Robotic guidance for ultrasound-guided percutaneous procedures: a prospective multispecialty feasibility study

PURPOSE: To evaluate the safety, feasibility, and technical performance of a handheld robotic guidance system for ultrasound-guided percutaneous procedures across multiple procedural domains. METHODS: A prospective, single-center, open-label feasibility study evaluated the Focalist™ handheld robotic guidance system (Mendaera Inc., USA) across multiple ultrasound-guided percutaneous procedures was performed between December 3 and December 11, 2025. Patients undergoing clinically indicated organ access, vascular access, or pain management injection procedures for urologic conditions were prospectively enrolled. Primary endpoints were safety and technical success. Secondary endpoints included procedural performance metrics and operator-reported user experience. RESULTS: A total of 61 ultrasound-guided percutaneous procedures were performed in 46 patients by 11 operators. Procedure categories included 20 organ access procedures, 31 pain management injections, and 10 vascular access procedures, including 15 specific procedures that had never previously been performed using the system. Technical success was achieved in 60 of 61 procedures (98.4%), with no device-related adverse events. First-attempt success occurred in 45 procedures (74%), with a mean of 1.4 needle attempts per procedure. Mean robotic procedure time (sterile scan start to needle at target) was 5.6 min. Operator survey responses indicated agreement that the system was easy to use (mean 3.7/5) and enhanced confidence during procedures (mean 3.7/5). CONCLUSION: In this prospective feasibility study, a novel handheld robotic ultrasound guidance system was used successfully across a diverse range of ultrasound-guided percutaneous procedures. High technical success and absence of device-related adverse events support consideration of its inclusion in standard of care workflows. CTRI REGISTRATION: CTRI/2025/11/098124. GOV IDENTIFIER: NCT07358351.

Authors

Institutions

Publication Details

Journal
World Journal of Urology
Published
2026-07-25
DOI
https://doi.org/10.1007/s00345-026-06618-4
Primary Topic
Minimally Invasive Surgical Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Robotic guidance for ultrasound-guided percutaneous procedures: a prospective multispecialty feasibility study

Abhishek Singh, Mihir Desai, Kunal Kukreja, R.B. Sabnis et al.
World Journal of Urology
Minimally Invasive Surgical Techniques
article

Robotic guidance for ultrasound-guided percutaneous procedures: a prospective multispecialty feasibility study

Abhishek Singh, Mihir Desai, Kunal Kukreja, R.B. Sabnis, Arvind Ganpule, Mahesh Desai, Shreyans Jain, Abhijit Patil
article en

Abstract

PURPOSE: To evaluate the safety, feasibility, and technical performance of a handheld robotic guidance system for ultrasound-guided percutaneous procedures across multiple procedural domains. METHODS: A prospective, single-center, open-label feasibility study evaluated the Focalist™ handheld robotic guidance system (Mendaera Inc., USA) across multiple ultrasound-guided percutaneous procedures was performed between December 3 and December 11, 2025. Patients undergoing clinically indicated organ access, vascular access, or pain management injection procedures for urologic conditions were prospectively enrolled. Primary endpoints were safety and technical success. Secondary endpoints included procedural performance metrics and operator-reported user experience. RESULTS: A total of 61 ultrasound-guided percutaneous procedures were performed in 46 patients by 11 operators. Procedure categories included 20 organ access procedures, 31 pain management injections, and 10 vascular access procedures, including 15 specific procedures that had never previously been performed using the system. Technical success was achieved in 60 of 61 procedures (98.4%), with no device-related adverse events. First-attempt success occurred in 45 procedures (74%), with a mean of 1.4 needle attempts per procedure. Mean robotic procedure time (sterile scan start to needle at target) was 5.6 min. Operator survey responses indicated agreement that the system was easy to use (mean 3.7/5) and enhanced confidence during procedures (mean 3.7/5). CONCLUSION: In this prospective feasibility study, a novel handheld robotic ultrasound guidance system was used successfully across a diverse range of ultrasound-guided percutaneous procedures. High technical success and absence of device-related adverse events support consideration of its inclusion in standard of care workflows. CTRI REGISTRATION: CTRI/2025/11/098124. GOV IDENTIFIER: NCT07358351.

World Journal of UrologyVol. 44(1)
University of Southern California (US), Muljibhai Patel Urological Hospital (IN)
Openalex Percentile: Top 6%
Minimally Invasive Surgical Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.