A New Faecal Immunochemical Test Buffer With Improved Haemoglobin Stability Under Varying Temperature and Storage Times

BACKGROUND: Measurement of faecal haemoglobin (f-Hb) using the faecal immunochemical test (FIT) in population screening for colorectal cancer (CRC) has reduced cancer-related incidence and mortality. However, temperature-induced f-Hb degradation can decrease FIT accuracy. This study assessed the f-Hb preservation capacity of a new FIT Hb-stabilising buffer at different temperatures and durations. METHODS: Fresh whole faecal samples from individuals with active colorectal bleeding were used for collection of samples using the OC-Sensor FIT (EIKEN CHEMICAL CO. LTD, Japan) containing either the standard (SOC3) or new formulation (SOC4) buffer. Pooled faeces-buffer mixtures (n = 29) were divided into aliquots for incubation at 4°C, 23°C, 35°C, 45°C, and 50°C and analysed for f-Hb concentration at days 0, 3, 7, 10, 14, 21, and 28. Relative f-Hb (percentage change of day 0) in both buffer types was compared between temperatures at each timepoint. Generalised linear modelling identified variables influencing f-Hb degradation. RESULTS: SOC3 exhibited significant f-Hb decline from day 10 at 23°C and from day 3 at all higher temperatures assessed (p < 0.05). In contrast, f-Hb concentrations remained stable within SOC4 for up to 28 days at 23°C, with reductions observed after 7, 7, and 14 days at 50°C, 45°C, and 35°C, respectively (p < 0.05). Higher temperatures and longer storage duration significantly decreased f-Hb concentration, while SOC4 demonstrated superior f-Hb preservation relative to SOC3 (generalised linear model estimate = 15.4, p < 0.001). CONCLUSION: The SOC4 FIT buffer improves f-Hb stability compared with the current formulation and may improve CRC screening accuracy, particularly in high-temperature environments.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Laboratory Analysis
Published
2026-07-26
DOI
https://doi.org/10.1002/jcla.70317
Primary Topic
Colorectal Cancer Screening and Detection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A New Faecal Immunochemical Test Buffer With Improved Haemoglobin Stability Under Varying Temperature and Storage Times

Rachel Parker, Jean M. Winter, Geraldine Laven‐Law, Graeme P. Young et al.
Journal of Clinical Laboratory Analysis
Colorectal Cancer Screening and Detection
article

A New Faecal Immunochemical Test Buffer With Improved Haemoglobin Stability Under Varying Temperature and Storage Times

Rachel Parker, Jean M. Winter, Geraldine Laven‐Law, Graeme P. Young, Erin L. Symonds, Rafha Rafeeu, Karen Wang
article en

Abstract

BACKGROUND: Measurement of faecal haemoglobin (f-Hb) using the faecal immunochemical test (FIT) in population screening for colorectal cancer (CRC) has reduced cancer-related incidence and mortality. However, temperature-induced f-Hb degradation can decrease FIT accuracy. This study assessed the f-Hb preservation capacity of a new FIT Hb-stabilising buffer at different temperatures and durations. METHODS: Fresh whole faecal samples from individuals with active colorectal bleeding were used for collection of samples using the OC-Sensor FIT (EIKEN CHEMICAL CO. LTD, Japan) containing either the standard (SOC3) or new formulation (SOC4) buffer. Pooled faeces-buffer mixtures (n = 29) were divided into aliquots for incubation at 4°C, 23°C, 35°C, 45°C, and 50°C and analysed for f-Hb concentration at days 0, 3, 7, 10, 14, 21, and 28. Relative f-Hb (percentage change of day 0) in both buffer types was compared between temperatures at each timepoint. Generalised linear modelling identified variables influencing f-Hb degradation. RESULTS: SOC3 exhibited significant f-Hb decline from day 10 at 23°C and from day 3 at all higher temperatures assessed (p < 0.05). In contrast, f-Hb concentrations remained stable within SOC4 for up to 28 days at 23°C, with reductions observed after 7, 7, and 14 days at 50°C, 45°C, and 35°C, respectively (p < 0.05). Higher temperatures and longer storage duration significantly decreased f-Hb concentration, while SOC4 demonstrated superior f-Hb preservation relative to SOC3 (generalised linear model estimate = 15.4, p < 0.001). CONCLUSION: The SOC4 FIT buffer improves f-Hb stability compared with the current formulation and may improve CRC screening accuracy, particularly in high-temperature environments.

Journal of Clinical Laboratory Analysis
Flinders University (AU), Flinders Medical Centre (AU), Adelaide Institute for Sleep Health (AU), South Australian Health and Medical Research Institute (AU)
Good health and well-being
Openalex Percentile: Top 11%
Colorectal Cancer Screening and Detection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.