Repeatability of Epithelial Thickness Mapping in Normal and Keratoconus Eyes Using a Novel Hybrid Scheimpflug—Optical Coherence Tomography Device

PURPOSE: A novel diagnostic device combining Scheimpflug and anterior segment optical coherence tomography promises increased accuracy in corneal diagnostics. The aim of this study was to analyze the epithelial thickness measurement repeatability in healthy and keratoconus eyes and define severity-dependent thresholds. METHODS: This prospective, cross-sectional, observational, monocentric study included a total of 217 eyes of 217 subjects divided into a keratoconus group (166 eyes of 166 subjects) and a healthy control group (51 eyes of 51 subjects). Keratoconus diagnosis and stage was established using multimodal imaging. Three measurements were performed with the novel combined Scheimpflug and optical coherence tomography device (Pentacam Cornea optical coherence tomography) on the same day. Repeatability coefficients (RCs) with two-sided 95% confidence intervals (CIs) were calculated for epithelial parameters. The repeatability was compared between groups and according to keratoconus severity and crosslinking status. RESULTS: The median repeatability of the epithelial thickness was slightly better in the healthy group (RC = 3.3) compared with the keratoconus group (RC = 3.9; P < 0.001) when the tear film was excluded. The overall median repeatability with (RC = 3.7) and without tear film (RC = 3.8) did not differ significantly (P > 0.05). Repeatability worsened significantly in severe keratoconus stages (P < 0.01), whereas crosslinking status had only a marginal effect. CONCLUSIONS: The hybrid tomographic device demonstrated high intrasubject repeatability for epithelial thickness assessment in healthy and keratoconus eyes, supporting its technical feasibility for clinical use. Further studies are needed to analyze the benefit of incorporating epithelium into established keratoconus scores using the hybrid system.

Authors

Institutions

Publication Details

Journal
Cornea
Published
2026-07-27
DOI
https://doi.org/10.1097/ico.0000000000004242
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Repeatability of Epithelial Thickness Mapping in Normal and Keratoconus Eyes Using a Novel Hybrid Scheimpflug—Optical Coherence Tomography Device

Hyeck-Soo Son, Maximilian Friedrich, Gerd Uwe Auffarth, Victor Aristide Augustin et al.
Cornea
Corneal surgery and disorders
article

Repeatability of Epithelial Thickness Mapping in Normal and Keratoconus Eyes Using a Novel Hybrid Scheimpflug—Optical Coherence Tomography Device

Hyeck-Soo Son, Maximilian Friedrich, Gerd Uwe Auffarth, Victor Aristide Augustin, Emilia Felter, Julia Keim, Luisa Finken
article en

Abstract

PURPOSE: A novel diagnostic device combining Scheimpflug and anterior segment optical coherence tomography promises increased accuracy in corneal diagnostics. The aim of this study was to analyze the epithelial thickness measurement repeatability in healthy and keratoconus eyes and define severity-dependent thresholds. METHODS: This prospective, cross-sectional, observational, monocentric study included a total of 217 eyes of 217 subjects divided into a keratoconus group (166 eyes of 166 subjects) and a healthy control group (51 eyes of 51 subjects). Keratoconus diagnosis and stage was established using multimodal imaging. Three measurements were performed with the novel combined Scheimpflug and optical coherence tomography device (Pentacam Cornea optical coherence tomography) on the same day. Repeatability coefficients (RCs) with two-sided 95% confidence intervals (CIs) were calculated for epithelial parameters. The repeatability was compared between groups and according to keratoconus severity and crosslinking status. RESULTS: The median repeatability of the epithelial thickness was slightly better in the healthy group (RC = 3.3) compared with the keratoconus group (RC = 3.9; P < 0.001) when the tear film was excluded. The overall median repeatability with (RC = 3.7) and without tear film (RC = 3.8) did not differ significantly (P > 0.05). Repeatability worsened significantly in severe keratoconus stages (P < 0.01), whereas crosslinking status had only a marginal effect. CONCLUSIONS: The hybrid tomographic device demonstrated high intrasubject repeatability for epithelial thickness assessment in healthy and keratoconus eyes, supporting its technical feasibility for clinical use. Further studies are needed to analyze the benefit of incorporating epithelium into established keratoconus scores using the hybrid system.

Cornea
Heidelberg University (DE), Augenklinik Heidelberg (DE), Apple (Germany) (DE)
Openalex Percentile: Top 9%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.