Default-threshold operating-point validation of a commercial chest radiograph AI system for selected CT-anchored thoracic findings: a bi-national multicenter retrospective study

Abstract This retrospective bi-national multicenter diagnostic accuracy study evaluated the locked default-threshold performance of a commercial chest radiograph AI system for selected thoracic findings using a CT-anchored, radiographic-detectability reference framework. Consecutive eligible adult patients who underwent frontal chest radiography and chest CT at four tertiary-care centers in Turkey and Bulgaria between June and December 2024 were included. Reference labels were assigned using temporally paired CT, with adjudication of whether CT-confirmed abnormalities had a corresponding radiographic manifestation on the paired chest radiograph. For potentially dynamic findings, the allowable CT–CXR interval was restricted to ≤ 2 days. The AI system was evaluated at the manufacturer’s default threshold of 0.50. Diagnostic performance was summarized using prevalence, sensitivity, specificity, positive predictive value, negative predictive value, and false-positive burden with 95% confidence intervals. The study included 940 patients with paired CXR-CT examinations. Reference-positive prevalence ranged from 2.2% for pneumothorax to 21.0% for pleural effusion. Sensitivity was highest for fracture (91%) and pneumothorax (90%) and lowest for atelectasis (64%); specificity ranged from 82% for consolidation/opacity to 98% for fracture and pneumothorax. PPV ranged from 43% to 63%, indicating a non-trivial false-positive burden at the evaluated operating point. Because continuous probability scores, human-reader comparison, and workflow outcomes were unavailable, these findings should be interpreted as default-threshold technical validation and support further evaluation of the system as radiologist-supervised decision support with local performance monitoring, rather than standalone diagnosis.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-07-25
DOI
https://doi.org/10.1038/s41598-026-55487-9
Primary Topic
COVID-19 diagnosis using AI
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Default-threshold operating-point validation of a commercial chest radiograph AI system for selected CT-anchored thoracic findings: a bi-national multicenter retrospective study

Yeliz Başar, Selin Ardalı Düzgün, Mustafa Ege Şeker, Gamze Durhan et al.
Scientific Reports
COVID-19 diagnosis using AI
article

Default-threshold operating-point validation of a commercial chest radiograph AI system for selected CT-anchored thoracic findings: a bi-national multicenter retrospective study

Yeliz Başar, Selin Ardalı Düzgün, Mustafa Ege Şeker, Gamze Durhan, Deniz Alış, Ercan Karaarslan, Recep Savaş, E. Karakoc, Nuri Saraç, Galina Kirova-Nedyalkova, Cemre Kavi, Mehmet Onur Önal, İlke Taşçı, Ahmet Karabulut, Melih Karadag, Elina Plamenova Milkovska, Boyan Emilov Manov, Kaan Buyukkirli, Atahan Tamturk
article en

Abstract

Abstract This retrospective bi-national multicenter diagnostic accuracy study evaluated the locked default-threshold performance of a commercial chest radiograph AI system for selected thoracic findings using a CT-anchored, radiographic-detectability reference framework. Consecutive eligible adult patients who underwent frontal chest radiography and chest CT at four tertiary-care centers in Turkey and Bulgaria between June and December 2024 were included. Reference labels were assigned using temporally paired CT, with adjudication of whether CT-confirmed abnormalities had a corresponding radiographic manifestation on the paired chest radiograph. For potentially dynamic findings, the allowable CT–CXR interval was restricted to ≤ 2 days. The AI system was evaluated at the manufacturer’s default threshold of 0.50. Diagnostic performance was summarized using prevalence, sensitivity, specificity, positive predictive value, negative predictive value, and false-positive burden with 95% confidence intervals. The study included 940 patients with paired CXR-CT examinations. Reference-positive prevalence ranged from 2.2% for pneumothorax to 21.0% for pleural effusion. Sensitivity was highest for fracture (91%) and pneumothorax (90%) and lowest for atelectasis (64%); specificity ranged from 82% for consolidation/opacity to 98% for fracture and pneumothorax. PPV ranged from 43% to 63%, indicating a non-trivial false-positive burden at the evaluated operating point. Because continuous probability scores, human-reader comparison, and workflow outcomes were unavailable, these findings should be interpreted as default-threshold technical validation and support further evaluation of the system as radiologist-supervised decision support with local performance monitoring, rather than standalone diagnosis.

Scientific Reports
University of Wisconsin–Madison (US), Adıyaman University (TR), Tokuda Hospital (BG), Acıbadem University (TR), Ege University (TR), Acibadem City Clinic (BG), Istanbul University-Cerrahpaşa (TR), Kent Hastanesi (TR), Acıbadem Adana Hospital (TR), Hacettepe University (TR)
Peace, Justice and strong institutions
Openalex Percentile: Top 8%
COVID-19 diagnosis using AI
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.