First imaging of secondary electron bremsstrahlung x-rays emitted from a patient during proton beam irradiation

Abstract Objective. A low-energy x-rays camera–based system has been introduced to image secondary electron bremsstrahlung x-rays, enabling external visualization of proton beam profiles during irradiation. To date, this approach has been demonstrated only in phantom studies and simulations, and has not yet been successfully applied in patients. To overcome this limitation, it is essential to establish the feasibility of SEB x-ray imaging in a clinical condition. Approach. We first optimized the energy window and threshold settings for range estimation by acquiring SEB x-ray images of a water phantom comparable in size to the human lower abdomen, using a cerium doped yttrium aluminum perovskite (YAP(Ce)) x-ray camera with a pinhole collimator. We then performed the first in vivo SEB x-ray measurement by imaging proton beams during treatment of a patient of prostate cancer with two opposed fields, each delivering an RBE-weighted dose of 1.5 Gy. Main results. In the patient study, SEB x-ray distributions were observed in the measured images using the same energy window as that established in the water phantom study. From these images, the beam ranges were estimated to be within approximately 1 cm of the values calculated by the treatment planning system using the optimized profile threshold. Dynamic imaging at 1 s time intervals was achieved, enabling temporal evaluation of count rate curves. Significance. We present, for the first time, measurements of SEB x-ray images acquired in a patient during proton beam irradiation. These preliminary results indicate that in vivo SEB x-ray imaging could be a promising approach for clinical real-time beam monitoring in proton therapy. Ethics approval. Ethics approval was obtained from the institutional review board of Kobe Proton Center: approval number: 06-10.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-07-27
DOI
https://doi.org/10.1088/1361-6560/ae88b0
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

First imaging of secondary electron bremsstrahlung x-rays emitted from a patient during proton beam irradiation

Katia Parodi, Takuya Yabe, Mitsutaka Yamaguchi, Seiichi Yamamoto et al.
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
article

First imaging of secondary electron bremsstrahlung x-rays emitted from a patient during proton beam irradiation

Katia Parodi, Takuya Yabe, Mitsutaka Yamaguchi, Seiichi Yamamoto, J Kataoka, A Yoshikawa, Tomohiro Yamashita, Naoki Kawachi, Kei Kamada
article en

Abstract

Abstract Objective. A low-energy x-rays camera–based system has been introduced to image secondary electron bremsstrahlung x-rays, enabling external visualization of proton beam profiles during irradiation. To date, this approach has been demonstrated only in phantom studies and simulations, and has not yet been successfully applied in patients. To overcome this limitation, it is essential to establish the feasibility of SEB x-ray imaging in a clinical condition. Approach. We first optimized the energy window and threshold settings for range estimation by acquiring SEB x-ray images of a water phantom comparable in size to the human lower abdomen, using a cerium doped yttrium aluminum perovskite (YAP(Ce)) x-ray camera with a pinhole collimator. We then performed the first in vivo SEB x-ray measurement by imaging proton beams during treatment of a patient of prostate cancer with two opposed fields, each delivering an RBE-weighted dose of 1.5 Gy. Main results. In the patient study, SEB x-ray distributions were observed in the measured images using the same energy window as that established in the water phantom study. From these images, the beam ranges were estimated to be within approximately 1 cm of the values calculated by the treatment planning system using the optimized profile threshold. Dynamic imaging at 1 s time intervals was achieved, enabling temporal evaluation of count rate curves. Significance. We present, for the first time, measurements of SEB x-ray images acquired in a patient during proton beam irradiation. These preliminary results indicate that in vivo SEB x-ray imaging could be a promising approach for clinical real-time beam monitoring in proton therapy. Ethics approval. Ethics approval was obtained from the institutional review board of Kobe Proton Center: approval number: 06-10.

Physics in Medicine and BiologyVol. 71(14)
Proton (Malaysia) (MY), Waseda University (JP), Tohoku University (JP), National Institutes for Quantum Science and Technology (JP), Ludwig-Maximilians-Universität München (DE)
Openalex Percentile: Top 10%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.