, Heidelberg, Germany) and the IOLMaster 700 (Carl Zeiss Meditec AG, Jena, Germany) in patient undergoing cataract surgery. Methods Ocular biometry was performed by a single investigator for each patient scheduled for cataract surgery, with ANTERION and IOLMaster 700. Intraocular lens (IOL) power for TECNIS ® 1-Piece Aspheric IOL (Model ZCB00) and corresponding PSEq were calculated using IOL formulas including Barrett Universal II, Cooke K6, EVO 2.0, Hill-RBF 3.0, Hoffer ® QST, Kane, Pearl-DGS and SRK/T. The main outcomes assessed were the mean difference in PSEq between the two devices and the proportion of eyes where the PSEq difference exceeded 0.35 dioptres (D), considered clinically significant as it would likely result in a different IOL power being implanted. Results A total of 155 right eyes from 155 patients were analysed. The ANTERION consistently produced more hyperopic PSEq compared to the IOLMaster, with mean differences across all IOL formulas ranging from + 0.11 to + 0.18 D. These differences were statistically significant ( P < 0.05). Agreement analysis demonstrated that the majority of data points between the devices were within the 95% limits of agreement (− 0.39 D to 0.79 D), regardless of the IOL power calculation formula applied. However, 15–25% of eyes showed a PSEq difference greater than 0.35 D. Conclusions PSEq derived from ANTERION is slightly hyperopic compared to IOLMaster, with discrepancies significant enough to affect IOL power selection in up to 25% of eyes.

Graefe s Archive for Clinical and Experimental Ophthalmology
Griffith University (AU), University of Auckland (NZ), Queensland University of Technology (AU), The University of Queensland (AU), Princess Alexandra Hospital (AU), Vision Australia (AU)
University of Queensland
Openalex Percentile: Top 9%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.