A Novel Titanium End-Disc-Assisted Allograft Preparation System for Anterior Cervical Corpectomy and Fusion: Development and Preliminary Biomechanical Evaluation

Background: Allograft bone remains an attractive option for anterior cervical corpectomy and fusion (ACCF), but its use can be limited by variable graft geometry, insufficient endplate contact, and operator-dependent trimming. This study aimed to develop a titanium end-disc-assisted allograft construct with a standardized cutting and assembly system and to preliminarily evaluate its static mechanical integrity under axial compression, compression-shear, and torsional loading. Methods: The construct consisted of a controlled-length bone graft, two Ti6Al4V end-discs, and four countersunk fixation screws, forming a graft–disc construct with a maximum total height not exceeding 40 mm. The end-discs incorporated angled screw holes, bone-graft filling spaces, and enlarged contact surfaces. A modular preparation system was designed for graft clamping, length definition, guided cutting, alignment, pre-drilling, and screw fixation. Porcine rib specimens were used as an allograft substitute. Native bone and assembled constructs were tested under static compression, compression-shear, and torsion. Results: The system produced standardized constructs. Except for axial compressive stiffness, the assembled constructs showed numerically higher compressive yield force, shear stiffness, shear yield force, torsional stiffness, and yield torque than native bone; however, these findings were interpreted descriptively because of the limited sample size. Failure occurred mainly within the bone, with no obvious failure at the bone–titanium interface, titanium end-discs, or fixation screws. Conclusions: The proposed system enabled standardized graft preparation and construct assembly. Further fatigue, subsidence, and clinically relevant cervical spine model testing are required.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-08-03
DOI
https://doi.org/10.3390/bioengineering13080893
Primary Topic
Cervical and Thoracic Myelopathy
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Novel Titanium End-Disc-Assisted Allograft Preparation System for Anterior Cervical Corpectomy and Fusion: Development and Preliminary Biomechanical Evaluation

Chun‐Li Lin, Chih-Chang Chang, Pin-Cang Huang, Shao-Fu Huang
Bioengineering
Cervical and Thoracic Myelopathy
article

A Novel Titanium End-Disc-Assisted Allograft Preparation System for Anterior Cervical Corpectomy and Fusion: Development and Preliminary Biomechanical Evaluation

Chun‐Li Lin, Chih-Chang Chang, Pin-Cang Huang, Shao-Fu Huang
article en

Abstract

Background: Allograft bone remains an attractive option for anterior cervical corpectomy and fusion (ACCF), but its use can be limited by variable graft geometry, insufficient endplate contact, and operator-dependent trimming. This study aimed to develop a titanium end-disc-assisted allograft construct with a standardized cutting and assembly system and to preliminarily evaluate its static mechanical integrity under axial compression, compression-shear, and torsional loading. Methods: The construct consisted of a controlled-length bone graft, two Ti6Al4V end-discs, and four countersunk fixation screws, forming a graft–disc construct with a maximum total height not exceeding 40 mm. The end-discs incorporated angled screw holes, bone-graft filling spaces, and enlarged contact surfaces. A modular preparation system was designed for graft clamping, length definition, guided cutting, alignment, pre-drilling, and screw fixation. Porcine rib specimens were used as an allograft substitute. Native bone and assembled constructs were tested under static compression, compression-shear, and torsion. Results: The system produced standardized constructs. Except for axial compressive stiffness, the assembled constructs showed numerically higher compressive yield force, shear stiffness, shear yield force, torsional stiffness, and yield torque than native bone; however, these findings were interpreted descriptively because of the limited sample size. Failure occurred mainly within the bone, with no obvious failure at the bone–titanium interface, titanium end-discs, or fixation screws. Conclusions: The proposed system enabled standardized graft preparation and construct assembly. Further fatigue, subsidence, and clinically relevant cervical spine model testing are required.

BioengineeringVol. 13(8)
National Yang Ming Chiao Tung University (TW), Taipei Veterans General Hospital (TW)
Ministry of Science and Technology, Taiwan
Openalex Percentile: Top 7%
Cervical and Thoracic Myelopathy
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies