Comparative thermal analysis of titanium and zirconia implants with various depths of insertion in different bone types: a 3D finite element analysis

Abstract Background Thermal changes in the oral environment can influence dental implant performance by affecting the surrounding bone. Titanium remains the most widely used implant material due to its strength and predictable osseointegration. However, zirconia has gained attention because of its lower thermal conductivity and favourable biological properties. Despite this, comparative evidence under clinically relevant thermal conditions remains limited. Aim To evaluate and compare the thermal behaviour of titanium and zirconia dental implants under different thermal loads, insertion depths, and bone types using three-dimensional finite element analysis (FEA). Materials and methods A three-dimensional finite element model including the implant, abutment, surrounding bone, and mucosa was developed. Two implant materials (titanium and zirconia), two bone types (D1 and D2), and two insertion depths (0 mm and 1 mm) were analysed. Thermal loads ranging from 5 °C to 55 °C were applied at the implant surface, with a baseline bone temperature of 35 °C and a vertical occlusal load of 140 N. Mesh convergence resulted in a model with 242,118 elements. Temperature distribution at the bone–implant interface was evaluated. As this was a deterministic simulation study, comparisons were based on numerical differences. Results Temperature increased across all models with increasing thermal load and insertion depth. Titanium implants showed greater heat transfer to surrounding bone. At 55 °C, titanium exceeded the critical threshold of 47 °C in D1 bone at 0 mm insertion (bone: 50.16 °C; interface: 50.04 °C). Zirconia generally showed lower thermal transmission and remained below the threshold in several conditions, particularly in D2 bone. However, under certain conditions, such as deeper insertion in D1 bone, zirconia showed interface temperatures comparable to or slightly higher than titanium. Conclusion Within the limitations of this study, zirconia implants generally reduced heat transfer compared to titanium, although this effect depended on specific conditions. Implant material and insertion depth may influence thermal exposure at the bone–implant interface. Further experimental and clinical validation is required.

Authors

Institutions

Publication Details

Journal
BMC Oral Health
Published
2026-07-28
DOI
https://doi.org/10.1186/s12903-026-09256-9
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative thermal analysis of titanium and zirconia implants with various depths of insertion in different bone types: a 3D finite element analysis

Srikanth Gadicherla, Sandipan Mukherjee, Lipika Sen, Sayeed Fatma Hasan et al.
BMC Oral Health
Dental Implant Techniques and Outcomes
article

Comparative thermal analysis of titanium and zirconia implants with various depths of insertion in different bone types: a 3D finite element analysis

Srikanth Gadicherla, Sandipan Mukherjee, Lipika Sen, Sayeed Fatma Hasan, Pradyumn Bhat, Rohan Sharma
article en

Abstract

Abstract Background Thermal changes in the oral environment can influence dental implant performance by affecting the surrounding bone. Titanium remains the most widely used implant material due to its strength and predictable osseointegration. However, zirconia has gained attention because of its lower thermal conductivity and favourable biological properties. Despite this, comparative evidence under clinically relevant thermal conditions remains limited. Aim To evaluate and compare the thermal behaviour of titanium and zirconia dental implants under different thermal loads, insertion depths, and bone types using three-dimensional finite element analysis (FEA). Materials and methods A three-dimensional finite element model including the implant, abutment, surrounding bone, and mucosa was developed. Two implant materials (titanium and zirconia), two bone types (D1 and D2), and two insertion depths (0 mm and 1 mm) were analysed. Thermal loads ranging from 5 °C to 55 °C were applied at the implant surface, with a baseline bone temperature of 35 °C and a vertical occlusal load of 140 N. Mesh convergence resulted in a model with 242,118 elements. Temperature distribution at the bone–implant interface was evaluated. As this was a deterministic simulation study, comparisons were based on numerical differences. Results Temperature increased across all models with increasing thermal load and insertion depth. Titanium implants showed greater heat transfer to surrounding bone. At 55 °C, titanium exceeded the critical threshold of 47 °C in D1 bone at 0 mm insertion (bone: 50.16 °C; interface: 50.04 °C). Zirconia generally showed lower thermal transmission and remained below the threshold in several conditions, particularly in D2 bone. However, under certain conditions, such as deeper insertion in D1 bone, zirconia showed interface temperatures comparable to or slightly higher than titanium. Conclusion Within the limitations of this study, zirconia implants generally reduced heat transfer compared to titanium, although this effect depended on specific conditions. Implant material and insertion depth may influence thermal exposure at the bone–implant interface. Further experimental and clinical validation is required.

BMC Oral Health
Manipal Academy of Higher Education (IN)
Openalex Percentile: Top 7%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.