Clinical evaluation of deep learning–accelerated 3D FLAIR imaging at 1.5T and 3T for visualization of cerebral white matter lesions

Background Fluid-attenuated inversion recovery (FLAIR) imaging is a standard sequence used for the detection of white matter lesions; however, 1-mm isotropic FLAIR can take 5–7 min at 1.5T and 3T using routine methods, limiting routine use of high-resolution 3D FLAIR imaging. Purpose To evaluate a deep learning (DL)–based unrolled optimization 3D FLAIR technique, Sonic DL™ 3D (SDL), in comparison to a standard 3D FLAIR protocol, both combined with DL-based post-processing (AIR™ Recon DL) at 1.5T and 3T. Material and Methods A total of 67 patients (15 scanned at 1.5T and 52 at 3T) with suspected white matter lesions were included. Two neuroradiologists performed blinded, head-to-head image quality evaluation (motion, noise, and overall diagnostic quality). Quantitative lesion volume and lesion counts were derived from automated segmentation using SPM 12. Cohen's kappa, Wilcoxon signed-rank tests, and paired t -tests were applied. Results At 1.5T, SDL FLAIR (4 min 53 s) reduced scan times by 15% compared to standard 3D FLAIR (5 min 43 s) and performed similarly or better in lesion visualization and image quality, with significantly ( P <0.05) improved noise quality. At 3T, SDL FLAIR (2 min 45 s) achieved a reduction in scan time of approximately 35% compared to standard 3D FLAIR (4 min 15 s) and showed improved lesion conspicuity (periventricular, juxtacortical, and other regions) and image quality metrics. However, except for noise quality, these differences did not reach statistical significance despite moderate to almost perfect interrater agreement (Cohen's kappa 0.55–0.85; P <0.001). Conclusion Accelerated SDL FLAIR demonstrated improved diagnostic quality, comparable lesion conspicuity, and volumetric measurements to standard 3D FLAIR sequence at 1.5T and 3T.

Authors

Institutions

Publication Details

Journal
Acta Radiologica
Published
2026-07-27
DOI
https://doi.org/10.1177/02841851261460842
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical evaluation of deep learning–accelerated 3D FLAIR imaging at 1.5T and 3T for visualization of cerebral white matter lesions

Ivan Jambor, Chiang Ch, Alexandra Filippakis, Susie Y. Huang et al.
Acta Radiologica
Advanced MRI Techniques and Applications
article

Clinical evaluation of deep learning–accelerated 3D FLAIR imaging at 1.5T and 3T for visualization of cerebral white matter lesions

Ivan Jambor, Chiang Ch, Alexandra Filippakis, Susie Y. Huang, Sittaya Buathong, Mary Thomas, E Milshteyn, D Rettmann, Christopher Hemond
article en

Abstract

Background Fluid-attenuated inversion recovery (FLAIR) imaging is a standard sequence used for the detection of white matter lesions; however, 1-mm isotropic FLAIR can take 5–7 min at 1.5T and 3T using routine methods, limiting routine use of high-resolution 3D FLAIR imaging. Purpose To evaluate a deep learning (DL)–based unrolled optimization 3D FLAIR technique, Sonic DL™ 3D (SDL), in comparison to a standard 3D FLAIR protocol, both combined with DL-based post-processing (AIR™ Recon DL) at 1.5T and 3T. Material and Methods A total of 67 patients (15 scanned at 1.5T and 52 at 3T) with suspected white matter lesions were included. Two neuroradiologists performed blinded, head-to-head image quality evaluation (motion, noise, and overall diagnostic quality). Quantitative lesion volume and lesion counts were derived from automated segmentation using SPM 12. Cohen's kappa, Wilcoxon signed-rank tests, and paired t -tests were applied. Results At 1.5T, SDL FLAIR (4 min 53 s) reduced scan times by 15% compared to standard 3D FLAIR (5 min 43 s) and performed similarly or better in lesion visualization and image quality, with significantly ( P <0.05) improved noise quality. At 3T, SDL FLAIR (2 min 45 s) achieved a reduction in scan time of approximately 35% compared to standard 3D FLAIR (4 min 15 s) and showed improved lesion conspicuity (periventricular, juxtacortical, and other regions) and image quality metrics. However, except for noise quality, these differences did not reach statistical significance despite moderate to almost perfect interrater agreement (Cohen's kappa 0.55–0.85; P <0.001). Conclusion Accelerated SDL FLAIR demonstrated improved diagnostic quality, comparable lesion conspicuity, and volumetric measurements to standard 3D FLAIR sequence at 1.5T and 3T.

Acta Radiologica
National Institutes of Health (US), Siriraj Hospital (TH), Harvard University (US), University of Turku (FI), University of Massachusetts Chan Medical School (US), New England College (US), General Electric (Spain) (ES), Mahidol University (TH), Massachusetts General Hospital (US), Memorial Medical Center (US), National Institute of Neurological Disorders and Stroke (US), Athinoula A. Martinos Center for Biomedical Imaging (US), Wentworth Douglass Hospital (US), Taipei Medical University-Shuang Ho Hospital (TW), UMass Memorial Medical Center (US), Taipei Medical University (TW)
Openalex Percentile: Top 9%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.