Assessment of Ultraviolet-C Inactivation of Pseudomonas aeruginosa and Aspergillus flavus in the Anesthesia Workspace

Objective To determine the ultraviolet-C (UV-C) irradiation doses required to inactivate Pseudomonas aeruginosa and Aspergillus flavus contamination on polycarbonate surfaces using a triangular emitter configuration. Patients and Methods In this laboratory experiment, polycarbonate test carriers inoculated with >10 8 colony-forming units of P aeruginosa (MYA-15442) or A flavus (American Type Culture Collection 9643) were positioned 2.74 m from 3 linear UV-C emitters to determine minimally effective doses. Ten independent irradiation doses were evaluated for P aeruginosa (25-250 mJ/cm 2 ) and A flavus (50-2000 mJ/cm 2 ). Larger carriers were additionally evaluated for A flavus to model reduced cell clumping, simulating prior surface disinfection. Residual colony-forming units were quantified and compared with untreated controls to calculate log reductions. Finally, the efficacy of delivering these minimally effective doses through a triangular UV-C emitter configuration was assessed. Results For P aeruginosa , no growth was detected at any dose, including 25 mJ/cm 2 . For A flavus on standard carriers, inactivation occurred at 725 mJ/cm 2 , though no growth conditions were not achieved at any tested dose. However, on larger carriers (reduced clumping), no growth conditions for A flavus were achieved at 650 mJ/cm 2 . Delivery of 25 mJ/cm 2 for P aeruginosa and 650 mJ/cm 2 for A flavus through the triangular configuration resulted in undetectable growth. Conclusion A UV-C dose of 25 mJ/cm 2 is sufficient to attenuate P aeruginosa surface contamination. Conversely, A flavus inactivation will likely require manual high-level surface disinfection augmented with 650 mJ/cm 2 of UV-C irradiation.

Authors

Institutions

Publication Details

Journal
Mayo Clinic Proceedings Innovations Quality & Outcomes
Published
2026-07-28
DOI
https://doi.org/10.1016/j.mayocpiqo.2026.100740
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of Ultraviolet-C Inactivation of Pseudomonas aeruginosa and Aspergillus flavus in the Anesthesia Workspace

Jonathan E. Charnin, Michelle C. Parra, Brendan T. Wanta, Chase P. Loftus et al.
Mayo Clinic Proceedings Innovations Quality & Outcomes
Medical Device Sterilization and Disinfection
article

Assessment of Ultraviolet-C Inactivation of Pseudomonas aeruginosa and Aspergillus flavus in the Anesthesia Workspace

Jonathan E. Charnin, Michelle C. Parra, Brendan T. Wanta, Chase P. Loftus, Randy W. Loftus, Carmen T. Brindeiro, Cornelie CL. Banguid, Franklin Dexter
article en

Abstract

Objective To determine the ultraviolet-C (UV-C) irradiation doses required to inactivate Pseudomonas aeruginosa and Aspergillus flavus contamination on polycarbonate surfaces using a triangular emitter configuration. Patients and Methods In this laboratory experiment, polycarbonate test carriers inoculated with >10 8 colony-forming units of P aeruginosa (MYA-15442) or A flavus (American Type Culture Collection 9643) were positioned 2.74 m from 3 linear UV-C emitters to determine minimally effective doses. Ten independent irradiation doses were evaluated for P aeruginosa (25-250 mJ/cm 2 ) and A flavus (50-2000 mJ/cm 2 ). Larger carriers were additionally evaluated for A flavus to model reduced cell clumping, simulating prior surface disinfection. Residual colony-forming units were quantified and compared with untreated controls to calculate log reductions. Finally, the efficacy of delivering these minimally effective doses through a triangular UV-C emitter configuration was assessed. Results For P aeruginosa , no growth was detected at any dose, including 25 mJ/cm 2 . For A flavus on standard carriers, inactivation occurred at 725 mJ/cm 2 , though no growth conditions were not achieved at any tested dose. However, on larger carriers (reduced clumping), no growth conditions for A flavus were achieved at 650 mJ/cm 2 . Delivery of 25 mJ/cm 2 for P aeruginosa and 650 mJ/cm 2 for A flavus through the triangular configuration resulted in undetectable growth. Conclusion A UV-C dose of 25 mJ/cm 2 is sufficient to attenuate P aeruginosa surface contamination. Conversely, A flavus inactivation will likely require manual high-level surface disinfection augmented with 650 mJ/cm 2 of UV-C irradiation.

Mayo Clinic Proceedings Innovations Quality & OutcomesVol. 10(5)
University of Iowa (US), Mayo Clinic in Arizona (US), Bioinformatics Institute (RU)
Openalex Percentile: Top 10%
Medical Device Sterilization and Disinfection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.