Left arm twitching caused by failure of pacemaker lead insulation

An 89-year-old woman with dementia was referred to our neurology clinic after presenting to her family physician with progressively worsening involuntary movements over 1 month.She had undergone permanent pacemaker implantation for complete atrioventricular block 14 years earlier, with generator replacement 3 years before presentation.Routine device follow-up had not been performed after she moved into a long-term care facility.On examination, we observed intermittent left arm twitching at approximately 1 Hz (Appendix 1, Video 1, available at www.cmaj.ca/lookup/doi/10.1503/cmaj.260045/tab-related-content).We considered myoclonus, fasciculation, and tremor, but these were inconsistent with the regular, coarse, low-frequency movements.The movements were simultaneous with pacing on mon itor electrocardiography, suggesting pacemaker involvement.Chest radiography revealed an abnormal configuration of the atrial pacing lead (Figure 1).We diagnosed pacemaker lead malfunction caused by insulation failure.Reprogramming the pacemaker mode from dual-chamber (DDD) to ventricular (VVI) pacing

Authors

Institutions

Publication Details

Journal
Canadian Medical Association Journal
Published
2026-07-27
DOI
https://doi.org/10.1503/cmaj.260045
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Left arm twitching caused by failure of pacemaker lead insulation

Satoshi Suda, Shunsuke Kido, Junya Aoki
Canadian Medical Association Journal
Cardiac pacing and defibrillation studies
article

Left arm twitching caused by failure of pacemaker lead insulation

Satoshi Suda, Shunsuke Kido, Junya Aoki
article en

Abstract

An 89-year-old woman with dementia was referred to our neurology clinic after presenting to her family physician with progressively worsening involuntary movements over 1 month.She had undergone permanent pacemaker implantation for complete atrioventricular block 14 years earlier, with generator replacement 3 years before presentation.Routine device follow-up had not been performed after she moved into a long-term care facility.On examination, we observed intermittent left arm twitching at approximately 1 Hz (Appendix 1, Video 1, available at www.cmaj.ca/lookup/doi/10.1503/cmaj.260045/tab-related-content).We considered myoclonus, fasciculation, and tremor, but these were inconsistent with the regular, coarse, low-frequency movements.The movements were simultaneous with pacing on mon itor electrocardiography, suggesting pacemaker involvement.Chest radiography revealed an abnormal configuration of the atrial pacing lead (Figure 1).We diagnosed pacemaker lead malfunction caused by insulation failure.Reprogramming the pacemaker mode from dual-chamber (DDD) to ventricular (VVI) pacing

Canadian Medical Association JournalVol. 198(27)
Saitama International Medical Center (JP)
Affordable and clean energy
Openalex Percentile: Top 9%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.