Serum versus lithium-heparin plasma for lipid testing: matrix-related bias and calculated LDL-C classification

AIMS: To assess whether serum and lithium-heparin plasma can be used interchangeably for routine lipid testing and whether matrix-related bias affects calculated low-density lipoprotein (LDL)-cholesterol (LDL-C) classification near clinical decision thresholds. METHODS: This prospective paired-sample method-comparison study included 104 fasting adult ambulatory patients referred for routine lipid testing. Serum and lithium-heparin plasma were collected during the same venipuncture and analysed on a Cobas 6000 c501 analyser. LDL-C was calculated using the Friedewald equation when triglycerides (TGs) were <4.5 mmol/L (n=103). Agreement was assessed using paired testing, Passing-Bablok regression, Bland-Altman analysis, intraclass correlation coefficients (ICCs), Cohen kappa and McNemar testing at selected decision thresholds. Interchangeability was interpreted using analytical bias, measurement uncertainty, limits of agreement and threshold discordance. RESULTS: Lithium-heparin plasma yielded lower total cholesterol and calculated LDL-C than serum, with mean biases (plasma - serum) of -0.15 mmol/L (-3.06%) and -0.14 mmol/L (-4.77%), respectively. TGs and HDL-cholesterol did not differ significantly. Serum-plasma correlations and ICCs were high. LDL-C classification was discordant in 3/103, 2/103 and 7/103 paired samples at thresholds of 1.4, 1.8 and 2.6 mmol/L, respectively; all discordances classified plasma below and serum above the threshold. CONCLUSIONS: Under these local analytical and pre-analytical conditions, the mean bias was modest but directional and sufficient to alter calculated LDL-C classification in a minority of samples near decision thresholds. Laboratories using lithium-heparin plasma for lipid testing should verify matrix agreement locally, document the specimen matrix and avoid alternating serum and plasma during longitudinal monitoring unless equivalence has been demonstrated.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Pathology
Published
2026-07-30
DOI
https://doi.org/10.1136/jcp-2026-210814
Primary Topic
Diabetes, Cardiovascular Risks, and Lipoproteins
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Serum versus lithium-heparin plasma for lipid testing: matrix-related bias and calculated LDL-C classification

Aouatef Jmal, L. Jmal, Khédija Zaouche, Raida GUIDARA et al.
Journal of Clinical Pathology
Diabetes, Cardiovascular Risks, and Lipoproteins
article

Serum versus lithium-heparin plasma for lipid testing: matrix-related bias and calculated LDL-C classification

Aouatef Jmal, L. Jmal, Khédija Zaouche, Raida GUIDARA, Nejd Kouched, Amel Dhieb
article en

Abstract

AIMS: To assess whether serum and lithium-heparin plasma can be used interchangeably for routine lipid testing and whether matrix-related bias affects calculated low-density lipoprotein (LDL)-cholesterol (LDL-C) classification near clinical decision thresholds. METHODS: This prospective paired-sample method-comparison study included 104 fasting adult ambulatory patients referred for routine lipid testing. Serum and lithium-heparin plasma were collected during the same venipuncture and analysed on a Cobas 6000 c501 analyser. LDL-C was calculated using the Friedewald equation when triglycerides (TGs) were <4.5 mmol/L (n=103). Agreement was assessed using paired testing, Passing-Bablok regression, Bland-Altman analysis, intraclass correlation coefficients (ICCs), Cohen kappa and McNemar testing at selected decision thresholds. Interchangeability was interpreted using analytical bias, measurement uncertainty, limits of agreement and threshold discordance. RESULTS: Lithium-heparin plasma yielded lower total cholesterol and calculated LDL-C than serum, with mean biases (plasma - serum) of -0.15 mmol/L (-3.06%) and -0.14 mmol/L (-4.77%), respectively. TGs and HDL-cholesterol did not differ significantly. Serum-plasma correlations and ICCs were high. LDL-C classification was discordant in 3/103, 2/103 and 7/103 paired samples at thresholds of 1.4, 1.8 and 2.6 mmol/L, respectively; all discordances classified plasma below and serum above the threshold. CONCLUSIONS: Under these local analytical and pre-analytical conditions, the mean bias was modest but directional and sufficient to alter calculated LDL-C classification in a minority of samples near decision thresholds. Laboratories using lithium-heparin plasma for lipid testing should verify matrix agreement locally, document the specimen matrix and avoid alternating serum and plasma during longitudinal monitoring unless equivalence has been demonstrated.

Journal of Clinical Pathology
Tunis University (TN), Military Hospital of Tunis (TN), Faculté de médecine de Tunis, Tunis El Manar University (TN)
Peace, Justice and strong institutions
Openalex Percentile: Top 9%
Diabetes, Cardiovascular Risks, and Lipoproteins
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.