Dosimetric evaluation of the RSCAN PP20 prototype plane-parallel ionization chamber in electron and kilovoltage x-ray beams

The RSCAN PP20 prototype plane-parallel ionization chamber was evaluated for electron beam dosimetry and medium-energy kilovoltage X-ray beam measurements. This study investigated its response constancy, linearity, depth-ionization characteristics, voltage and polarity dependence, and signal stability in water using a 6 MeV electron beam and selected kilovoltage X-ray beam qualities. Additional time-resolved measurements were performed to assess dark-field charge transients, post-irradiation charge relaxation, in-beam current transients, and cable-related parasitic currents, with a PTW Advanced Markus chamber used as a reference-control detector. In the 6 MeV electron beam, the RSCAN PP20 prototype produced physically plausible depth-ionization curves but showed insufficient quantitative stability. The coefficient of variation reached 1.57%, and the relative range reached 5.10%. The apparent linearity deviation reached 16.8% when background compensation was performed only once per measurement series, but decreased to 0.64% when repeated before each measurement. A polarity correction factor of approximately 0.91 was obtained across the tested voltage range, indicating measurement-chain asymmetry rather than a conventional intrinsic polarity effect. Time-resolved recordings revealed dark-field charge transients, post-beam charge accumulation, and mechanically induced bipolar current spikes, consistent with parasitic currents and possible cable-related triboelectric effects. Fixed-depth measurements in kilovoltage X-ray beams showed better short-term stability, whereas depth-response scans demonstrated cable-dependent distortions. The main limitation of the current RSCAN PP20 prototype is not its basic plane-parallel geometry, which is close to established clinical designs, but instability of the prototype chamber-cable assembly within the tested measurement setup. The prototype is not yet suitable for routine clinical dosimetry in its current configuration and requires hardware improvements followed by comprehensive recharacterization.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-07-27
DOI
https://doi.org/10.1088/2057-1976/ae90b0
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dosimetric evaluation of the RSCAN PP20 prototype plane-parallel ionization chamber in electron and kilovoltage x-ray beams

A. Yu. Smyslov, V A Savelev, A N Moiseev, A V Anan'ina et al.
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

Dosimetric evaluation of the RSCAN PP20 prototype plane-parallel ionization chamber in electron and kilovoltage x-ray beams

A. Yu. Smyslov, V A Savelev, A N Moiseev, A V Anan'ina, A A Kokontsev
article en

Abstract

The RSCAN PP20 prototype plane-parallel ionization chamber was evaluated for electron beam dosimetry and medium-energy kilovoltage X-ray beam measurements. This study investigated its response constancy, linearity, depth-ionization characteristics, voltage and polarity dependence, and signal stability in water using a 6 MeV electron beam and selected kilovoltage X-ray beam qualities. Additional time-resolved measurements were performed to assess dark-field charge transients, post-irradiation charge relaxation, in-beam current transients, and cable-related parasitic currents, with a PTW Advanced Markus chamber used as a reference-control detector. In the 6 MeV electron beam, the RSCAN PP20 prototype produced physically plausible depth-ionization curves but showed insufficient quantitative stability. The coefficient of variation reached 1.57%, and the relative range reached 5.10%. The apparent linearity deviation reached 16.8% when background compensation was performed only once per measurement series, but decreased to 0.64% when repeated before each measurement. A polarity correction factor of approximately 0.91 was obtained across the tested voltage range, indicating measurement-chain asymmetry rather than a conventional intrinsic polarity effect. Time-resolved recordings revealed dark-field charge transients, post-beam charge accumulation, and mechanically induced bipolar current spikes, consistent with parasitic currents and possible cable-related triboelectric effects. Fixed-depth measurements in kilovoltage X-ray beams showed better short-term stability, whereas depth-response scans demonstrated cable-dependent distortions. The main limitation of the current RSCAN PP20 prototype is not its basic plane-parallel geometry, which is close to established clinical designs, but instability of the prototype chamber-cable assembly within the tested measurement setup. The prototype is not yet suitable for routine clinical dosimetry in its current configuration and requires hardware improvements followed by comprehensive recharacterization.

Biomedical Physics & Engineering Express
Federal State Budgetary Institution Russian Scientific Center of Roentgenoradiology (RU), Covance (United States) (US)
Affordable and clean energy
Openalex Percentile: Top 10%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.